Transprosthetic Fenestration With Electrified Wires. Experimental Evaluation of Three Multifilament Endografts

View through CrossRef
Purpose: In situ fenestration of aortic endografts is an alternative endovascular technique for treatment of complex aortic aneurysms. While this technique has been carried out also to pass stent-grafts in individual cases, its feasibility and safety using different stent-grafts needs to be evaluated. Methods: In a saline bath at water temperature of 37°C, a 0.018” Astato 30 guidewire was advanced through 3 different stent-grafts (RelayPro, Zenith and Endurant II) by applying external current of 180 W via an electrosurgery pencil. Puncture efficacy and quality of the fenestration after ballooning with a 6 mm percutaneous transluminal angioplasty (PTA) catheter were assessed. Then, balloon-expandable covered stents were deployed in the fenestrations and evaluated for stenosis, using microscopy and radiography. Results: Crossing of the electrified guidewire was instantaneous in the Zenith (n:10) and RelayPro (n:10) groups but not in 3 of 10 punctures in the Endurant group (p<.05). The fenestration area created after PTA was significantly larger in the RelayPro (5.3 mm2 ± 1.8, interquartile range [IQR] 1.6) and Zenith group (6.7 mm2 ± 0.7, IQR 0.5) compared to Endurant (2.3 mm2 ± 0.4, IQR 0.5, p<.001). Fraying was observed in all groups while graft shredding was found in 8 cases after PTA of the Zenith and Endurant endografts and in 5 of the RelayPro group, but the difference was not significant. Vertical tearing was detected after RelayPro (2 out of 10) and Zenith (6 out of 10) fenestrations, no damage was found in the Endurant group (p<.01). Residual stenosis at the level of the fenestration after implantation of a 6 × 79 mm VBX stent had to be corrected in all Endurant cases with a high-pressure PTA catheter. No stenosis was found in the RelayPro and Zenith groups before and after flaring. Conclusions: The “electrified wire” technique is a feasible tool that can be used to perform in situ fenestration by perforation of the endograft fabric. Based on this experimental evaluation the “ideal graft” for this technique could not be identified. Long-term fatigue tests and comparison with other fenestration techniques are required. Clinical Impact In situ endograft fenestration can be a useful technique in emergent aortic repair. Recently, the electrified wire technique has been proposed as alternative option to laser, radiofrequency and needle-based techniques. In comparison to these methods, the use of electrified wires can be performed without modifications of routine equipment. Additionally, the material costs can be substantially reduced. However, the effectiveness of this approach for fenestration of different prosthetic grafts is unknown. Based on our experimental studies, the electrified wire technique is feasible but the Endurant endograft requires more attempts, and the placement of a bridging stent should be completed with high-pressure balloons.
Title: Transprosthetic Fenestration With Electrified Wires. Experimental Evaluation of Three Multifilament Endografts
Description:
Purpose: In situ fenestration of aortic endografts is an alternative endovascular technique for treatment of complex aortic aneurysms.
While this technique has been carried out also to pass stent-grafts in individual cases, its feasibility and safety using different stent-grafts needs to be evaluated.
Methods: In a saline bath at water temperature of 37°C, a 0.
018” Astato 30 guidewire was advanced through 3 different stent-grafts (RelayPro, Zenith and Endurant II) by applying external current of 180 W via an electrosurgery pencil.
Puncture efficacy and quality of the fenestration after ballooning with a 6 mm percutaneous transluminal angioplasty (PTA) catheter were assessed.
Then, balloon-expandable covered stents were deployed in the fenestrations and evaluated for stenosis, using microscopy and radiography.
Results: Crossing of the electrified guidewire was instantaneous in the Zenith (n:10) and RelayPro (n:10) groups but not in 3 of 10 punctures in the Endurant group (p<.
05).
The fenestration area created after PTA was significantly larger in the RelayPro (5.
3 mm2 ± 1.
8, interquartile range [IQR] 1.
6) and Zenith group (6.
7 mm2 ± 0.
7, IQR 0.
5) compared to Endurant (2.
3 mm2 ± 0.
4, IQR 0.
5, p<.
001).
Fraying was observed in all groups while graft shredding was found in 8 cases after PTA of the Zenith and Endurant endografts and in 5 of the RelayPro group, but the difference was not significant.
Vertical tearing was detected after RelayPro (2 out of 10) and Zenith (6 out of 10) fenestrations, no damage was found in the Endurant group (p<.
01).
Residual stenosis at the level of the fenestration after implantation of a 6 × 79 mm VBX stent had to be corrected in all Endurant cases with a high-pressure PTA catheter.
No stenosis was found in the RelayPro and Zenith groups before and after flaring.
Conclusions: The “electrified wire” technique is a feasible tool that can be used to perform in situ fenestration by perforation of the endograft fabric.
Based on this experimental evaluation the “ideal graft” for this technique could not be identified.
Long-term fatigue tests and comparison with other fenestration techniques are required.
Clinical Impact In situ endograft fenestration can be a useful technique in emergent aortic repair.
Recently, the electrified wire technique has been proposed as alternative option to laser, radiofrequency and needle-based techniques.
In comparison to these methods, the use of electrified wires can be performed without modifications of routine equipment.
Additionally, the material costs can be substantially reduced.
However, the effectiveness of this approach for fenestration of different prosthetic grafts is unknown.
Based on our experimental studies, the electrified wire technique is feasible but the Endurant endograft requires more attempts, and the placement of a bridging stent should be completed with high-pressure balloons.

