Improved Understanding of Stent Malapposition Using Virtual Bench Testing
View through CrossRef
Intravascular imaging techniques such as optical coherence tomography (OCT) and intravascular ultrasound (IVUS) are often used to assess strut apposition, but only provide limited insight into the three-dimensional appositioning behaviour of stents. Recently, a new approach has been introduced to study the phenomenon of incomplete stent apposition (ISA) based on finite element simulations. In this study, we employed this virtual strut apposition assessment technique in the setting of coronary bifurcation stenting and compared simulated strut–artery distances of two stent designs with actual measurements based on OCT imaging using a silicone model. Stenting of the main branch leads to malapposed struts in the proximal part and the average strut–artery distance in that region for the Integrity stent is 126 μm based on the simulation and 117±14 μm based on the OCT analysis. For the Multi-Link 8 stent, this average distance is 150 μm and 174±7 µm for the simulation and thein vitroOCT measurements respectively. In conclusion, the virtual assessment of strut appositioning results in similar strut–artery distances when compared with measurements based on OCT-visualisedin vitrostent deployments and could be used to optimise devices and procedures.
Title: Improved Understanding of Stent Malapposition Using Virtual Bench Testing
Description:
Intravascular imaging techniques such as optical coherence tomography (OCT) and intravascular ultrasound (IVUS) are often used to assess strut apposition, but only provide limited insight into the three-dimensional appositioning behaviour of stents.
Recently, a new approach has been introduced to study the phenomenon of incomplete stent apposition (ISA) based on finite element simulations.
In this study, we employed this virtual strut apposition assessment technique in the setting of coronary bifurcation stenting and compared simulated strut–artery distances of two stent designs with actual measurements based on OCT imaging using a silicone model.
Stenting of the main branch leads to malapposed struts in the proximal part and the average strut–artery distance in that region for the Integrity stent is 126 μm based on the simulation and 117±14 μm based on the OCT analysis.
For the Multi-Link 8 stent, this average distance is 150 μm and 174±7 µm for the simulation and thein vitroOCT measurements respectively.
In conclusion, the virtual assessment of strut appositioning results in similar strut–artery distances when compared with measurements based on OCT-visualisedin vitrostent deployments and could be used to optimise devices and procedures.
Related Results
The promising role of OCT-VFR in guiding post-PCI optimization in CHIP patients: a VISION trial subanalysis
The promising role of OCT-VFR in guiding post-PCI optimization in CHIP patients: a VISION trial subanalysis
Abstract
Background
The integration of imaging and physiological assessments may provide optimal guidance for complex pro...
Novel functionalized and patterned surfaces for cardiovascular applications
Novel functionalized and patterned surfaces for cardiovascular applications
Nowadays, cardiovascular diseases are mainly treated by implantation of a metallic or polymeric mesh, called stent, which maintains the artery widely open. This technique shows ver...
Different vessel response between chronic total occlusions and non-CTO lesion after Sirolimus-eluting stent implantation: a serial optical coherence tomography study
Different vessel response between chronic total occlusions and non-CTO lesion after Sirolimus-eluting stent implantation: a serial optical coherence tomography study
Objectives
To investigate the neointimal coverage and malapposition of sirolimus-eluting stent in treatment of CTOs and non-CTO lesions (including lipid-rich plaq...
CLINICAL APPLICATION OF DOMESTIC LONG BIODEGRADABLE POLYMER COATED SIROLIMUS ELUTING STENTS FOR TREATMENT OF LONG CORONARY ARTERY LESIONS
CLINICAL APPLICATION OF DOMESTIC LONG BIODEGRADABLE POLYMER COATED SIROLIMUS ELUTING STENTS FOR TREATMENT OF LONG CORONARY ARTERY LESIONS
Objectives
To evaluate the efficacy and safety of domestic long biodegradable polymer coated sirolimus eluting stents (Excel stent, Shandong JW) for treatment of ...
Effect of co-trimoxazole and N-acetylcysteine alone and in combination on bacterial adherence on ureteral stent surface
Effect of co-trimoxazole and N-acetylcysteine alone and in combination on bacterial adherence on ureteral stent surface
Abstract
To assess the effect of co-trimoxazole and N-acetylcysteine (NAC), alone and in combination, on bacterial adherence to biofilm forme...
Minimum stent area, stent expansion and ultrasonic flow ratio post pci: insights from multivessel talent trial
Minimum stent area, stent expansion and ultrasonic flow ratio post pci: insights from multivessel talent trial
Abstract
Background
Ultrasonic flow ratio (UFR) is a novel method for fast computation of fractional flow reserve (FFR) from int...
One stent versus two stents for distal LM PCI: insights from the experience of a high volume center
One stent versus two stents for distal LM PCI: insights from the experience of a high volume center
Abstract
Introduction
Distal left main (LM) PCIremains a challenge. One of the most debated issues is whether to use a single vs...
Fluid and Solid Mechanics in Stented Arteries
Fluid and Solid Mechanics in Stented Arteries
Abstract
Vascular stents are being used in increasing numbers to correct arterial flow limiting disorders. Although data exist on the in vivo performance of these de...