Novel approach for quantification of porosity for biomaterial implants using microcomputed tomography (μCT)
View through CrossRef
AbstractPorous bioceramics have been widely investigated in orthopaedic tissue engineering. Attention has been given to manufacturing of a porous bioceramic that mimics the trabecular bone structure for proper bone regeneration. With the advance of biomedical imaging through microcomputed tomography (μCT), this study attempted to quantify the pore structure of different bioceramics. Two bioceramic blocks (BSC and ChronOS) were synthesized by two methods. The specification claimed the porosity of the bioceramic ranged from 40% to 70%. Six blocks of each bioceramic were evaluated by conventional water immersion method and μCT. The pore size and connectivity were evaluated with standardized protocols. By the water immersion method, the porosity of BSC and ChronOS was 60.4% and 74.7%, respectively. The three‐dimensional results of μCT showed that BSC porosity was 26.2% and ChronOS was 60.0%. The pore connectivity was evaluated to be 2.6 for BSC and 39.7 for ChronOS. ChronOS had functional pores with 200 μm to 400 μm in diameter (87.8% ± 0.5%), which is significantly more than 52.8% ± 11.5% of pores in BSC (p < 0.05). Providing information on the functional pores objectively, the μCT evaluation serves as a good standard for specification of the bioceramic‐related implants. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2005
Title: Novel approach for quantification of porosity for biomaterial implants using microcomputed tomography (μCT)
Description:
AbstractPorous bioceramics have been widely investigated in orthopaedic tissue engineering.
Attention has been given to manufacturing of a porous bioceramic that mimics the trabecular bone structure for proper bone regeneration.
With the advance of biomedical imaging through microcomputed tomography (μCT), this study attempted to quantify the pore structure of different bioceramics.
Two bioceramic blocks (BSC and ChronOS) were synthesized by two methods.
The specification claimed the porosity of the bioceramic ranged from 40% to 70%.
Six blocks of each bioceramic were evaluated by conventional water immersion method and μCT.
The pore size and connectivity were evaluated with standardized protocols.
By the water immersion method, the porosity of BSC and ChronOS was 60.
4% and 74.
7%, respectively.
The three‐dimensional results of μCT showed that BSC porosity was 26.
2% and ChronOS was 60.
0%.
The pore connectivity was evaluated to be 2.
6 for BSC and 39.
7 for ChronOS.
ChronOS had functional pores with 200 μm to 400 μm in diameter (87.
8% ± 0.
5%), which is significantly more than 52.
8% ± 11.
5% of pores in BSC (p < 0.
05).
Providing information on the functional pores objectively, the μCT evaluation serves as a good standard for specification of the bioceramic‐related implants.
© 2005 Wiley Periodicals, Inc.
J Biomed Mater Res Part B: Appl Biomater, 2005.
Related Results
EFFECT OF MINERAL COMPOSITION OF CARBONATE ROCKS ON THEIR PETROPHYSICAL PROPERTIES
EFFECT OF MINERAL COMPOSITION OF CARBONATE ROCKS ON THEIR PETROPHYSICAL PROPERTIES
ABSTRACT. The construction of digital models from a set of x-ray tomography images of high resolution (μCT) acquired in rock samples is an innovative technique in growing use nowad...
Iodine doping of a photocuring, one-component embedding medium for µCT and histology
Iodine doping of a photocuring, one-component embedding medium for µCT and histology
Objective: Evaluation of μCT scans of bone implant complexes often shows a specific problem: if an implant material has a very similar radiopacity as the embedding medium (e.g. met...
Standardization of Microcomputed Tomography for Tracheal Tissue Engineering Analysis
Standardization of Microcomputed Tomography for Tracheal Tissue Engineering Analysis
Tracheal tissue engineering has become an active area of interest among clinical and scientific communities; however, methods to evaluate success of
in vivo
...
Porosity Signature of Lunar Impact Basins
Porosity Signature of Lunar Impact Basins
The upper portion of the lunar highland crust is known to show a wide range of porosities between 3 and 23% (Wieczorek et al., 2013). Impact cratering seems to be the prim...
A New Method of Porosity Determination by D-T Neutron Generator and Dual CLYC Detector
A New Method of Porosity Determination by D-T Neutron Generator and Dual CLYC Detector
Porosity is one of the essential parameters in conventional oil and gas reservoir evaluation, as well as plays an important role in the calculation of formation saturation and rese...
3D Characterization of Materials: a Key Enabler in Material Development
3D Characterization of Materials: a Key Enabler in Material Development
The aim of new material development is often to achieve enhanced and reliable performance, and to provide higher customer flexibility, at a lower cost. Characterization of the stru...
Measuring varve thickness using µCT: a comparison with thin section
Measuring varve thickness using µCT: a comparison with thin section
Abstract. X-ray micro-computed tomography (μCT) scans were performed on four varved sediment cores collected in Grand Lake (Labrador) and previously studied with thin sections. The...
Clinical outcomes of zirconia implants: a systematic review and meta-analysis
Clinical outcomes of zirconia implants: a systematic review and meta-analysis
Abstract
Purpose
To assess the clinical outcomes of zirconia dental implants based on an updated systematic literature review.
...