Calibration Method of Accelerometer Based on Rotation Principle Using Double Turntable Centrifuge
View through CrossRef
For linear accelerometers, calibration with a precision centrifuge is a key technology, and the input acceleration imposed on the accelerometer should be accurately obtained in the calibration. However, there are often errors in the installation of sample that make the calibration inaccurate. To solve installation errors and obtain the input acceleration in the calibration of the accelerometer, a calibration method based on the rotation principle using a double turntable centrifuge is proposed in this work. The key operation is that the sub-turntable is rotated to make the input axis of the accelerometer perpendicular to the direction of the centripetal acceleration vector. Models of installation errors of angle and radius were built. Based on these models, the static radius and input acceleration can be obtained accurately, and the calibration of the scale factor, nonlinearity and asymmetry can be implemented. Using this method, measurements of the MEMS accelerometer with a range of ±30 g were carried out. The results show that the discrepancy of performance obtained from different installation positions was smaller than 100 ppm after calibrating the input acceleration. Moreover, the results using this method were consistent with those using the back-calculation method. These results demonstrate that the effectiveness of our proposed method was confirmed. This method can measure the static radius directly eliminating the installation errors of angle and radius, and it simplifies the accelerometer calibration procedure.
Title: Calibration Method of Accelerometer Based on Rotation Principle Using Double Turntable Centrifuge
Description:
For linear accelerometers, calibration with a precision centrifuge is a key technology, and the input acceleration imposed on the accelerometer should be accurately obtained in the calibration.
However, there are often errors in the installation of sample that make the calibration inaccurate.
To solve installation errors and obtain the input acceleration in the calibration of the accelerometer, a calibration method based on the rotation principle using a double turntable centrifuge is proposed in this work.
The key operation is that the sub-turntable is rotated to make the input axis of the accelerometer perpendicular to the direction of the centripetal acceleration vector.
Models of installation errors of angle and radius were built.
Based on these models, the static radius and input acceleration can be obtained accurately, and the calibration of the scale factor, nonlinearity and asymmetry can be implemented.
Using this method, measurements of the MEMS accelerometer with a range of ±30 g were carried out.
The results show that the discrepancy of performance obtained from different installation positions was smaller than 100 ppm after calibrating the input acceleration.
Moreover, the results using this method were consistent with those using the back-calculation method.
These results demonstrate that the effectiveness of our proposed method was confirmed.
This method can measure the static radius directly eliminating the installation errors of angle and radius, and it simplifies the accelerometer calibration procedure.
Related Results
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
Lectin C gene analysis v1
Lectin C gene analysis v1
Mammalian Tissue Total RNA Purification Protocol by GeneJET RNA Purification Kit (Thermo Scientific, USA) Before starting: • Supplement the required amount of Lysis Buffer with β-...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Protocol for antigen labeling in eukaryotic cells and quantification by flow cytometry v1
Protocol for antigen labeling in eukaryotic cells and quantification by flow cytometry v1
Goal: This document aims to standardize the protocol used for labeling intracellular or extracellular antigens in eukaryotic cells, using antibodies already associated with fluoroc...
Flow cytometric analysis of EBUS-TBNA samples v1
Flow cytometric analysis of EBUS-TBNA samples v1
1) First preparation ・Sample is collected in 5% FBS in 1 ml DMEM high glucose. ・Transfer the sample to a 15 ml centrifuge tube and add PBS containing 10% wash and recovery solution...
The Reaction of Blinded Goldfish to Rotation in A Centrifuge*
The Reaction of Blinded Goldfish to Rotation in A Centrifuge*
ABSTRACT
Blinded goldfish behave in a centrifuge as if they were orienting to absolute rotation by means of a Foucault pendulum. That is, they swim so as to compensa...
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-S...
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Aiming at solving the problems such as long calibration time, low precision, and complex operation in hand-eye calibration of welding robot, an automatic hand-eye calibration algor...