Abstract 6025: Optimal Stent Deployment. Five Seconds Is Not Enough

View through CrossRef
Background . The risk of thrombosis which had been well described since the advent of stenting, is currently being re-addressed in the era of drug-eluting stents (DES). Among the predictors of thrombosis risk, there are some factors which may be favourably influenced by the operator. Indeed, the quality of stent deployment is crucial, which has been well demonstrated with IVUS. Inadequate deployment may be associated with numerous parameters. The time factor has never been explored accurately so far. Objectives. The purpose of this study was to analyse the relationship between the duration of stent balloon inflation and the quality of stent deployment as measured by StentBoost ™ in patients with de novo lesions. Methods. StentBoost ™ is a software developed by Philips on the flat panel detector, which averages frames, correcting for motion by using the balloon markers. This creates an enhanced image of the stent with an improved signal-to-noise ratio and allows precise measurements. Pressure of stent deployment was selected according to the reference diameter of the lesion and the diameter/pressure chart of the stent delivery system. A constant pressure was applied for 25 seconds and stent size measured at 5, 15, 25 sec., after deflation and post dilatation when performed. Primary endpoint was optimal stent expansion defined as minimal stent diameter/reference diameter at the lesion site > 80%. Results. Data were collected prospectively in 77 patients aged 67.8± 12.4 yrs, diabetic (38%), acute coronary syndrome (37 %), in whom 109 stents (length 18.7±6.4 mm; diameter 2.96±0.49 mm) were deployed at a mean pressure of 13.5±2.48 atm. Lesion reference diameter was 3.06±0.58 mm. Conclusion. Better stent implantation may be achieved by increasing the duration of stent deployment. This may influence the risk of stent thrombosis.
Title: Abstract 6025: Optimal Stent Deployment. Five Seconds Is Not Enough
Description:
Background .
The risk of thrombosis which had been well described since the advent of stenting, is currently being re-addressed in the era of drug-eluting stents (DES).
Among the predictors of thrombosis risk, there are some factors which may be favourably influenced by the operator.
Indeed, the quality of stent deployment is crucial, which has been well demonstrated with IVUS.
Inadequate deployment may be associated with numerous parameters.
The time factor has never been explored accurately so far.
Objectives.
The purpose of this study was to analyse the relationship between the duration of stent balloon inflation and the quality of stent deployment as measured by StentBoost ™ in patients with de novo lesions.
Methods.
StentBoost ™ is a software developed by Philips on the flat panel detector, which averages frames, correcting for motion by using the balloon markers.
This creates an enhanced image of the stent with an improved signal-to-noise ratio and allows precise measurements.
Pressure of stent deployment was selected according to the reference diameter of the lesion and the diameter/pressure chart of the stent delivery system.
A constant pressure was applied for 25 seconds and stent size measured at 5, 15, 25 sec.
, after deflation and post dilatation when performed.
Primary endpoint was optimal stent expansion defined as minimal stent diameter/reference diameter at the lesion site > 80%.
Results.
Data were collected prospectively in 77 patients aged 67.
8± 12.
4 yrs, diabetic (38%), acute coronary syndrome (37 %), in whom 109 stents (length 18.
7±6.
4 mm; diameter 2.
96±0.
49 mm) were deployed at a mean pressure of 13.
5±2.
48 atm.
Lesion reference diameter was 3.
06±0.
58 mm.
Conclusion.
Better stent implantation may be achieved by increasing the duration of stent deployment.
This may influence the risk of stent thrombosis.

Related Results

Novel functionalized and patterned surfaces for cardiovascular applications
Novel functionalized and patterned surfaces for cardiovascular applications
Nowadays, cardiovascular diseases are mainly treated by implantation of a metallic or polymeric mesh, called stent, which maintains the artery widely open. This technique shows ver...
Longitudinal stent elongation or shortening after deployment in the coronary arteries: which is dominant?
Longitudinal stent elongation or shortening after deployment in the coronary arteries: which is dominant?
Abstract Background Stent manufacturers always record stent shortening data while they do not record stent elongation data. The aim of this study is...
One stent versus two stents for distal LM PCI: insights from the experience of a high volume center
One stent versus two stents for distal LM PCI: insights from the experience of a high volume center
Abstract Introduction Distal left main (LM) PCIremains a challenge. One of the most debated issues is whether to use a single vs...
Fluid and Solid Mechanics in Stented Arteries
Fluid and Solid Mechanics in Stented Arteries
Abstract Vascular stents are being used in increasing numbers to correct arterial flow limiting disorders. Although data exist on the in vivo performance of these de...
Stent diameter and stent-related symptoms, does size matter? A systematic review and meta-analysis
Stent diameter and stent-related symptoms, does size matter? A systematic review and meta-analysis
The ureteral insertion of a silicone tube was first performed in 1967. A validated ureteral stent symptom questionnaire (USSQ) is used for an objective assessment of patient-report...

Back to Top