SU‐FF‐I‐43: Workload Weighed Transmission Curves of Brazilian Shielding Materials: Experimental Results
View through CrossRef
Purpose: Evaluation of attenuation properties of typical Brazilian shielding materials (Barite Concrete) used in diagnostic rooms from transmission measurements considering the local workload distribution. Method and Materials: Broad beam attenuation properties of three shielding materials were obtained following the method proposed by Archer et. al (Med. Phys. 21(9), 1994). These attenuation properties were measured by using a constant potential X‐ray machine operated in voltages from 60 to 150 kV. The primary and transmitted radiation from different material thicknesses were measured by using a Radcal 10×5–6 ion chamber. Moreover, workload spectra (Simpkin, Med. Phys. 23(4) 1996) of typical diagnostic rooms were obtained by observing 1060 patients. These data were combined following the method presented in NCRP 147. Results: The workload weighed transmission curves were obtained considering the different workload spectra and shielding material manufacturer. Results shows that for a same attenuation factor, for example 0.01, the thicknesses of material required for shielding a conventional radiation room to primary radiation ranges from 10 to 14.5 cm depending on the manufacturer. If the workload spectra are not considered, this range of thicknesses grows to 19 to 39 cm considering the 120 kV transmission curve. Conclusion: The new methods published in NCRP 147 for shielding calculation of diagnostic rooms incorporated a strong dependence on the quantities workload distribution and transmission through the shielding materials. Local evaluations are needed in order to adapt this method to the local workloads and used shielding materials. The correct consideration of the combined workload spectra and attenuation properties allows the optimized determination of the shielding material needed for protecting the environment.
Title: SU‐FF‐I‐43: Workload Weighed Transmission Curves of Brazilian Shielding Materials: Experimental Results
Description:
Purpose: Evaluation of attenuation properties of typical Brazilian shielding materials (Barite Concrete) used in diagnostic rooms from transmission measurements considering the local workload distribution.
Method and Materials: Broad beam attenuation properties of three shielding materials were obtained following the method proposed by Archer et.
al (Med.
Phys.
21(9), 1994).
These attenuation properties were measured by using a constant potential X‐ray machine operated in voltages from 60 to 150 kV.
The primary and transmitted radiation from different material thicknesses were measured by using a Radcal 10×5–6 ion chamber.
Moreover, workload spectra (Simpkin, Med.
Phys.
23(4) 1996) of typical diagnostic rooms were obtained by observing 1060 patients.
These data were combined following the method presented in NCRP 147.
Results: The workload weighed transmission curves were obtained considering the different workload spectra and shielding material manufacturer.
Results shows that for a same attenuation factor, for example 0.
01, the thicknesses of material required for shielding a conventional radiation room to primary radiation ranges from 10 to 14.
5 cm depending on the manufacturer.
If the workload spectra are not considered, this range of thicknesses grows to 19 to 39 cm considering the 120 kV transmission curve.
Conclusion: The new methods published in NCRP 147 for shielding calculation of diagnostic rooms incorporated a strong dependence on the quantities workload distribution and transmission through the shielding materials.
Local evaluations are needed in order to adapt this method to the local workloads and used shielding materials.
The correct consideration of the combined workload spectra and attenuation properties allows the optimized determination of the shielding material needed for protecting the environment.
Related Results
SU‐FF‐I‐44: Workload Weighed Transmission Curves of Brazilian Shielding Materials: Computer Simulations
SU‐FF‐I‐44: Workload Weighed Transmission Curves of Brazilian Shielding Materials: Computer Simulations
Purpose: Evaluation of attenuation properties of typical Brazilian shielding materials (Barite Concrete) used in diagnostic rooms from computer simulations and considering the loca...
Inverse Jacobian and related topics for certain superelliptic curves
Inverse Jacobian and related topics for certain superelliptic curves
Given an elliptic curve E over the complex numbers (CC) given by y^2 = x^3 + ax + b, there exists a lattice L in CC such that the group E(CC) of complex points on E is isomorphic ...
Bismuth Pelvic X‐Ray Shielding Reduces Radiation Dose Exposure in Pediatric Radiography
Bismuth Pelvic X‐Ray Shielding Reduces Radiation Dose Exposure in Pediatric Radiography
Background. Radiation using conventional X‐ray is associated with exposure of radiosensitive organs and typically requires the use of protection. This study is aimed at evaluating ...
WE‐B‐330A‐01: Shielding Design Workshop: PET/CT
WE‐B‐330A‐01: Shielding Design Workshop: PET/CT
The application of the structural shielding design techniques and goals as outlined in AAPM Task Group Report 108: PET and PET/CT Shielding Requirements Medical Physics (Vol. 33., ...
Storage Workload Identification
Storage Workload Identification
Storage workload identification is the task of characterizing a workload in a storage system (more specifically, network storage system—NAS or SAN) and matching it with the previou...
Analysis of the Effect of Working Hours on Mental Workload Using the Subjective Workload Assessment Technique Method
Analysis of the Effect of Working Hours on Mental Workload Using the Subjective Workload Assessment Technique Method
This research contains a study of the effect of working hours on mental workload using the Subjective Workload Assessment Technique (SWAT) method. Workload is a consequence of acti...
Effect of Sodium Ion Addition on Copper Selenide/Chitosan Film Towards Electrical and Shielding Efficiency Improvement
Effect of Sodium Ion Addition on Copper Selenide/Chitosan Film Towards Electrical and Shielding Efficiency Improvement
The operation of electronic devices can be disrupted by unwanted electromagnetic signals, affecting its operation. Deploying electromagnetic shielding is a viable solution to minim...
CO
2
, Fiber, and
UV
Direct Laser Writing of Graphene for Flexible
EMI
Shielding Applications: A Comparative Stu
CO
2
, Fiber, and
UV
Direct Laser Writing of Graphene for Flexible
EMI
Shielding Applications: A Comparative Stu
ABSTRACT
As electronic devices become an inseparable part of modern life, the challenge of electromagnetic interference (EMI) continues to gr...