Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
View through CrossRef
On average, arterial oxygen saturation measured by pulse oximetry (SpO2) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO2), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO2 on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm). The SpO2 readings from the two POxs showed higher values than the invasive SaO2 readings, and the disparity increased with decreasing SaO2. SpO2 measured using the two infrared wavelengths showed better correlation with SaO2 than SpO2 measured using the red and infrared wavelengths. After appropriate calibration, the standard deviation of the individual SpO2−SaO2 differences for the two-infrared POx was smaller (3.6%) than that for the red and infrared POx (6.5%, p < 0.05). The overestimation of SpO2 readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths. The two-infrared POx can reduce the overestimation of SpO2 measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia.
Title: Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
Description:
On average, arterial oxygen saturation measured by pulse oximetry (SpO2) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO2), thereby increasing the risk of occult hypoxemia.
In the current article, measurements of SpO2 on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm).
The SpO2 readings from the two POxs showed higher values than the invasive SaO2 readings, and the disparity increased with decreasing SaO2.
SpO2 measured using the two infrared wavelengths showed better correlation with SaO2 than SpO2 measured using the red and infrared wavelengths.
After appropriate calibration, the standard deviation of the individual SpO2−SaO2 differences for the two-infrared POx was smaller (3.
6%) than that for the red and infrared POx (6.
5%, p < 0.
05).
The overestimation of SpO2 readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths.
The two-infrared POx can reduce the overestimation of SpO2 measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia.
Related Results
Pulse Oximetry in the Evaluation of Peripheral Vascular Disease
Pulse Oximetry in the Evaluation of Peripheral Vascular Disease
The role of pulse oximetry in the evaluation of peripheral vascular disease (PVD) was investigated. In addition, the value of elevating the limb to improve the sensitivity of detec...
Retinal Oximetry
Retinal Oximetry
Abstract.Purpose:Malfunction of retinal blood flow or oxygenation is believed to be involved in various diseases. Among them are retinal vessel occlusions, diabetic retinopathy and...
Clinical use of pulse oximetry in adults
Clinical use of pulse oximetry in adults
The article is devoted to the practical aspects of pulse oximetry. Pulse oximetry is a photoplethysmographic method for determining heart rate and the percentage of oxyhemoglobin i...
Noninvasive ventilation vs. high-flow nasal cannula oxygen for preoxygenation before intubation in patients with obesity: a post hoc analysis of a randomized controlled trial
Noninvasive ventilation vs. high-flow nasal cannula oxygen for preoxygenation before intubation in patients with obesity: a post hoc analysis of a randomized controlled trial
Abstract
Background
Critically ill patients with obesity may have an increased risk of difficult intubation and subsequen...
Optical Fogging is an Effective Method of Reducing Overestimation of Myopia
Optical Fogging is an Effective Method of Reducing Overestimation of Myopia
Abstract
Background
To evaluate the effectiveness of optical fogging in reducing overestimation of myopia using an open-field auto-refractor and partial coherence interfero...
Protocol and statistical analysis plan for the PREOXI trial of preoxygenation with noninvasive ventilation vs oxygen mask
Protocol and statistical analysis plan for the PREOXI trial of preoxygenation with noninvasive ventilation vs oxygen mask
ABSTRACT
Background
Hypoxemia is a common and life-threatening complication during emergency tracheal intubation of critically ...
High Concentration Oxygen and Hypercapnia in Respiratory Disease
High Concentration Oxygen and Hypercapnia in Respiratory Disease
<p>Oxygen-induced elevations in arterial carbon dioxide tension have been demonstrated in patients with chronic obstructive pulmonary disease (COPD), asthma, pneumonia, obesi...
Risk factors for intraprocedural hypoxemia in patients with acute cerebral ischemia treated with vascular intervention and its impact on prognosis: A retrospective cohort study
Risk factors for intraprocedural hypoxemia in patients with acute cerebral ischemia treated with vascular intervention and its impact on prognosis: A retrospective cohort study
Abstract:
BACKGROUND:
Acute cerebral infarction (ACI) is one of the most common ischemic cerebrovascular diseases in neurology, with high morbidi...