Heliox Delivery with Noninvasive Positive Pressure ventilation: A Laboratory Study
View through CrossRef
Background
There is clinical interest in the use of heliox (helium-oxygen mixture) during noninvasive positive pressure ventilation (NPPV), but delivery of heliox with ventilators designed for NPPV has not been reported. We studied helium concentration ([He]) when an 80%:20% helium:oxygen mixture (heliox) was used with 5 NPPV ventilators (Knightstar, Quantum, BiPAP S/T-D30, Sullivan, and BiPAP Vision).
Methods
A simulated spontaneous breathing lung model was connected to the ventilators with a circuit incorporating a standard leak. Heliox flows of 0, 5, 10, and 18 L/min and oxygen flows of 0 and 10 L/min were titrated into the system at either a proximal position near the lung model or a distal position near the ventilator (titration method). Because the BiPAP Vision has an oxygen delivery module, it was also studied using heliox connected to the air inlet of an oxygen blender, with the blender outlet connected to the oxygen module of the ventilator (blender method). All ventilators were evaluated in spontaneous/timed mode at inspira- tory/expiratory pressures of 10/5, 15/5, and 20/5 cm H
2
O. After 5 minutes, [He], oxygen concen- tration, and pressure in the lung model were recorded.
Results
Heliox flow, NPPV settings, site of heliox infusion, and type of ventilator significantly (p < 0.05) affected [He]. [He] was > 60% when heliox flow was 18 L/min in some combinations of settings. The BiPAP S/T-D30 and Quantum occasionally functioned erratically. The BiPAP Vision (blender method) ventilator performed er- ratically with heliox unless the exhalation port test was bypassed on startup. The addition of heliox flow had no important effect on inspiratory or expiratory positive airway pressure on those breaths during which the ventilators functioned correctly.
Conclusion
Heliox flow was the most im- portant determinant of [He] when using heliox with NPPV. With heliox there was a potential for ventilator malfunction in some conditions. The clinical implications of these findings remain to be determined.
Title: Heliox Delivery with Noninvasive Positive Pressure ventilation: A Laboratory Study
Description:
Background
There is clinical interest in the use of heliox (helium-oxygen mixture) during noninvasive positive pressure ventilation (NPPV), but delivery of heliox with ventilators designed for NPPV has not been reported.
We studied helium concentration ([He]) when an 80%:20% helium:oxygen mixture (heliox) was used with 5 NPPV ventilators (Knightstar, Quantum, BiPAP S/T-D30, Sullivan, and BiPAP Vision).
Methods
A simulated spontaneous breathing lung model was connected to the ventilators with a circuit incorporating a standard leak.
Heliox flows of 0, 5, 10, and 18 L/min and oxygen flows of 0 and 10 L/min were titrated into the system at either a proximal position near the lung model or a distal position near the ventilator (titration method).
Because the BiPAP Vision has an oxygen delivery module, it was also studied using heliox connected to the air inlet of an oxygen blender, with the blender outlet connected to the oxygen module of the ventilator (blender method).
All ventilators were evaluated in spontaneous/timed mode at inspira- tory/expiratory pressures of 10/5, 15/5, and 20/5 cm H
2
O.
After 5 minutes, [He], oxygen concen- tration, and pressure in the lung model were recorded.
Results
Heliox flow, NPPV settings, site of heliox infusion, and type of ventilator significantly (p < 0.
05) affected [He].
[He] was > 60% when heliox flow was 18 L/min in some combinations of settings.
The BiPAP S/T-D30 and Quantum occasionally functioned erratically.
The BiPAP Vision (blender method) ventilator performed er- ratically with heliox unless the exhalation port test was bypassed on startup.
The addition of heliox flow had no important effect on inspiratory or expiratory positive airway pressure on those breaths during which the ventilators functioned correctly.
Conclusion
Heliox flow was the most im- portant determinant of [He] when using heliox with NPPV.
With heliox there was a potential for ventilator malfunction in some conditions.
The clinical implications of these findings remain to be determined.
Related Results
Noninvasive Ventilation with Heliox for Respiratory Distress Syndrome in Preterm Infant: A Systematic Review and Meta-Analysis
Noninvasive Ventilation with Heliox for Respiratory Distress Syndrome in Preterm Infant: A Systematic Review and Meta-Analysis
Objectives. To assess whether noninvasive ventilation with Heliox reduces the need for endotracheal ventilation and subsequent complications in preterm infants with respiratory dis...
Application of heliox for optimized drug delivery through respiratory tract
Application of heliox for optimized drug delivery through respiratory tract
Understanding the transportation and deposition (TD) of inhaled particles in the upper respiratory tract is crucial for predicting health risks and treating pulmonary diseases. The...
Heliox for inducible laryngeal obstruction (vocal cord dysfunction): A systematic literature review
Heliox for inducible laryngeal obstruction (vocal cord dysfunction): A systematic literature review
ObjectivesTo perform a systematic literature review on the use of Heliox with patients with inducible laryngeal obstruction/vocal cord dysfunction to: i) identify current evidence ...
Combined Ventilation of Two Subjects with a Single Mechanical Ventilator Using a New Medical Device: An In Vitro Study
Combined Ventilation of Two Subjects with a Single Mechanical Ventilator Using a New Medical Device: An In Vitro Study
Introduction. The SARS-CoV-2 pandemic has created a sudden lack of ventilators. DuplicARⓇ is a novel device that allows simultaneous and independent ventilation of two subjects wit...
TU‐H‐202‐02: Applications, Validation, Clinical Endpoints and Opportunities for CT Ventilation
TU‐H‐202‐02: Applications, Validation, Clinical Endpoints and Opportunities for CT Ventilation
CT ventilation imaging is an exciting functional imaging modality that combines four‐dimensional computed tomography (4DCT) and deformable image registration (DIR) to provide exqui...
Invasive versus non-invasive ventilation for acute respiratory failure
Invasive versus non-invasive ventilation for acute respiratory failure
Abstract
Abstract:
Background: This case-control study aimed to assess the efficacy and outcomes of noninvasive ventilation (NIV) compared to invasive positive pressure ven...
The use of Heliox plasma to improve the adhesion of a self-adhesive resin cement to NaOCl treated dentin
The use of Heliox plasma to improve the adhesion of a self-adhesive resin cement to NaOCl treated dentin
Abstract
Objectives To evaluate the influence of Heliox plasma to improve the adhesion of self-adhesive resin cement to sodium hypochlorite (NaOCl) treated dentin, in compa...
Respiratory muscle strength no influence the need for noninvasive ventilation after heart surgery
Respiratory muscle strength no influence the need for noninvasive ventilation after heart surgery
The aim of this study was to investigate the relation between preoperative respiratory muscle strength and indication and performance of noninvasive ventilation in the planned hear...