Lumen imaging in calcified coronary arteries using high-resolution Gd-enhanced color K-edge imaging with spectral photon-counting CT: a phantom study
View through CrossRef
Abstract
Objective
Conventional coronary computed tomography angiography (CCTA) lumen assessment is hampered by the similar attenuation of iodine and calcium. We assessed the lumen for Gd-enhanced CCTA on high-resolution color Gd K-edge imaging using a clinical spectral photon-counting computed tomography (SPCCT) prototype in an anthropomorphic phantom.
Materials and methods
A hollow cylindrical coronary artery phantom (10-mm outer diameter, 5-mm inner diameter) containing five cylindrical calcifications of different densities (75, 100, 200, 400, and 800 mg/cm
3
hydroxyapatite, deemed very low, low, medium, high, and very high, respectively) and equal size was scanned on a clinical SPCCT prototype using a clinical CCTA protocol. The artery model was filled with Gd mixed with saline to achieve 400 HU at 70 keV. The luminal area was compared with the physical area (
i.e
., 19.6 mm
2
), and spectral results were compared to conventional acquisition.
Results
In the absence of calcification, physical lumen size was overestimated by 7% and 16% on color Gd K-edge and conventional images, respectively. In the presence of calcification, only color Gd K-edge images enabled the measurement of the lumen,
i.e
., 22.25 ± 0.43 mm
2
, 22.85 ± 1.77 mm
2
, 14.63 ± 1.17 mm
2
, and 15.37 ± 0.78 for very low to very high calcium density, respectively. Largest underestimation (-26%) and overestimation (16%) were shown for the high- and low-density calcification in comparison to physical size, respectively.
Conclusion
Color Gd K-edge imaging with Gd-enhanced CCTA outperformed conventional imaging in assessing the coronary lumen in both noncalcified and calcified vessels, whilst accuracy depends on the presence and density of calcifications.
Relevance statement
In a phantom study, high-resolution color Gd K-edge imaging targeting Gd enabled accurate visualization of the coronary lumen in both calcified and noncalcified arteries, outperforming conventional CT angiography. In the presence of calcifications, only color Gd K-edge imaging provided objective lumen assessment, with accuracy influenced by calcium extent and density.
Key Points
Color Gd K-edge Gd-enhanced CCTA using SPCCT is feasible in a coronary artery phantom.
Color Gd K-edge imaging improves lumen assessment in calcified vessels.
Color Gd K-edge imaging potentially enhances diagnostic precision and treatment planning for patients with coronary artery disease.
Graphical Abstract
Title: Lumen imaging in calcified coronary arteries using high-resolution Gd-enhanced color K-edge imaging with spectral photon-counting CT: a phantom study
Description:
Abstract
Objective
Conventional coronary computed tomography angiography (CCTA) lumen assessment is hampered by the similar attenuation of iodine and calcium.
We assessed the lumen for Gd-enhanced CCTA on high-resolution color Gd K-edge imaging using a clinical spectral photon-counting computed tomography (SPCCT) prototype in an anthropomorphic phantom.
Materials and methods
A hollow cylindrical coronary artery phantom (10-mm outer diameter, 5-mm inner diameter) containing five cylindrical calcifications of different densities (75, 100, 200, 400, and 800 mg/cm
3
hydroxyapatite, deemed very low, low, medium, high, and very high, respectively) and equal size was scanned on a clinical SPCCT prototype using a clinical CCTA protocol.
The artery model was filled with Gd mixed with saline to achieve 400 HU at 70 keV.
The luminal area was compared with the physical area (
i.
e
.
, 19.
6 mm
2
), and spectral results were compared to conventional acquisition.
Results
In the absence of calcification, physical lumen size was overestimated by 7% and 16% on color Gd K-edge and conventional images, respectively.
In the presence of calcification, only color Gd K-edge images enabled the measurement of the lumen,
i.
e
.
, 22.
25 ± 0.
43 mm
2
, 22.
85 ± 1.
77 mm
2
, 14.
63 ± 1.
17 mm
2
, and 15.
37 ± 0.
78 for very low to very high calcium density, respectively.
Largest underestimation (-26%) and overestimation (16%) were shown for the high- and low-density calcification in comparison to physical size, respectively.
Conclusion
Color Gd K-edge imaging with Gd-enhanced CCTA outperformed conventional imaging in assessing the coronary lumen in both noncalcified and calcified vessels, whilst accuracy depends on the presence and density of calcifications.
Relevance statement
In a phantom study, high-resolution color Gd K-edge imaging targeting Gd enabled accurate visualization of the coronary lumen in both calcified and noncalcified arteries, outperforming conventional CT angiography.
In the presence of calcifications, only color Gd K-edge imaging provided objective lumen assessment, with accuracy influenced by calcium extent and density.
Key Points
Color Gd K-edge Gd-enhanced CCTA using SPCCT is feasible in a coronary artery phantom.
Color Gd K-edge imaging improves lumen assessment in calcified vessels.
Color Gd K-edge imaging potentially enhances diagnostic precision and treatment planning for patients with coronary artery disease.
Graphical Abstract.
Related Results
Crescimento de feijoeiro sob influência de carvão vegetal e esterco bovino
Crescimento de feijoeiro sob influência de carvão vegetal e esterco bovino
<p align="justify"><span style="color: #000000;"><span style="font-family: 'Times New Roman', serif;"><span><span lang="pt-BR">É indiscutível a import...
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
<span style="font-size:11pt"><span style="background:#f9f9f4"><span style="line-height:normal"><span style="font-family:Calibri,sans-serif"><b><spa...
Feasibility of energy-resolved angiography
Feasibility of energy-resolved angiography
<p dir="ltr">Cardiovascular disease (CVD) is the leading cause of death globally, with coronary heart disease being one of the most common types of CVD. Two-dimensional x-ray...
Feasibility of energy-resolved angiography
Feasibility of energy-resolved angiography
<p dir="ltr">Cardiovascular disease (CVD) is the leading cause of death globally, with coronary heart disease being one of the most common types of CVD. Two-dimensional x-ray...
Ritelatur Review: Prosedur Pemeriksaan Coronary Angiogrhpy (Cag) Pada Kasus Acute Trasmular Myocardial Infraction
Ritelatur Review: Prosedur Pemeriksaan Coronary Angiogrhpy (Cag) Pada Kasus Acute Trasmular Myocardial Infraction
Literature review: coronary angiography examination procedure in cases of acute trasmular myocardial infarction. Background: Acute transmural myocardial infarction (ATMI) is a crit...
e0517 Forearm arteries with ultrasound for percutaneous coronary procedures
e0517 Forearm arteries with ultrasound for percutaneous coronary procedures
Background
The radial artery has become a widely used approach for coronary angiography and intervention in patients, and the ulnar artery is another approach for...
Magnetic resonance measurement of coronary blood flow
Magnetic resonance measurement of coronary blood flow
Magnetic resonance (MR) flow measurement in the coronary artery can be achieved with either a breath‐hold acquisition or a respiration‐triggered acquisition. MR measurements of car...
e0543 Clinical and coronary angiography characteristics between young (<45) and old (>60) patients with coronary artery disease
e0543 Clinical and coronary angiography characteristics between young (<45) and old (>60) patients with coronary artery disease
Objective
To study the clinical Clinical and coronary angiography characteristics between young (≤45) and old (>60) patients with coronary artery disease.
...