Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
View through CrossRef
Background:
The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom.
Methods:
The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom. Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems. The images from the CDRAD phantom were evaluated by three observers. The results were displayed in the form of a contrast-detail (CD) curve. In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance.
Results:
Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems. For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used. The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5.
Conclusion:
The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study. A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality. Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios.
Ovid Technologies (Wolters Kluwer Health)
Title: Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
Description:
Background:
The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom.
Methods:
The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom.
Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems.
The images from the CDRAD phantom were evaluated by three observers.
The results were displayed in the form of a contrast-detail (CD) curve.
In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance.
Results:
Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems.
For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used.
The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5.
Conclusion:
The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study.
A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality.
Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios.
Related Results
Access Denied
Access Denied
Introduction
As social-distancing mandates in response to COVID-19 restricted in-person data collection methods such as participant observation and interviews, researchers turned t...
Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Background: Hundred patients were examined to determine the effectiveness of digital radiography in detecting Orthopedic injuries. The decisive findings showed that digital radiogr...
The role of chest radiography in the diagnosis of bacteriologically confirmed pulmonary tuberculosis in hospitalised Xpert MTB/RIF-negative patients
The role of chest radiography in the diagnosis of bacteriologically confirmed pulmonary tuberculosis in hospitalised Xpert MTB/RIF-negative patients
The role of chest radiography to diagnose active tuberculosis in symptomatic patients who have a negative Xpert MTB/RIF (Xpert) test result is unclear. This study aimed to assess t...
Comparative Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Bone Fractures
Comparative Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Bone Fractures
Background: Bone fractures represent a substantial proportion of cases presenting to emergency and orthopedic departments worldwide, including in resource-limited settings like Nis...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract
The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...
The influence of micro influencers and digital marketing on product purchasing decisions at tiktok shop in bengkulu city
The influence of micro influencers and digital marketing on product purchasing decisions at tiktok shop in bengkulu city
THE INFLUENCE OF MICRO-INFLUENCERS AND DIGITAL MARKETING ON PURCHASE DECISIONS OF TIKTOK SHOP CUSTOMERS IN BENGKULU CITY
Andhes Tiani Putri, Meylaty F
12Faculty Of Economic
E...
Research on experiment and Engineering Application of 9%Ni LNG Steel Storage Tank butt joint using digital radiography testing technique
Research on experiment and Engineering Application of 9%Ni LNG Steel Storage Tank butt joint using digital radiography testing technique
9%Ni steel process good mechanical properties, welding ability and low-temperature toughness, and has been widely used in LNG low-temperature storage tanks. Due to the existence of...
Pengembangan Parameter Penilaian Keamanan Pelayanan Kesehatan Tradisional Empiris
Pengembangan Parameter Penilaian Keamanan Pelayanan Kesehatan Tradisional Empiris
Abstract
In the context of protecting the community against the security of traditional empirical health services, more detailed safety assessment parameters have been dev...