The difference of removal force values on various preload application on 5º taper implant transfer abutment to determine the optimal preload
View through CrossRef
Background: Dental implants successfully become long-term solution in Indonesia with the two-piece variety being prevalent. The 5º tapered internal connections are favored for their stability compared to external connections. Preload is essential for stability relying on friction between the implant-abutment connection. Purpose: Evaluate the differences in removal force value resulting from various preload applications on 5º taper implant transfer abutment to determine the optimal preload. Methods: Implant abutments with 5º taper implant transfer abutments were evaluated in the study. The control group of this study derived from the manufacturer’s preload recommendation of 35 Ncm. Preloads from 10 Ncm, 15 Ncm, 20 Ncm, 25 Ncm, 30 Ncm, 40 Ncm, 45 Ncm, 50 Ncm were used as independent variables in this experiment (n=3). The abutment screw was removed before the tensile test using a universal testing machine (UTM). Statistical data analysis was conducted using the One-way ANOVA and Tukey HSD (p<0,05). Results: The 35 Ncm preload has the highest result with a mean value of (723 N ± 53.8). The 10 Ncm preload had the lowest results with a mean value of (267 N ± 71.4). There was no remarkable difference between the 30 Ncm and the 35 Ncm preload groups. Conclusion: There is a difference in removal force values at various preloads given. The removal force value at 35 Ncm is the optimal preload recommended by the manufacturer in line with a preload of 30 Ncm which is not significantly different from 35 Ncm.
Title: The difference of removal force values on various preload application on 5º taper implant transfer abutment to determine the optimal preload
Description:
Background: Dental implants successfully become long-term solution in Indonesia with the two-piece variety being prevalent.
The 5º tapered internal connections are favored for their stability compared to external connections.
Preload is essential for stability relying on friction between the implant-abutment connection.
Purpose: Evaluate the differences in removal force value resulting from various preload applications on 5º taper implant transfer abutment to determine the optimal preload.
Methods: Implant abutments with 5º taper implant transfer abutments were evaluated in the study.
The control group of this study derived from the manufacturer’s preload recommendation of 35 Ncm.
Preloads from 10 Ncm, 15 Ncm, 20 Ncm, 25 Ncm, 30 Ncm, 40 Ncm, 45 Ncm, 50 Ncm were used as independent variables in this experiment (n=3).
The abutment screw was removed before the tensile test using a universal testing machine (UTM).
Statistical data analysis was conducted using the One-way ANOVA and Tukey HSD (p<0,05).
Results: The 35 Ncm preload has the highest result with a mean value of (723 N ± 53.
8).
The 10 Ncm preload had the lowest results with a mean value of (267 N ± 71.
4).
There was no remarkable difference between the 30 Ncm and the 35 Ncm preload groups.
Conclusion: There is a difference in removal force values at various preloads given.
The removal force value at 35 Ncm is the optimal preload recommended by the manufacturer in line with a preload of 30 Ncm which is not significantly different from 35 Ncm.
Related Results
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Implant‐Abutment Interface: Biomechanical Study of Flat Top versus Conical
Implant‐Abutment Interface: Biomechanical Study of Flat Top versus Conical
ABSTRACT Background: Overloading has been identified as a primary factor behind dental implant failure. The peak bone stresses normally appear in the marginal bone. The anchorage s...
Dental Implant Abutments and Their Selection - A Review
Dental Implant Abutments and Their Selection - A Review
Dental implants have become the most promising and accepted prosthetic alternative to missing teeth. Innovative and technological advances in implant dentistry have improved the pe...
REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
Introduction. Replacement of a small included defect in the frontal area with a fixed structure based on the implant is the optimal solution. It is not always required to remove th...
Stress analysis in implant, abutment, and peripheral bone with different restorative crown and abutment materials: A three-dimensional finite element analysis study
Stress analysis in implant, abutment, and peripheral bone with different restorative crown and abutment materials: A three-dimensional finite element analysis study
Background:
Stresses on prosthetic crown directly influences the survival rate of implants hence it should be considered while selecting prosthetic material.
...
P199 Glucocorticoid taper success after one and two years treatment in the Imperial College Takayasu Arteritis Cohort
P199 Glucocorticoid taper success after one and two years treatment in the Imperial College Takayasu Arteritis Cohort
Abstract
Background/Aims
The large vessel vasculitis Takayasu arteritis (TA) is initially treated with high dose glucocorticoids...
Influence of Roxolid Implant Material on The Implant Stability of Maxillary Implant Retained Overdenture
Influence of Roxolid Implant Material on The Implant Stability of Maxillary Implant Retained Overdenture
Abstract
Background
Long-term success of implant restoration depends on many factors one of them is the sufficient implant stability which is lowered in compromised bone d...
<span style="mso-ansi-language: EN;">Retrieval of Fractured Abutment Screw of Dental Implant. Case Report<span style="font-size: 10.0pt; mso-ansi-font-size: 18.0pt; font-family: 'Times New Roman',serif; mso-ascii-font-family: 'Palatino Linotype';
<span style="mso-ansi-language: EN;">Retrieval of Fractured Abutment Screw of Dental Implant. Case Report<span style="font-size: 10.0pt; mso-ansi-font-size: 18.0pt; font-family: 'Times New Roman',serif; mso-ascii-font-family: 'Palatino Linotype';
Background: The advancement of dental implants marks a significant milestone in modern dentistry, offering an effective solution for tooth replacement while improving both function...