Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs
View through CrossRef
Abstract
This study aims to measure entrance surface doses during routine chest and abdomen x-ray examinations of adult and child patients. Radiation dose measurements were performed using thermoluminescent dosimeters TLD-100s in three major public hospitals in northern Jordan on a total of 100 patients. Wide variations in entrance surface doses were observed within and between hospitals, which might be attributed to significant variations of the selected exposure parameters. For adult patients, the results have shown that the majority of entrance surface dose values from both chest and abdomen examinations were within recommended values of diagnostic reference levels. For child patients, the mean entrance surface dose from chest examinations in three age groups were 0.131 mGy (0–1 y), 0.136 mGy (1–5 y), and 0.191 mGy (5–10 y). These values were considered relatively high compared to the European reference levels and published results in the literature. However, for abdomen examinations, entrance surface dose values were relatively lower than European reference levels. Patient effective doses were estimated using a PCXMC 2.0 Monte Carlo program. The results for both adults and children were found to be relatively lower than the values reported by international publications. Due to the wide variations of entrance surface dose and the higher radiation doses delivered to child patients, this study recommends implementing a quality assurance program in such hospitals to achieve optimization between good image quality and minimum dose according to the as low as reasonably achievable principle. Moreover, the results of this work will provide a useful base for establishing local diagnostic reference levels for chest and abdomen examinations in Jordan.
Ovid Technologies (Wolters Kluwer Health)
Title: Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs
Description:
Abstract
This study aims to measure entrance surface doses during routine chest and abdomen x-ray examinations of adult and child patients.
Radiation dose measurements were performed using thermoluminescent dosimeters TLD-100s in three major public hospitals in northern Jordan on a total of 100 patients.
Wide variations in entrance surface doses were observed within and between hospitals, which might be attributed to significant variations of the selected exposure parameters.
For adult patients, the results have shown that the majority of entrance surface dose values from both chest and abdomen examinations were within recommended values of diagnostic reference levels.
For child patients, the mean entrance surface dose from chest examinations in three age groups were 0.
131 mGy (0–1 y), 0.
136 mGy (1–5 y), and 0.
191 mGy (5–10 y).
These values were considered relatively high compared to the European reference levels and published results in the literature.
However, for abdomen examinations, entrance surface dose values were relatively lower than European reference levels.
Patient effective doses were estimated using a PCXMC 2.
0 Monte Carlo program.
The results for both adults and children were found to be relatively lower than the values reported by international publications.
Due to the wide variations of entrance surface dose and the higher radiation doses delivered to child patients, this study recommends implementing a quality assurance program in such hospitals to achieve optimization between good image quality and minimum dose according to the as low as reasonably achievable principle.
Moreover, the results of this work will provide a useful base for establishing local diagnostic reference levels for chest and abdomen examinations in Jordan.
Related Results
Study of the Response and Calibration Procedures of LiF:Mg, Ti TLD-100 for the Dosimetry of Photon Beam
Study of the Response and Calibration Procedures of LiF:Mg, Ti TLD-100 for the Dosimetry of Photon Beam
To guarantee quality assurance and quality control of the received radiation dose and treatment level, Thermoluminescent Dosimeter (TLD) is frequently employed in personal control ...
Autonomy on Trial
Autonomy on Trial
Photo by CHUTTERSNAP on Unsplash
Abstract
This paper critically examines how US bioethics and health law conceptualize patient autonomy, contrasting the rights-based, individualist...
Suffering of Patients with Neurogenic Thoracic Outlet Syndrome (TOS); The First Qualitative study in TOS
Suffering of Patients with Neurogenic Thoracic Outlet Syndrome (TOS); The First Qualitative study in TOS
Abstract
Background
Diagnosis of neurogenic thoracic outlet syndrome (nTOS) is hindered by symptom overlap with cervical radiculopathy, carpal tunnel syndrome, or psychosomatic dis...
Provocative Tests in Diagnosis of Thoracic Outlet Syndrome: A Narrative Review
Provocative Tests in Diagnosis of Thoracic Outlet Syndrome: A Narrative Review
Abstract
Thoracic outlet syndrome (TOS) is a group of conditions caused by the compression of the neurovascular bundle within the thoracic outlet. It is classified into three main ...
Assessment of patient radiation dose in conventional lumbar spine radiography: A multicenter study in the Souss Massa region, Morocco
Assessment of patient radiation dose in conventional lumbar spine radiography: A multicenter study in the Souss Massa region, Morocco
ABSTRACT
Background
Patient radiation exposure in diagnostic radiology is an important concern for radiation protection and pat...
Studi Paparan Radiasi pada Pekerja Radiasi Cathlab dengan Menggunakan My Dose Mini sebagai Upaya Keselamatan Radiasi di RSUP Adam Malik Medan
Studi Paparan Radiasi pada Pekerja Radiasi Cathlab dengan Menggunakan My Dose Mini sebagai Upaya Keselamatan Radiasi di RSUP Adam Malik Medan
Cathlab radiation workers, when performing interventional procedures, are at high risk of the effects of radiation exposure. The risk of radiation exposure is deterministic and sto...
Radiation Evaluation System for Radiation Releasing Environment of Nuclear Power Plant Based on Domain-Driven Design
Radiation Evaluation System for Radiation Releasing Environment of Nuclear Power Plant Based on Domain-Driven Design
In order to meet the sustainable development demand for energy, developing nuclear power actively has become an important means for the country to improve energy supply pattern and...
Practical guidance on the assessment of radiation risks for diagnostic radiological examinations
Practical guidance on the assessment of radiation risks for diagnostic radiological examinations
Abstract
Patient doses cannot be limited; instead, radiological examinations should be justified and optimised to ensure the necessary diagnostic or therapeutic effe...