Vendor-independent software shows limited variability in speckle tracking strain measurements on images of different vendors
View through CrossRef
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Vendors use proprietary speckle tracking software algorithms for echocardiographic strain measurements, which results in high inter-vendor variability. Little is known about potential advantages or disadvantages of using vendor-independent software in clinical practice.
Purpose
We therefore investigated the reproducibility, accuracy, and ability to identify scar of strain measurements on images from different vendors by using a vendor-independent software.
Methods
A vendor-independent software (TomTec Image Arena) was used to analyze datasets of 63 patients which were obtained on four ultrasound machines from different vendors (GE, Philips, Siemens, Toshiba). We measured the tracking feasibility, inter-vendor bias, the relative and absolute test-re-test variability of strain measurements and their ability to detect scar. Cardiac magnetic resonance delayed enhancement images were used as the reference standard of scar definition.
Results
Tracking feasibility differed depending on the image source (p < 0.05). Variability of global longitudinal strain (GLS) (Figure 1A) was similar (ANOVA p = 0.124) among the images of different vendors whereas variability of segmental longitudinal strain (SLS) (Figure 1B) showed modest difference (ANOVA- peak systolic strain (PS); p = 0.077, end-systolic strain (ES); p = 0.171, post-systolic strain (PSS); p = 0.020). Relative test-re-test variability of GLS showed no differences (ANOVA p = 0.360). Absolute test-re-test errors of SLS measurements showed modest differences among images of different vendors (ANOVA- PS; p = 0.018, ES; p = 0.001, PSS; p = 0.090). No relevant difference in scar detection capability was observed (Figure 1C).
Conclusions
Vendor independent software leads to low bias among strain measurements on images from different vendors. Likewise, measurement variability and the ability to identify scar becomes similar. Our findings suggest that a vendor independent speckle tracking software could help to overcome inter-vendor bias. To which extend such measurements would be more accurate compared to vendor specific software remains to be determined.
Abstract Figure 1
Oxford University Press (OUP)
Title: Vendor-independent software shows limited variability in speckle tracking strain measurements on images of different vendors
Description:
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Vendors use proprietary speckle tracking software algorithms for echocardiographic strain measurements, which results in high inter-vendor variability.
Little is known about potential advantages or disadvantages of using vendor-independent software in clinical practice.
Purpose
We therefore investigated the reproducibility, accuracy, and ability to identify scar of strain measurements on images from different vendors by using a vendor-independent software.
Methods
A vendor-independent software (TomTec Image Arena) was used to analyze datasets of 63 patients which were obtained on four ultrasound machines from different vendors (GE, Philips, Siemens, Toshiba).
We measured the tracking feasibility, inter-vendor bias, the relative and absolute test-re-test variability of strain measurements and their ability to detect scar.
Cardiac magnetic resonance delayed enhancement images were used as the reference standard of scar definition.
Results
Tracking feasibility differed depending on the image source (p < 0.
05).
Variability of global longitudinal strain (GLS) (Figure 1A) was similar (ANOVA p = 0.
124) among the images of different vendors whereas variability of segmental longitudinal strain (SLS) (Figure 1B) showed modest difference (ANOVA- peak systolic strain (PS); p = 0.
077, end-systolic strain (ES); p = 0.
171, post-systolic strain (PSS); p = 0.
020).
Relative test-re-test variability of GLS showed no differences (ANOVA p = 0.
360).
Absolute test-re-test errors of SLS measurements showed modest differences among images of different vendors (ANOVA- PS; p = 0.
018, ES; p = 0.
001, PSS; p = 0.
090).
No relevant difference in scar detection capability was observed (Figure 1C).
Conclusions
Vendor independent software leads to low bias among strain measurements on images from different vendors.
Likewise, measurement variability and the ability to identify scar becomes similar.
Our findings suggest that a vendor independent speckle tracking software could help to overcome inter-vendor bias.
To which extend such measurements would be more accurate compared to vendor specific software remains to be determined.
Abstract Figure 1.
Related Results
Comparing the economic and social security of franchise street vendors and independent street vendors : case studies of Klong Toei Community in Bangkok, Thailand
Comparing the economic and social security of franchise street vendors and independent street vendors : case studies of Klong Toei Community in Bangkok, Thailand
Informal economy workers suffer from a lack of social protection. However, most of independent street vendors in Thailand work informally with little social protection. Number of f...
Mechanical Parameters Determination of a Polymeric Membrane by Digital Image Correlation
Mechanical Parameters Determination of a Polymeric Membrane by Digital Image Correlation
Digital image correlation (DIC) is a powerful method for full-field strain test of polymeric materials and speckle pattern on the material's surface is a critical factor for the me...
Vendor-Agnostic Reconfiguration of Kubernetes Clusters in Cloud Federations
Vendor-Agnostic Reconfiguration of Kubernetes Clusters in Cloud Federations
Kubernetes (K8s) defines standardized APIs for container-based cluster orchestration so it becomes possible for application managers to deploy their applications in a unified manne...
High-resolution solar image reconstruction based on non-rigid alignment
High-resolution solar image reconstruction based on non-rigid alignment
Abstract
Suppressing the interference of atmospheric turbulence and obtaining observational data with high spatial resolution is an important problem to be solved in ground...
DAMPAK PENATAAN PEDAGANG KAKI LIMA DI KAWASAN GEMBONG KOTA SURABAYA
DAMPAK PENATAAN PEDAGANG KAKI LIMA DI KAWASAN GEMBONG KOTA SURABAYA
Sektor informal pada kota-kota besar seringkali merujuk pada aktivitas perekonomian kecil. Pedagang kaki lima merupakan beberapa pelaku usaha sektor informal yang terkait langsung ...
Speckle Tracking as Predictor of Coronary Artery Disease in Diabetic Female with Acute Chest Pain
Speckle Tracking as Predictor of Coronary Artery Disease in Diabetic Female with Acute Chest Pain
Background: Cardiovascular disease is the most common reason of mortality and morbidity all-over the world and is the major complication of diabetes. Diabetes mellitus (DM) has rea...
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Objective: To compare the left atrial function utilising speckle tracking echocardiography in patients with low-risk and high-risk non-valvular atrial fibrillation.Methods: The des...
Strategi Komunikasi Pengelola Pura dalam Relokasi Pedagang di Kawasan Pura Besakih
Strategi Komunikasi Pengelola Pura dalam Relokasi Pedagang di Kawasan Pura Besakih
The Nangun Sat Kertih Loka Bali program encompasses the arrangement of the Besakih Temple area in the construction of parking facilities, kiosks, sanitation facilities, and other s...