SU‐E‐T‐195: Commissioning the Neutron Production of a Varian TrueBeam Linac
View through CrossRef
Purpose:The purpose of this work is the characterization of a new Varian TrueBeam™ facility in terms of neutron production, in order to estimate neutron equivalent dose in organs during radiotherapy treatments.Methods:The existing methodology [1] was used with the reference SRAMnd detector, calibrated in terms of thermal neutron fluence at the reference field operated by PTB (Physikalisch‐Technische‐Bundesanstalt) at the GeNF (Geesthacht‐Neutron‐Facility) with the GKSS reactor FRG‐1 [2]. Thermal neutron fluence for the 5 available possibilities was evaluated: 15 MV and 10&6 MV with and without Flattening Filter (FF and FFF, respectively). Irradiation conditions are as described in [3]. In addition, three different collimator‐MLC configurations were studied for 15 MV: (a) collimator of 10×10 cm2 and MLC fully retracted (reference), (b) field sizes of 20×20 cm2 and 10×10 cm2 for collimator and MLC respectively, and (c) collimator and MLC aperture of 10×10 cm2.Results:Thermal fluence rate at the “reference point” [3], as a consequence of the neutron production, obtained for (a) conformation in 15 MV is (1.45±0.11) x10⁴ n•cm2/MU. Configurations (b) and (c) gave fluences of 96.6% and 97.8% of the reference (a). Neutron production decreases up to 8.6% and 5.7% for the 10 MV FF and FFF beams, respectively. Finally, it decreases up to 2.8% and 0.1% for the 6 MV FF and FFF modes, respectively.Conclusion:This work evaluates thermal neutron production of Varian TrueBeam™ system for organ equivalent dose estimation. The small difference in collimator‐MLC configuration shows the universality of the methodology [3]. A decrease in this production is shown when decreasing energy from 15 to 10 MV and an almost negligible production was found for 6 MV. Moreover, a lower neutron contribution is observed for the FFF modes.[1]Phys Med Biol,2012;57:6167–6191.[2]Radiat Meas,2010;45:1513–1517.[3]Med Phys,2015;42:276–281.
Title: SU‐E‐T‐195: Commissioning the Neutron Production of a Varian TrueBeam Linac
Description:
Purpose:The purpose of this work is the characterization of a new Varian TrueBeam™ facility in terms of neutron production, in order to estimate neutron equivalent dose in organs during radiotherapy treatments.
Methods:The existing methodology [1] was used with the reference SRAMnd detector, calibrated in terms of thermal neutron fluence at the reference field operated by PTB (Physikalisch‐Technische‐Bundesanstalt) at the GeNF (Geesthacht‐Neutron‐Facility) with the GKSS reactor FRG‐1 [2].
Thermal neutron fluence for the 5 available possibilities was evaluated: 15 MV and 10&6 MV with and without Flattening Filter (FF and FFF, respectively).
Irradiation conditions are as described in [3].
In addition, three different collimator‐MLC configurations were studied for 15 MV: (a) collimator of 10×10 cm2 and MLC fully retracted (reference), (b) field sizes of 20×20 cm2 and 10×10 cm2 for collimator and MLC respectively, and (c) collimator and MLC aperture of 10×10 cm2.
Results:Thermal fluence rate at the “reference point” [3], as a consequence of the neutron production, obtained for (a) conformation in 15 MV is (1.
45±0.
11) x10⁴ n•cm2/MU.
Configurations (b) and (c) gave fluences of 96.
6% and 97.
8% of the reference (a).
Neutron production decreases up to 8.
6% and 5.
7% for the 10 MV FF and FFF beams, respectively.
Finally, it decreases up to 2.
8% and 0.
1% for the 6 MV FF and FFF modes, respectively.
Conclusion:This work evaluates thermal neutron production of Varian TrueBeam™ system for organ equivalent dose estimation.
The small difference in collimator‐MLC configuration shows the universality of the methodology [3].
A decrease in this production is shown when decreasing energy from 15 to 10 MV and an almost negligible production was found for 6 MV.
Moreover, a lower neutron contribution is observed for the FFF modes.
[1]Phys Med Biol,2012;57:6167–6191.
[2]Radiat Meas,2010;45:1513–1517.
[3]Med Phys,2015;42:276–281.
Related Results
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Comparative study of the effect of neutrons emitted from neutron source 241Am-Be and curcumin on MCF-7 breast cancer cells in 3D culture medium
Comparative study of the effect of neutrons emitted from neutron source 241Am-Be and curcumin on MCF-7 breast cancer cells in 3D culture medium
AbstractIntroductionCancer is one of the major medical problems threatening human health. Breast cancer is the most prevalent type of cancer in women. Reports indicate that treatme...
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Abstract
Background: Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geometrical description of th...
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Abstract
Background
Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geomet...
Study of coded source neutron imaging based on a compact accelerator
Study of coded source neutron imaging based on a compact accelerator
Compact accelerator based neutron source has lower cost and better flexibility than nuclear reactor. Neutron imaging using such a neutron source has attracted more and more attenti...
A New Method of Porosity Determination by D-T Neutron Generator and Dual CLYC Detector
A New Method of Porosity Determination by D-T Neutron Generator and Dual CLYC Detector
Porosity is one of the essential parameters in conventional oil and gas reservoir evaluation, as well as plays an important role in the calculation of formation saturation and rese...
Penyebaran sistem vokal DMP di Yaha
Penyebaran sistem vokal DMP di Yaha
Kajian ini dilakukan bertujuan melihat ciri linguistik bagi setiap varian yang wujud di Yaha. Bentuk vokal dilihat dari segi kesejajaran dan penyebarannya dalam ruang geografi, sam...
MicroRNA‐195 regulates blood pressure by inhibiting NKCC2A
MicroRNA‐195 regulates blood pressure by inhibiting NKCC2A
We previously showed that inhibition of renal tumor necrosis factor‐alpha (TNF) production induced by high salt intake increases Na+‐K+‐2Cl−cotransporter isoform A (NKCC2A) mRNA, p...