Rancang Bangun Fetal Doppler Simulator untuk Kalibrasi Fetal Doppler
View through CrossRef
The Fetal Doppler Simulator for calibration is a device designed to ensure the accuracy and reliability of Fetal Doppler equipment in monitoring fetal heart rate. Traditionally, calibration of Fetal Doppler devices has relied on limited methods with relatively high costs. Therefore, an alternative calibration medium that is more economical, practical, and reliable is needed. This simulator functions by generating fetal heartbeat signals with various rhythm patterns, ranging from normal conditions to pathological variations such as tachycardia and bradycardia. The research method applied was Research and Development (R&D), which included needs analysis, design planning, feature integration, prototype construction, and performance testing. The test results showed that the Fetal Doppler Simulator was able to generate stable heartbeat simulations with measurement values that remained within acceptable tolerance limits (≤5%). Based on these findings, it can be concluded that the development of this Fetal Doppler Simulator is feasible and effective as a calibration medium. This research is expected to offer a practical alternative for healthcare facilities to perform Fetal Doppler calibration at a lower cost.
Politeknik Bina Trada Semarang
Title: Rancang Bangun Fetal Doppler Simulator untuk Kalibrasi Fetal Doppler
Description:
The Fetal Doppler Simulator for calibration is a device designed to ensure the accuracy and reliability of Fetal Doppler equipment in monitoring fetal heart rate.
Traditionally, calibration of Fetal Doppler devices has relied on limited methods with relatively high costs.
Therefore, an alternative calibration medium that is more economical, practical, and reliable is needed.
This simulator functions by generating fetal heartbeat signals with various rhythm patterns, ranging from normal conditions to pathological variations such as tachycardia and bradycardia.
The research method applied was Research and Development (R&D), which included needs analysis, design planning, feature integration, prototype construction, and performance testing.
The test results showed that the Fetal Doppler Simulator was able to generate stable heartbeat simulations with measurement values that remained within acceptable tolerance limits (≤5%).
Based on these findings, it can be concluded that the development of this Fetal Doppler Simulator is feasible and effective as a calibration medium.
This research is expected to offer a practical alternative for healthcare facilities to perform Fetal Doppler calibration at a lower cost.
Related Results
Estimasi Angka Ketidakpastian Asal Kalibrasi Instrumen Terhadap Pengujian Kualitas CPO
Estimasi Angka Ketidakpastian Asal Kalibrasi Instrumen Terhadap Pengujian Kualitas CPO
Telah dilakukan kajian pengaruh kalibrasi peralatan pengujian moisture and impurities (M&I) dan deterioration of bleachability index (DOBI) pada crude palm oil (CPO). Kalibrasi...
Fetal Doppler Simulator Based on Arduino
Fetal Doppler Simulator Based on Arduino
Heart rate of the fetal is the main indicator of the fetal life in the womb. Monitoring fetal heart rate can’t be done, so a tool is needed to monitoring fetal heart rate. Fetal he...
Fetal Arrhythmias: A Clinical Review
Fetal Arrhythmias: A Clinical Review
ABSTRACT
Fetal rhythm abnormalities occur in 2% of pregnancies. They are usually identified by the obstetrician or midwifes after 20 weeks. There are four different methods used to...
JAUBAK: Simulator 3D Peninjauan Penembakan Mobile
JAUBAK: Simulator 3D Peninjauan Penembakan Mobile
Kebutuhan teknis dan taktis yang dapat mensimulasikan kegiatan peninjauan secara nyata seperti kegiatan yang sebenarnya di lapangan sehingga gumil atau pelatih mendapatkan sarana/p...
ANALISIS DAN KALIBRASI PARAMETER DRIVING BEHAVIOR PADA SIMPANG TAK BERSINYAL JALAN PERKOTAAN
ANALISIS DAN KALIBRASI PARAMETER DRIVING BEHAVIOR PADA SIMPANG TAK BERSINYAL JALAN PERKOTAAN
VISSIM adalah software mikro-simulasi arus lalu lintas multimoda yang digunakan secara luas dalam menentukan kinerja lalu lintas dan perencanaan alternatif. Namun, karena pengatura...
Simulator approaches in otorhinolaryngology ‐ Looking beyond the ophthalmological horizon
Simulator approaches in otorhinolaryngology ‐ Looking beyond the ophthalmological horizon
Purpose: To improve medical students' learning of handheld otoscopy technique and findings based on a standardized simulator‐based procedure.Methods: A group of 120 medical student...
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Purpose
The purpose of this study was to analyze the effects of past participation in athletics, the playing of musical instruments and video games and other va...
Determination of fetal weight by ultrasonographic evaluation of fetal mid-thigh soft-tissue thickness in late third trimester
Determination of fetal weight by ultrasonographic evaluation of fetal mid-thigh soft-tissue thickness in late third trimester
Background: Estimation of fetal weight is of utmost importance in the planning and management of uneventful labor. This study was done to compare the estimated fetal weight determi...