SU‐E‐I‐27: Establishing Target Exposure Index Values for Computed Radiography

View through CrossRef
Purpose:To develop a standard set of target exposure index (TEI) values to be applied to Agfa Computed Radiography (CR) readers in accordance with International Electrotechnical Committee 62494‐1 (ed. 1.0).Methods:A large data cohort was collected from six USAF Medical Treatment Facilities that exclusively use Agfa CR Readers. Dose monitoring statistics were collected from each reader. The data was analyzed based on anatomic region, view, and processing speed class. The Agfa specific exposure metric, logarithmic mean (LGM), was converted to exposure index (EI) for each data set. The optimum TEI value was determined by minimizing the number of studies that fell outside the acceptable deviation index (DI) range of +/− 2 for phototimed techniques or a range of +/−3 for fixed techniques. An anthropomorphic radiographic phantom was used to corroborate the TEI recommendations. Images were acquired of several anatomic regions and views using standard techniques. The images were then evaluated by two radiologists as either acceptable or unacceptable. The acceptable image with the lowest exposure and EI value was compared to the recommended TEI values using a passing DI range.Results:Target EI values were determined for a comprehensive list of anatomic regions and views.Conclusion:Target EI values must be established on each CR unit in order to provide a positive feedback system for the technologist. This system will serve as a mechanism to prevent under or overexposures of patients. The TEI recommendations are a first attempt at a large scale process improvement with the goal of setting reasonable and standardized TEI values. The implementation and effectiveness of the recommended TEI values should be monitored and adjustments made as necessary.
Title: SU‐E‐I‐27: Establishing Target Exposure Index Values for Computed Radiography
Description:
Purpose:To develop a standard set of target exposure index (TEI) values to be applied to Agfa Computed Radiography (CR) readers in accordance with International Electrotechnical Committee 62494‐1 (ed.
1.
0).
Methods:A large data cohort was collected from six USAF Medical Treatment Facilities that exclusively use Agfa CR Readers.
Dose monitoring statistics were collected from each reader.
The data was analyzed based on anatomic region, view, and processing speed class.
The Agfa specific exposure metric, logarithmic mean (LGM), was converted to exposure index (EI) for each data set.
The optimum TEI value was determined by minimizing the number of studies that fell outside the acceptable deviation index (DI) range of +/− 2 for phototimed techniques or a range of +/−3 for fixed techniques.
An anthropomorphic radiographic phantom was used to corroborate the TEI recommendations.
Images were acquired of several anatomic regions and views using standard techniques.
The images were then evaluated by two radiologists as either acceptable or unacceptable.
The acceptable image with the lowest exposure and EI value was compared to the recommended TEI values using a passing DI range.
Results:Target EI values were determined for a comprehensive list of anatomic regions and views.
Conclusion:Target EI values must be established on each CR unit in order to provide a positive feedback system for the technologist.
This system will serve as a mechanism to prevent under or overexposures of patients.
The TEI recommendations are a first attempt at a large scale process improvement with the goal of setting reasonable and standardized TEI values.
The implementation and effectiveness of the recommended TEI values should be monitored and adjustments made as necessary.

Related Results

Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Background: Hundred patients were examined to determine the effectiveness of digital radiography in detecting Orthopedic injuries. The decisive findings showed that digital radiogr...
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
Background: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) ...
Plain radiography, excretion radiography, and contrast radiography
Plain radiography, excretion radiography, and contrast radiography
Abstract Plain radiography is both a primary abdominal examination technique and an initial component of subsequent excretion and contrast radiographic studies. Excr...
Simulating upright cervical lordosis in the supine position
Simulating upright cervical lordosis in the supine position
Cervical alignment or lordosis evolution is still attained by direct radiography in standing position because an ideal cervical curvature is essential to maintain a horizontal gaze...
Low-dose, X-ray-based Imaging of Lung Function
Low-dose, X-ray-based Imaging of Lung Function
<p>Chronic obstructive pulmonary disease (COPD) is characterized by chronic airflow limitation resulting from emphysema and small airway disease. Functional lung imaging of C...

Back to Top