Javascript must be enabled to continue!
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
View through CrossRef
Abstract
Background
Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geometrical description of the linac head. Since the geometry of Varian TrueBeam machines has not been disclosed, the manufacturer distributes phase-space files of the linac patient-independent part to allow researchers to compute absorbed dose distributions using the Monte Carlo method. This approach limits the possibility of achieving an arbitrarily small statistical uncertainty. This work investigates the use of the geometry of the Varian Clinac 2100, which is included in the Monte Carlo system PRIMO, as a surrogate.
Methods
Energy, radial and angular distributions extracted from the TrueBeam phase space files published by the manufacturer and from phase spaces tallied with PRIMO for the Clinac 2100 were compared for the 6, 8, 10 and 15 MV flattened-filtered beams. Dose distributions in water computed for the two sets of PSFs were compared with the Varian Representative Beam Data (RBD) for square fields with sides ranging from 3 to 30 cm. Output factors were calculated for square fields with sides ranging from 2 to 40 cm.
Results
Excellent agreement with the RBD was obtained for the simulations that employed the phase spaces distributed by Varian as well as for those that used the surrogate geometry, reaching in both cases Gamma ($$2\%$$
2
%
, 2 mm) pass rates larger than $$99\%$$
99
%
, except for the 15 MV surrogate. This result supports previous investigations that suggest a change in the material composition of the TrueBeam 15 MV flattening filter. In order to get the said agreement, PRIMO simulations were run using enlarged transport parameters to compensate the discrepancies between the actual and surrogate geometries.
Conclusions
This work sustains the claim that the simulation of the 6, 8 and 10 MV flattening-filtered beams of the TrueBeam linac can be performed using the Clinac 2100 model of PRIMO without significant loss of accuracy.
Springer Science and Business Media LLC
Title: Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Description:
Abstract
Background
Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geometrical description of the linac head.
Since the geometry of Varian TrueBeam machines has not been disclosed, the manufacturer distributes phase-space files of the linac patient-independent part to allow researchers to compute absorbed dose distributions using the Monte Carlo method.
This approach limits the possibility of achieving an arbitrarily small statistical uncertainty.
This work investigates the use of the geometry of the Varian Clinac 2100, which is included in the Monte Carlo system PRIMO, as a surrogate.
Methods
Energy, radial and angular distributions extracted from the TrueBeam phase space files published by the manufacturer and from phase spaces tallied with PRIMO for the Clinac 2100 were compared for the 6, 8, 10 and 15 MV flattened-filtered beams.
Dose distributions in water computed for the two sets of PSFs were compared with the Varian Representative Beam Data (RBD) for square fields with sides ranging from 3 to 30 cm.
Output factors were calculated for square fields with sides ranging from 2 to 40 cm.
Results
Excellent agreement with the RBD was obtained for the simulations that employed the phase spaces distributed by Varian as well as for those that used the surrogate geometry, reaching in both cases Gamma ($$2\%$$
2
%
, 2 mm) pass rates larger than $$99\%$$
99
%
, except for the 15 MV surrogate.
This result supports previous investigations that suggest a change in the material composition of the TrueBeam 15 MV flattening filter.
In order to get the said agreement, PRIMO simulations were run using enlarged transport parameters to compensate the discrepancies between the actual and surrogate geometries.
Conclusions
This work sustains the claim that the simulation of the 6, 8 and 10 MV flattening-filtered beams of the TrueBeam linac can be performed using the Clinac 2100 model of PRIMO without significant loss of accuracy.
Related Results
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Abstract
Background: Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geometrical description of th...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Evaluation of the Varian TrueBeam™ 6 MV phase-space files for the Monte Carlo simulation in small field dosimetry
Evaluation of the Varian TrueBeam™ 6 MV phase-space files for the Monte Carlo simulation in small field dosimetry
Background: The Monte Carlo (MC) simulation is an effective tool for determining the absorbed dose in small field sizes. To calculate accurate results, the MC simulation requires p...
Karakteristik Covid-19 Varian Delta dan Varian Omicron pada Anak
Karakteristik Covid-19 Varian Delta dan Varian Omicron pada Anak
Abstract: To date, COVID-19 has mutated into several types or variants. The variants of this virus, Delta and Omicron, become noticable and spread in all ages including children wo...
Monte-Carlo Simulation mit Risk Kit (Monte-Carlo Simulation with Risk Kit)
Monte-Carlo Simulation mit Risk Kit (Monte-Carlo Simulation with Risk Kit)
<b>German Abstract:</b> Monte-Carlo Simulationen spielen eine immer bedeutender werdende Rolle der Finanzwirtschaft, den Sozialwissenschaften und im Risk Management. Mo...
Penyebaran sistem vokal DMP di Yaha
Penyebaran sistem vokal DMP di Yaha
Kajian ini dilakukan bertujuan melihat ciri linguistik bagi setiap varian yang wujud di Yaha. Bentuk vokal dilihat dari segi kesejajaran dan penyebarannya dalam ruang geografi, sam...