WE‐G‐BRB‐03: On the Detectability of Jaw Decalibrations with the DAVID System

View through CrossRef
Purpose: To evaluate if the recently described problems with the calibration of jaws at Varian linear accelerators can be detected with the DAVID system. Methods: Recently Varian Clinac field size issues resulted in a widespread medical device correction notice, stating that incorrect jaw positioning due to excessive connector resistances may result in clinically relevant deviations from the intended field size. Since field size checks via the light field, separately for each administered field segment, are not practicable, it might be helpful to know that an independent, permanently active field size surveillance system is routinely used in our department since four years. The DAVID system (PTW‐Freiburg, Germany) is a translucent multiwire transmission ionisation chamber, placed in the linac accessory holder. Each detection wire of the chamber is positioned in theprojection of the midline of the associated MLC leaf pair, resulting in a signal proportional to the leaf pair aperture. High lateral resolution has been achieved by a deconvolution software (Looe et al, PMB 55, 2010, 3981– 3992). MLC decalibrations are detected by comparison between measured and reference signal profiles. Results: The side wings of the DAVID system signal profiles are sensibly influenced by the position of the backup jaw blocks limiting the field size in the direction at right angles to the MLC travel direction. Systematic variation of the backup jaw positions for several field sizes and IMRT plans has revealed high sensitivity of the signal profile side wings with respect to jaw decalibrations. Conclusions: The DAVID system is able to detect, without any delay, suddenly appearing jaw decalibrations down to positioning errors less than 1mm. Use of the DAVID system provides the independent and permanent surveillance of correct field sizes including the position of the backup jaws, making checks of light‐field congruence for each field segment unnecessary.
Title: WE‐G‐BRB‐03: On the Detectability of Jaw Decalibrations with the DAVID System
Description:
Purpose: To evaluate if the recently described problems with the calibration of jaws at Varian linear accelerators can be detected with the DAVID system.
Methods: Recently Varian Clinac field size issues resulted in a widespread medical device correction notice, stating that incorrect jaw positioning due to excessive connector resistances may result in clinically relevant deviations from the intended field size.
Since field size checks via the light field, separately for each administered field segment, are not practicable, it might be helpful to know that an independent, permanently active field size surveillance system is routinely used in our department since four years.
The DAVID system (PTW‐Freiburg, Germany) is a translucent multiwire transmission ionisation chamber, placed in the linac accessory holder.
Each detection wire of the chamber is positioned in theprojection of the midline of the associated MLC leaf pair, resulting in a signal proportional to the leaf pair aperture.
High lateral resolution has been achieved by a deconvolution software (Looe et al, PMB 55, 2010, 3981– 3992).
MLC decalibrations are detected by comparison between measured and reference signal profiles.
Results: The side wings of the DAVID system signal profiles are sensibly influenced by the position of the backup jaw blocks limiting the field size in the direction at right angles to the MLC travel direction.
Systematic variation of the backup jaw positions for several field sizes and IMRT plans has revealed high sensitivity of the signal profile side wings with respect to jaw decalibrations.
Conclusions: The DAVID system is able to detect, without any delay, suddenly appearing jaw decalibrations down to positioning errors less than 1mm.
Use of the DAVID system provides the independent and permanent surveillance of correct field sizes including the position of the backup jaws, making checks of light‐field congruence for each field segment unnecessary.

Related Results

Plastic roles of pericytes in the blood–retinal barrier
Plastic roles of pericytes in the blood–retinal barrier
AbstractThe blood–retinal barrier (BRB) consists of tightly interconnected capillary endothelial cells covered with pericytes and glia, but the role of the pericytes in BRB regulat...
Impact of Jaw Functional Status on Neck Flexor Endurance among Females
Impact of Jaw Functional Status on Neck Flexor Endurance among Females
Background: Opening and closing of the mouth, biting, and chewing are all sensory-motor tasks that require both jaw and neck muscles to be contracted simultaneously. The postural s...
Jaw Kinematics and Tongue Protraction-Retraction during Chewing and Drinking in the Pig
Jaw Kinematics and Tongue Protraction-Retraction during Chewing and Drinking in the Pig
ABSTRACTMastication and drinking are rhythmic and cyclic oral behaviors that require interactions between the tongue, jaw, and a food or liquid bolus, respectively. During masticat...
SU‐FF‐I‐113: Anatomical Power Spectrum and Detectability: An Analytical and Experimental Basis
SU‐FF‐I‐113: Anatomical Power Spectrum and Detectability: An Analytical and Experimental Basis
Purpose: Superposition of anatomical clutter in medical images is known to degrade detectability of underlying structures. This paper investigates background power spectrum (PS) in...
Improving Detectability of Illegal Fishing Activities Across Supply Chains
Improving Detectability of Illegal Fishing Activities Across Supply Chains
Abstract Improving detectability (i.e., enforcers’ capacity to detect illegal fishing activities) is crucial for fisheries management, food security and livelihoods. Identi...
Research on the biaxial compound pendulum jaw crusher based on seven-bar mechanism
Research on the biaxial compound pendulum jaw crusher based on seven-bar mechanism
Traditional compound pendulum jaw crushers have many disadvantages such as low efficiency and uneven broken materials. This paper proposes a new biaxial compound pendulum jaw crush...
Do Jaw Deformities Adversely Affect Largemouth Bass?
Do Jaw Deformities Adversely Affect Largemouth Bass?
Abstract Largemouth Bass Micropterus nigricans anglers have mainly adopted catch-and-release practices with limited direct fishing mortality. However, catch-and-rele...

Back to Top