Fetal Doppler Simulator Based on Arduino
View through CrossRef
Heart rate of the fetal is the main indicator of the fetal life in the womb. Monitoring fetal heart rate can’t be done, so a tool is needed to monitoring fetal heart rate. Fetal heart rate can be monitored with fetal doppler. To test the accuracy of Fetal Doppler, a calibration is needed with the Fetal Doppler Simulator. This tool will simulate the fetal heart rate with a BPM value that can be adjusted according to the settings on the device. This module using Arduino as the brain system. On the module there is a selection of BPM from 60 to 240 BPM with an increase of 30 BPM displayed on 2x16 character LCDs. Based on BPM measurement 6 times using Fetal Doppler, the measurement error in a BPM of 60 to BPM 210 is 0%, while at BPM 240 an error is 0.2%. This module has been compared with the standard devices (Fetal Simulator Brand Fluke Biomedical Ps320), the results of the comparison modules with the comparison tool has the same error value in 240 BPM is 0.2% and in BPM 210 there is a difference in the result of module Fetal Doppler reading of 210 BPM while in the comparison tool is 209 BPM. Of the measurement data and analysis, it can be concluded that the tool can work and the tool has the same accuracy as the standard device.
Title: Fetal Doppler Simulator Based on Arduino
Description:
Heart rate of the fetal is the main indicator of the fetal life in the womb.
Monitoring fetal heart rate can’t be done, so a tool is needed to monitoring fetal heart rate.
Fetal heart rate can be monitored with fetal doppler.
To test the accuracy of Fetal Doppler, a calibration is needed with the Fetal Doppler Simulator.
This tool will simulate the fetal heart rate with a BPM value that can be adjusted according to the settings on the device.
This module using Arduino as the brain system.
On the module there is a selection of BPM from 60 to 240 BPM with an increase of 30 BPM displayed on 2x16 character LCDs.
Based on BPM measurement 6 times using Fetal Doppler, the measurement error in a BPM of 60 to BPM 210 is 0%, while at BPM 240 an error is 0.
2%.
This module has been compared with the standard devices (Fetal Simulator Brand Fluke Biomedical Ps320), the results of the comparison modules with the comparison tool has the same error value in 240 BPM is 0.
2% and in BPM 210 there is a difference in the result of module Fetal Doppler reading of 210 BPM while in the comparison tool is 209 BPM.
Of the measurement data and analysis, it can be concluded that the tool can work and the tool has the same accuracy as the standard device.
Related Results
Rancang Bangun Fetal Doppler Simulator untuk Kalibrasi Fetal Doppler
Rancang Bangun Fetal Doppler Simulator untuk Kalibrasi Fetal Doppler
Fetal Doppler Simulator untuk kalibrasi merupakan alat yang dirancang guna memastikan akurasi serta keandalan perangkat Fetal Doppler dalam memantau detak jantung janin. Selama ini...
Simulator approaches in otorhinolaryngology ‐ Looking beyond the ophthalmological horizon
Simulator approaches in otorhinolaryngology ‐ Looking beyond the ophthalmological horizon
Purpose: To improve medical students' learning of handheld otoscopy technique and findings based on a standardized simulator‐based procedure.Methods: A group of 120 medical student...
Ovine fetal metabolism during norepinephrine infusion
Ovine fetal metabolism during norepinephrine infusion
Although stress in fetal life not only increases fetal catecholamine concentration but also decreases fetal growth, there have been few studies that define the specific role of cat...
Application of Arduino Devices in various IOT Application
Application of Arduino Devices in various IOT Application
In this paper, v segribe the wireless sensor network system is the next open source hardware we've graded or raspberry arduino refers to an open source electronic operating system ...
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Purpose
The purpose of this study was to analyze the effects of past participation in athletics, the playing of musical instruments and video games and other va...
Etiology and Outcome of Isolated Fetal Ascites
Etiology and Outcome of Isolated Fetal Ascites
OBJECTIVE:
To describe the etiology of isolated fetal ascites and associated perinatal outcomes, and to assess the progression of isolated fetal ascites to fetal hydrop...
ANALISIS EFEK KECELAKAAN WATER INGRESS TERHADAP REAKTIVITAS DOPPLER TERAS RGTT200K
ANALISIS EFEK KECELAKAAN WATER INGRESS TERHADAP REAKTIVITAS DOPPLER TERAS RGTT200K
Dalam high temperature reactor, koefisien reaktivitas temperatur yang didesain negatif menjamin reaksi fisi dalam teras tetap berada di bawah kendali dan panas peluruhan tidak akan...
World Congress of Perinatal Medicine
World Congress of Perinatal Medicine
SPEAKERS ABSTRACTS
The application of super slow review of ultrasound clips to a diagnosis of fetal tachyarrhythmia
Name and sur...