Related Results

Comparison of supra-arch in situ fenestration and chimney techniques for aortic dissection involving the left subclavian artery
Comparison of supra-arch in situ fenestration and chimney techniques for aortic dissection involving the left subclavian artery
Introduction Endovascular intervention involving the aortic arch, particularly in thoracic aortic dissection, remains challenging and controversial at current time when there is no...
Fenestration Condensation Resistance: Computer Simulation and In Situ Performance
Fenestration Condensation Resistance: Computer Simulation and In Situ Performance
Abstract The purpose of this paper is to examine condensation rating procedures, specifically the procedures as defined by the National Fenestration Rating Council (...
Damping Performance of Glass Fiber Reinforced Polymers With Embedded Shape Memory Alloy Wires
Damping Performance of Glass Fiber Reinforced Polymers With Embedded Shape Memory Alloy Wires
Abstract This paper explores the integration of shape memory alloy (SMA) wires within fiber reinforced polymer (FRP) composites to enhance their dynamic properties. ...
Comparing the Safety and Effectiveness of Dedicated Radiofrequency Transseptal Wires to Electrified Metal Guidewires
Comparing the Safety and Effectiveness of Dedicated Radiofrequency Transseptal Wires to Electrified Metal Guidewires
Application of electrocautery to a metal guidewire can be used to perform transseptal puncture (TSP). Dedicated radiofrequency guidewires (RF) may represent a better alternative. T...
A HIGH–PRESSURE MYSTERY: UNVEILING THE BODY‘S HIDDEN SHORTCUT
A HIGH–PRESSURE MYSTERY: UNVEILING THE BODY‘S HIDDEN SHORTCUT
Abstract An arteriovenous fistula (AVF) is an abnormal connection between an artery and a vein, bypassing the capillary network. Acquired AVFs are most commonly caus...
Long-Term Experimental Evaluation of In Situ Laser-Fenestrated Versus Custom-Made Fenestrated Endografts
Long-Term Experimental Evaluation of In Situ Laser-Fenestrated Versus Custom-Made Fenestrated Endografts
Objective: This study evaluates the biomechanical performance and long-term durability of in situ laser fenestration (ISLF) compared to custom-made fenestrated ...
Transcatheter Coil Occlusion of Surgical Fenestration After Fontan Operation
Transcatheter Coil Occlusion of Surgical Fenestration After Fontan Operation
Background Fenestration of the Fontan circulation that results in a residual right-to-left shunt has improved operative survival rates among high-risk patients. Late cl...

Back to Top