A Comparison of Operating Room Toric Placement Tools: CALLISTO Eye vs. Wavetec AnalyzOR (CORTCO)
View through CrossRef
Abstract
Background: To evaluate procedure times for two cataract planning systems (ZEISS CALLISTO eye and the Wavetec AnalyzOR) in predicting residual astigmatism (prediction error) and other visual outcomes in patients with corneal astigmatism (maximum allowable up to 3.0D) at postoperative month 1.
Methods
This was a prospective, single center, parallel treatment group, bilateral and unilateral, randomized, 1-month study on patients scheduled to undergo routine, small-incision cataract surgery with a toric intraocular lens implantation. Both groups underwent preop measurements with the IOLMaster 700 (Zeiss, Jena, Germany) and surgery with the LenSx device (Alcon). Lens selection in the CALLISTO eye group was based on Zeiss VERACITY Surgery Planner (a web-based tool) and on the Wavetec AnalyzOR component of the ORA system (a real-time intraoperative aberrometer) for those eyes in the ORA group. All procedure and intraoperative times were measured with a stopwatch. Postoperative visual outcomes were evaluated between 1 and 2 months after surgery.
Results
There were 23 eyes in the CALLISTO group and 28 eyes in the ORA group. The mean surgical time for the CALLISTO group was 28.09±1.72 minutes compared to 34.41±1.52 minutes for the ORA group (P=0.01). Toric lens placement mean time in the CALLISTO group was 2.47±0.34 minutes compared to 3.88±0.29 minutes in the ORA group (P=0.0034). At month 1 postoperatively, the manifest refractive spherical error (MRSE) in the CALLISTO eye group 0.022±0.388 compared to -0.174±0.322 in the ORA group; these were not statistically different. There was a higher percentage (75%) of eyes with an MRSE within 0.25D in the ORA group compared to the CALLISTO eye group (56.5%); at all other levels outcomes were numerically higher in the CALLISTO eye group.
Conclusions
Less surgical time was needed when using the CALLISTO eye than the ORA when performing cataract surgery with toric lens implantation. There were similar visual outcomes between the groups and no statistical differences.
Title: A Comparison of Operating Room Toric Placement Tools: CALLISTO Eye vs. Wavetec AnalyzOR (CORTCO)
Description:
Abstract
Background: To evaluate procedure times for two cataract planning systems (ZEISS CALLISTO eye and the Wavetec AnalyzOR) in predicting residual astigmatism (prediction error) and other visual outcomes in patients with corneal astigmatism (maximum allowable up to 3.
0D) at postoperative month 1.
Methods
This was a prospective, single center, parallel treatment group, bilateral and unilateral, randomized, 1-month study on patients scheduled to undergo routine, small-incision cataract surgery with a toric intraocular lens implantation.
Both groups underwent preop measurements with the IOLMaster 700 (Zeiss, Jena, Germany) and surgery with the LenSx device (Alcon).
Lens selection in the CALLISTO eye group was based on Zeiss VERACITY Surgery Planner (a web-based tool) and on the Wavetec AnalyzOR component of the ORA system (a real-time intraoperative aberrometer) for those eyes in the ORA group.
All procedure and intraoperative times were measured with a stopwatch.
Postoperative visual outcomes were evaluated between 1 and 2 months after surgery.
Results
There were 23 eyes in the CALLISTO group and 28 eyes in the ORA group.
The mean surgical time for the CALLISTO group was 28.
09±1.
72 minutes compared to 34.
41±1.
52 minutes for the ORA group (P=0.
01).
Toric lens placement mean time in the CALLISTO group was 2.
47±0.
34 minutes compared to 3.
88±0.
29 minutes in the ORA group (P=0.
0034).
At month 1 postoperatively, the manifest refractive spherical error (MRSE) in the CALLISTO eye group 0.
022±0.
388 compared to -0.
174±0.
322 in the ORA group; these were not statistically different.
There was a higher percentage (75%) of eyes with an MRSE within 0.
25D in the ORA group compared to the CALLISTO eye group (56.
5%); at all other levels outcomes were numerically higher in the CALLISTO eye group.
Conclusions
Less surgical time was needed when using the CALLISTO eye than the ORA when performing cataract surgery with toric lens implantation.
There were similar visual outcomes between the groups and no statistical differences.
Related Results
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length).
Methods: This single-...
Regional photometric study of Ganymede and Callisto
Regional photometric study of Ganymede and Callisto
AbstractWe carried out a regional study of Ganymede and Callisto’s photometry by estimating the parameters of the Hapke photometric models on a handful of regions of inte...
Existence and non-uniqueness of constant scalar curvature toric Sasaki metrics
Existence and non-uniqueness of constant scalar curvature toric Sasaki metrics
AbstractWe study compatible toric Sasaki metrics with constant scalar curvature on co-oriented compact toric contact manifolds of Reeb type of dimension at least five. These metric...
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Abstract
Purpose: To evaluate the effect of toric intraocular lens implantation in cataract patient with corneal opacity and high astigmatism. Methods: 31 eyes of 31 patien...
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Abstract
Purpose: To evaluate the effect of toric intraocular lens implantation in cataract patient with corneal opacity and high astigmatism.
Methods: 31 eyes of 31 patie...
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Abstract
Purpose: To evaluate the effect of toric intraocular lens implantation in cataract patient with corneal opacity and high astigmatism. Methods: 31 eyes of 31 patien...
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity
Abstract
Background: To evaluate the effect of toric intraocular lens implantation in cataract patient with corneal opacity and high astigmatism. Methods: 31 eyes of 31 pat...
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Toric Intraocular Lens Implantation in Cataract Patients with Corneal Opacity.
Abstract
Purpose: To evaluate the effect of toric intraocular lens implantation in cataract patient with corneal opacity and high astigmatism. Methods: 31 eyes of 31 patien...