Evaluation of the Varian TrueBeam™ 6 MV phase-space files for the Monte Carlo simulation in small field dosimetry

View through CrossRef
Background: The Monte Carlo (MC) simulation is an effective tool for determining the absorbed dose in small field sizes. To calculate accurate results, the MC simulation requires precise geometric and material descriptions of the linear accelerator head. Due to proprietary information issues, the description of the Varian TrueBeam™ linear accelerator (Varian Medical Systems, Palo Alto, CA) head geometry and material information are not available. Instead, the manufacturer provided a phase-space file just above the jaw for each photon energy level. Although several studies have validated the accuracy of this phase-space file, to the best of our knowledge, there are no reported data for a small field size (<2x2 cm2) of 6 MV photon beams. Objectives: The purpose of this study was to evaluate the Varian TrueBeam™ phase-space file of the 6 MV photon beam provided by the manufacturer for the Monte Carlo (MC) simulation in small field dosimetry. Materials and methods: The TrueBeam™ linear accelerator was simulated using an EGSnrc MC code with a Varian phase-space file as the input. The simulation was compared with the measurement using percent depth dose (PDD) and beam profile, and the field output factor (FOF) for the 0.6x0.6, 1x1, 2x2, 3x3, 4x4, 6x6, and 10x10 cm2 field sizes. Results: The agreement between the measurements and simulated PDD data was under 2.2% beyond the buildup region. The distance to agreement (DTA) in the buildup region was within 1.0 mm. The simulation data presented identical profiles with the measurement within 1.0% of the dose difference or 1.2 mm of the DTA. The mean dose difference in the radiation field was ≤1.5% for the ≥1x1 cm2 field size. The largest deviation was observed in the 0.6x0.6 cm2 inline beam profile. The deviation of the penumbra and full width at half maximum (FWHM) between simulation and measurement was <2 mm. The agreement of the simulated and measured FOF was within 1.0%, except for the 0.6x0.6 cm2 field size. Conclusion: Overall, the MC simulation demonstrates data that is consistent with the measurement for the ≥1x1 cm2 field sizes. These data assure that the 6 MV Varian phase-space file can be used as a radiation source for accurate MC dose calculation in a small field. However, a large discrepancy in beam profiles was observed at the 0.6x0.6 cm2 field size due to the different primary source sizes among TruebeamTM machines.
Title: Evaluation of the Varian TrueBeam™ 6 MV phase-space files for the Monte Carlo simulation in small field dosimetry
Description:
Background: The Monte Carlo (MC) simulation is an effective tool for determining the absorbed dose in small field sizes.
To calculate accurate results, the MC simulation requires precise geometric and material descriptions of the linear accelerator head.
Due to proprietary information issues, the description of the Varian TrueBeam™ linear accelerator (Varian Medical Systems, Palo Alto, CA) head geometry and material information are not available.
Instead, the manufacturer provided a phase-space file just above the jaw for each photon energy level.
Although several studies have validated the accuracy of this phase-space file, to the best of our knowledge, there are no reported data for a small field size (<2x2 cm2) of 6 MV photon beams.
Objectives: The purpose of this study was to evaluate the Varian TrueBeam™ phase-space file of the 6 MV photon beam provided by the manufacturer for the Monte Carlo (MC) simulation in small field dosimetry.
Materials and methods: The TrueBeam™ linear accelerator was simulated using an EGSnrc MC code with a Varian phase-space file as the input.
The simulation was compared with the measurement using percent depth dose (PDD) and beam profile, and the field output factor (FOF) for the 0.
6x0.
6, 1x1, 2x2, 3x3, 4x4, 6x6, and 10x10 cm2 field sizes.
Results: The agreement between the measurements and simulated PDD data was under 2.
2% beyond the buildup region.
The distance to agreement (DTA) in the buildup region was within 1.
0 mm.
The simulation data presented identical profiles with the measurement within 1.
0% of the dose difference or 1.
2 mm of the DTA.
The mean dose difference in the radiation field was ≤1.
5% for the ≥1x1 cm2 field size.
The largest deviation was observed in the 0.
6x0.
6 cm2 inline beam profile.
The deviation of the penumbra and full width at half maximum (FWHM) between simulation and measurement was <2 mm.
The agreement of the simulated and measured FOF was within 1.
0%, except for the 0.
6x0.
6 cm2 field size.
Conclusion: Overall, the MC simulation demonstrates data that is consistent with the measurement for the ≥1x1 cm2 field sizes.
These data assure that the 6 MV Varian phase-space file can be used as a radiation source for accurate MC dose calculation in a small field.
However, a large discrepancy in beam profiles was observed at the 0.
6x0.
6 cm2 field size due to the different primary source sizes among TruebeamTM machines.

Related Results

Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Abstract Background: Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geometrical description of th...
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Monte Carlo simulation of the Varian TrueBeam flattened-filtered beams using a surrogate geometry in PRIMO
Abstract Background Monte Carlo simulation of radiation transport for medical linear accelerators (linacs) requires accurate knowledge of the geomet...
Penyebaran sistem vokal DMP di Yaha
Penyebaran sistem vokal DMP di Yaha
Kajian ini dilakukan bertujuan melihat ciri linguistik bagi setiap varian yang wujud di Yaha. Bentuk vokal dilihat dari segi kesejajaran dan penyebarannya dalam ruang geografi, sam...
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
For discretely observed barrier options, there exists no closed solution under the Black-Scholes model. Thus, it is often helpful to use Monte Carlo simulations, which are easily a...
Research on Multi-Group Monte Carlo Calculations Based on Group Constants Generated by RMC
Research on Multi-Group Monte Carlo Calculations Based on Group Constants Generated by RMC
Abstract Nowadays, deterministic two-step or Monte Carlo methods are commonly used in core physics calculations. However, with the development of reactor core design, tradi...
Technologies for retrospective radiation dosimetry
Technologies for retrospective radiation dosimetry
Abstract Radiation dosimetry is an important task for assessing the biological damages created in human being due to ionising radiation exposure. Ionising radiation ...
Probabilistic Field Development in Presence of Uncertainty
Probabilistic Field Development in Presence of Uncertainty
Abstract Field developments are characterized by high levels of uncertainty and dynamic interconnected decisions with a complex value description. Typical decisio...

Back to Top