Javascript must be enabled to continue!
Effect of Amusement Park Rides on Programmable Shunt Valve Settings
View through CrossRef
<b><i>Background:</i></b> Magnetically programmable shunt valves are susceptible to environmental factors including magnetic fields and accelerative forces. It is unknown if rollercoasters with or without magnetic brakes or linear induction motors (LIMs) are capable of altering the setting of a programmable shunt valve. <b><i>Methods:</i></b> Two different valve types (type A, n = 10; type B, n = 9) were tested at varying resistance settings in 2 trials on 6 different amusement park rides including 2 rides with LIMs, 2 rides with magnetic brakes, and 2 rides without magnetic technology. <b><i>Results:</i></b> The performance level of valve type A and the setting of valve type B changed on rollercoasters with magnets (A = 2.5% [2/80]; B = 5.6% [4/72]) and without magnets (A = 7.5% [3/40]; B = 2.8% [1/36]). Neither valve setting changed when exposed to a Ferris wheel or during ambulation throughout the park. <b><i>Conclusion:</i></b> Magnetically programmable valves are susceptible to changes in pressure settings when exposed to amusement park rides with elevated vertical gravitational forces, irrespective of the presence of LIMs or magnetic brakes.
Title: Effect of Amusement Park Rides on Programmable Shunt Valve Settings
Description:
<b><i>Background:</i></b> Magnetically programmable shunt valves are susceptible to environmental factors including magnetic fields and accelerative forces.
It is unknown if rollercoasters with or without magnetic brakes or linear induction motors (LIMs) are capable of altering the setting of a programmable shunt valve.
<b><i>Methods:</i></b> Two different valve types (type A, n = 10; type B, n = 9) were tested at varying resistance settings in 2 trials on 6 different amusement park rides including 2 rides with LIMs, 2 rides with magnetic brakes, and 2 rides without magnetic technology.
<b><i>Results:</i></b> The performance level of valve type A and the setting of valve type B changed on rollercoasters with magnets (A = 2.
5% [2/80]; B = 5.
6% [4/72]) and without magnets (A = 7.
5% [3/40]; B = 2.
8% [1/36]).
Neither valve setting changed when exposed to a Ferris wheel or during ambulation throughout the park.
<b><i>Conclusion:</i></b> Magnetically programmable valves are susceptible to changes in pressure settings when exposed to amusement park rides with elevated vertical gravitational forces, irrespective of the presence of LIMs or magnetic brakes.
Related Results
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
American Institute of Mining, Metallurgical, and Petroleum Engineers, Inc.
Abstract
A surface-controlled subsurface safety valve...
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Objective: To compare the heart valve circumference before and after 10% formalin fixation. Materials and Methods: The study analyzed 63 Thai human cadaveric hearts. Each heart val...
Reciprocating compressor valve damage estimation under varying speeds through the acoustic emission technique
Reciprocating compressor valve damage estimation under varying speeds through the acoustic emission technique
Purpose
The purpose of this paper is to examine the effect of reciprocating compressor speeds and valve conditions on the roor-mean-square (RMS) value of burst acoustic emission (A...
Abstract 10999: Long-Term Outcomes of Atrioventricular Valve Regurgitation and Repair in Patients With Single Ventricle Physiology
Abstract 10999: Long-Term Outcomes of Atrioventricular Valve Regurgitation and Repair in Patients With Single Ventricle Physiology
Introduction:
In single ventricle patients, AV valve regurgitation increases the risk of adverse outcomes, and staged palliation with concomitant AV valve intervention ...
Ventriculoportal Shunt, a New Transomphalic Extraperitoneal Surgical Technique in Treatment of Hydrocephalus
Ventriculoportal Shunt, a New Transomphalic Extraperitoneal Surgical Technique in Treatment of Hydrocephalus
Aim. The aim of this article was to report a new transomphalic extraperitoneal surgical technique for treatment of hydrocephalus, called ventriculoportal shunt. Materials and Metho...
<b>COMPARISON OF SUCCESS RATE OF FRONTAL VERSUS OCCIPITAL ENTRY SITE FOR VENTRICULOPERITONEAL SHUNT INSERTION IN PATIENTS WITH HYDROCEPHALUS</b>
<b>COMPARISON OF SUCCESS RATE OF FRONTAL VERSUS OCCIPITAL ENTRY SITE FOR VENTRICULOPERITONEAL SHUNT INSERTION IN PATIENTS WITH HYDROCEPHALUS</b>
Background: Hydrocephalus is a common neurological disorder characterized by an abnormal accumulation of cerebrospinal fluid within the brain ventricles. Surgical interventions, su...
Branch Barrier Valve for Subsea Application
Branch Barrier Valve for Subsea Application
ABSTRACT
A new type subsea repairable pipeline branch barrier valve, invented by TOTAL & STATOIL, has been developed by CAMERON based on the ram type blowout ...
Role of intrathecal and topical vancomycin in prevention of ventriculoperitoneal shunt infections
Role of intrathecal and topical vancomycin in prevention of ventriculoperitoneal shunt infections
The objective of this study is to assess the role of intrathecal and topical vancomycin in the prevention of ventriculoperitoneal shunt infection. Infections of the ventriculoperit...