Calibration of Indian Dobson Spectrophotometer Nos. D112 and D036 during WMO Dobson Intercomparison Campaigns

View through CrossRef
The present study summarizes the results of calibration of Indian Dobson Spectrophotometer No. D112 and D036 during Dobson inter-comparison campaigns, held at Irene, South Africa from 07 – 18 October, 2019 and at Hohenpeissenberg, Germany from 15.07.2020 to 30.07.2020 respectively. In both the international inter-comparison campaigns, European Regional Standard Dobson Spectrophotometer No. D064 had been used as the reference instrument. According to the results of the inter-comparison of Dobson spectrophotometer instrument no. D112 and no. D036 with Regional Standard Dobson Instrument (D064), these instruments are reliable and their measurements are accurate and appropriate for further analysis as their deviations have been found within the acceptable limits i.e. less than - 0.3% for D112 and less than 0.01 for D036 respectively after final calibration.
Title: Calibration of Indian Dobson Spectrophotometer Nos. D112 and D036 during WMO Dobson Intercomparison Campaigns
Description:
The present study summarizes the results of calibration of Indian Dobson Spectrophotometer No.
D112 and D036 during Dobson inter-comparison campaigns, held at Irene, South Africa from 07 – 18 October, 2019 and at Hohenpeissenberg, Germany from 15.
07.
2020 to 30.
07.
2020 respectively.
In both the international inter-comparison campaigns, European Regional Standard Dobson Spectrophotometer No.
D064 had been used as the reference instrument.
According to the results of the inter-comparison of Dobson spectrophotometer instrument no.
D112 and no.
D036 with Regional Standard Dobson Instrument (D064), these instruments are reliable and their measurements are accurate and appropriate for further analysis as their deviations have been found within the acceptable limits i.
e.
less than - 0.
3% for D112 and less than 0.
01 for D036 respectively after final calibration.

Related Results

(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
EGU ESSI–WMO–UNESCO Synergies for Interoperable Hydrological Data
EGU ESSI–WMO–UNESCO Synergies for Interoperable Hydrological Data
Many promising initiatives are advancing FAIR data in hydrology, yet a substantial semantic and technical interoperability gap remains between water data used in operational servic...
Il WMO-RTC di IBIMET-CNR: storia e attività
Il WMO-RTC di IBIMET-CNR: storia e attività
La rete dei Centri Regionali di Formazione della Organizzazione Meteorologica Mondiale è formata da 27 centri dislocati nelle sei Associazioni Regionali del WMO e coinvolge oltre 4...
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-S...
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Aiming at solving the problems such as long calibration time, low precision, and complex operation in hand-eye calibration of welding robot, an automatic hand-eye calibration algor...
Decentralisatie: maatwerk of uniformiteit?
Decentralisatie: maatwerk of uniformiteit?
Decentralization: custom-made or uniformity? A main motive for policy decentralization is the belief that municipalities are better able to customize public policy to loc...
From Strategy to Action: Strengthening Global Hydrological Research within WMO
From Strategy to Action: Strengthening Global Hydrological Research within WMO
The WMO Research Board Task Team on Hydrology Research (TT-HydroResearch) operated from September 2023 to December 2025 with the objective of strengthening coordination, coherence,...

Back to Top