Javascript must be enabled to continue!
Inter-Finger Variability of SpO2 During Hypoxemia and Step Resaturation
View through CrossRef
Background: Pulse oximetry is a non-invasive method for continuous monitoring of peripheral blood oxygen saturation (SpO2) to estimate arterial oxygen saturation. Previous studies suggested that SpO2 measurements show variability depending on the particular finger that is used for measurement. To date, no study has compared all fingers for SpO2 under hypoxemia and during continuous simultaneous monitoring with randomization of finger sensor placement. Objectives: The aim of this study was to assess the inter-finger variability of SpO2 values during sequential desaturation and step resaturation. Methods: Forty-three out of forty-five healthy participants (age 23.0 ± 1.8 years, BMI 24.0 ± 4.4 kg·m–2) completed the experimental assessment with short-term induced hypoxemia by consecutive inhalation of three prepared gas mixtures with reduced oxygen concentrations (14%, 12%, and 10%). SpO2 was measured continuously with the Masimo Radical-97 (Masimo Corp., Irvine, CA, USA) pulse oximeters. Results: The SpO2 measured on the thumb was lower than all other fingers by 0.6% to 0.7% SpO2, a systematic difference that is less than the clinically accepted accuracy of oximeters. No difference in SpO2 dynamics was found between any of the fingers during step resaturation. Conclusions: A systematic difference in measured SpO2 exists between the thumb and the other fingers during desaturation, which should be considered at least as well as the impact of the performance of a particular oximeter, sensor placement or anatomical variability.
Title: Inter-Finger Variability of SpO2 During Hypoxemia and Step Resaturation
Description:
Background: Pulse oximetry is a non-invasive method for continuous monitoring of peripheral blood oxygen saturation (SpO2) to estimate arterial oxygen saturation.
Previous studies suggested that SpO2 measurements show variability depending on the particular finger that is used for measurement.
To date, no study has compared all fingers for SpO2 under hypoxemia and during continuous simultaneous monitoring with randomization of finger sensor placement.
Objectives: The aim of this study was to assess the inter-finger variability of SpO2 values during sequential desaturation and step resaturation.
Methods: Forty-three out of forty-five healthy participants (age 23.
0 ± 1.
8 years, BMI 24.
0 ± 4.
4 kg·m–2) completed the experimental assessment with short-term induced hypoxemia by consecutive inhalation of three prepared gas mixtures with reduced oxygen concentrations (14%, 12%, and 10%).
SpO2 was measured continuously with the Masimo Radical-97 (Masimo Corp.
, Irvine, CA, USA) pulse oximeters.
Results: The SpO2 measured on the thumb was lower than all other fingers by 0.
6% to 0.
7% SpO2, a systematic difference that is less than the clinically accepted accuracy of oximeters.
No difference in SpO2 dynamics was found between any of the fingers during step resaturation.
Conclusions: A systematic difference in measured SpO2 exists between the thumb and the other fingers during desaturation, which should be considered at least as well as the impact of the performance of a particular oximeter, sensor placement or anatomical variability.
Related Results
Pulse Oximetry Values from 33,080 Participants in the Apple Heart & Movement Study
Pulse Oximetry Values from 33,080 Participants in the Apple Heart & Movement Study
Abstract
Wearable devices that include pulse oximetry (SpO2) sensing afford the opportunity to capture oxygen saturation measurements from large cohorts in unconstrained li...
Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
On average, arterial oxygen saturation measured by pulse oximetry (SpO2) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO2), thereby increasing the r...
Risk factors for intraprocedural hypoxemia in patients with acute cerebral ischemia treated with vascular intervention and its impact on prognosis: A retrospective cohort study
Risk factors for intraprocedural hypoxemia in patients with acute cerebral ischemia treated with vascular intervention and its impact on prognosis: A retrospective cohort study
Abstract:
BACKGROUND:
Acute cerebral infarction (ACI) is one of the most common ischemic cerebrovascular diseases in neurology, with high morbidi...
Limitations of SpO2 / FiO2-ratio for classification and monitoring of acute respiratory distress syndrome—an observational cohort study
Limitations of SpO2 / FiO2-ratio for classification and monitoring of acute respiratory distress syndrome—an observational cohort study
Abstract
Background
The ratio of pulse-oximetric peripheral oxygen saturation to fraction of inspired oxygen (SpO2/FiO2) has been proposed as add...
Evaluation of Leading Smartwatches for the Detection of Hypoxemia: Comparison to Reference Oximeter
Evaluation of Leading Smartwatches for the Detection of Hypoxemia: Comparison to Reference Oximeter
Although smartwatches are not considered medical devices, experimental validation of their accuracy in detecting hypoxemia is necessary due to their potential use in monitoring con...
Closed-Loop Oxygen Control
Closed-Loop Oxygen Control
<p>Guidelines recommend that oxygen should be titrated to achieve a target oxygen saturation (SpO2 ) range in acutely unwell patients, a concept colloquially known as “swimmi...
A Randomized Controlled Trial of Naloxone for Optimization of Hypoxemia in Lung Donors After Brain Death
A Randomized Controlled Trial of Naloxone for Optimization of Hypoxemia in Lung Donors After Brain Death
Background.
Persistent hypoxemia is the principal reason lungs from otherwise eligible brain dead (BD) organ donors are not transplanted. Experimental models and retros...
P4675The utility of additional balloon pulmonary angioplasty for residual hypoxemia in normohemodynamic chronic thromboembolic pulmonary hypertension patients after invasive treatments
P4675The utility of additional balloon pulmonary angioplasty for residual hypoxemia in normohemodynamic chronic thromboembolic pulmonary hypertension patients after invasive treatments
Abstract
Background
Balloon pulmonary angioplasty (BPA) improves hemodynamics, symptoms and exercise capacity in patients with c...

