SU‐E‐T‐766: Treatment Planning Comparison Study On Two Different Multileaf Collimators Delivered with Volumetric Modulated Arc Therapy
View through CrossRef
Purpose:To compare and evaluate the performance of two different multileaf collimators(MLCi2 and Agility) delivery with volumetric modulated arc therapy techniques.Methods:Treatment plans were graded four (Low, Moderate, Moderate‐High and High complexity) accorrding to the complexity. This includes 1 Low complexity(brain metastasis), 2 Moderate complexity(Lung and Liver), 1 Moderate‐High complexity(prostate) and 1 High complexity ( head and neck) cases. Total dose of 60 Gy was given for all the plans. All cases were desigined two VMAT plans, one with MLCi2(group A) and the other with Agility(group B). All plans were done on Elekta VMAT with Monaco treatment planning system. All plans were generated with 6 MV X‐rays for both Plan A and Plan B. Plans were evaluated based on the ability to meet the dose volume histogram, radiation conformity index, estimated radiation delivery time, dose homogeneity index(HI) and monitor units(MU) needed to deliver the prescribed dose.Results:Plans of group B achieved the best HI (HI = 1.05 Vs. 1.06) at the Low complexity cases while plans of group A were slightly better at the high complexity cases (HI = 1.12 Vs. 1.14). Faster VMAT plan delivery with Agility than with MLCi2 as plan complexity increased (Low complexity:52s Vs.52s, Moderate complexity:58s Vs. 55s, Moderate‐High complexity: 171s Vs.152s, High complexity : 326s Vs. 202s ), especially for the most complex paradigms delivered time can be decresed 38%. No Significant changes were observed between the group B and group A plans in terms of the healthy tissue mean dose and MU. Both plans respected the planning objective for all organs at risk.Conclusion:The study concludes that VMAT plans with the novel Agility MLC can significant decrease the delivering time at the high complexity cases, while a slight compromise in the dose homogeneity index should be noted.This work was supported by The Medical Science Foundation of The health department of Hebei Province (No. 20130253)
Title: SU‐E‐T‐766: Treatment Planning Comparison Study On Two Different Multileaf Collimators Delivered with Volumetric Modulated Arc Therapy
Description:
Purpose:To compare and evaluate the performance of two different multileaf collimators(MLCi2 and Agility) delivery with volumetric modulated arc therapy techniques.
Methods:Treatment plans were graded four (Low, Moderate, Moderate‐High and High complexity) accorrding to the complexity.
This includes 1 Low complexity(brain metastasis), 2 Moderate complexity(Lung and Liver), 1 Moderate‐High complexity(prostate) and 1 High complexity ( head and neck) cases.
Total dose of 60 Gy was given for all the plans.
All cases were desigined two VMAT plans, one with MLCi2(group A) and the other with Agility(group B).
All plans were done on Elekta VMAT with Monaco treatment planning system.
All plans were generated with 6 MV X‐rays for both Plan A and Plan B.
Plans were evaluated based on the ability to meet the dose volume histogram, radiation conformity index, estimated radiation delivery time, dose homogeneity index(HI) and monitor units(MU) needed to deliver the prescribed dose.
Results:Plans of group B achieved the best HI (HI = 1.
05 Vs.
1.
06) at the Low complexity cases while plans of group A were slightly better at the high complexity cases (HI = 1.
12 Vs.
1.
14).
Faster VMAT plan delivery with Agility than with MLCi2 as plan complexity increased (Low complexity:52s Vs.
52s, Moderate complexity:58s Vs.
55s, Moderate‐High complexity: 171s Vs.
152s, High complexity : 326s Vs.
202s ), especially for the most complex paradigms delivered time can be decresed 38%.
No Significant changes were observed between the group B and group A plans in terms of the healthy tissue mean dose and MU.
Both plans respected the planning objective for all organs at risk.
Conclusion:The study concludes that VMAT plans with the novel Agility MLC can significant decrease the delivering time at the high complexity cases, while a slight compromise in the dose homogeneity index should be noted.
This work was supported by The Medical Science Foundation of The health department of Hebei Province (No.
20130253).
Related Results
SU‐FF‐T‐384: Reliable Radiation Dosimetry for Acquiring Beam Data for a 6 MV Linac Based Stereotactic Radiosurgery
SU‐FF‐T‐384: Reliable Radiation Dosimetry for Acquiring Beam Data for a 6 MV Linac Based Stereotactic Radiosurgery
Purpose: To evaluate the reliability of the dosimetry systems available at our institution in the measurement of beam data for 6 MV linac based stereotactic radiosurgery. Method an...
Geomorphology and Geochemistry of Back Arc Basins in the Havre Trough, Southwest Pacific
Geomorphology and Geochemistry of Back Arc Basins in the Havre Trough, Southwest Pacific
<p>The Havre Trough back arc system located behind the Kermadec Arc, in the southwest Pacific, is a classic example of an intra-oceanic back arc system. Subduction driven mag...
Back‐arc rifting in the Izu‐Bonin Island Arc: Structural evolution of Hachijo and Aoga Shima Rifts
Back‐arc rifting in the Izu‐Bonin Island Arc: Structural evolution of Hachijo and Aoga Shima Rifts
Abstract
Multi‐ and single‐channel seismic profiles are used to investigate the structural evolution of back‐arc rifting in the intra‐oceanic Izu‐Bonin Arc. Hachijo and Aoga ...
Life cycle of an Archean subduction zone from initiation to arc–polarity reversal: Insights from the Zunhua ophiolitic mélange, North China Craton 
Life cycle of an Archean subduction zone from initiation to arc–polarity reversal: Insights from the Zunhua ophiolitic mélange, North China Craton 
<p>Subduction initiation and arc&#8211;polarity reversal have rarely been recognized in the Archean rock record. We document Neoarchean subduction initiation,...
SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
Purpose:Different parameters Result in different plans. As the developing of volumetric modulated arc therapy for rectal cancer, the purpose of this study was to work out appropria...
Sedimentary, volcanic, and tectonic processes of the central Mariana Arc: Mariana Trough back‐arc basin formation and the West Mariana Ridge
Sedimentary, volcanic, and tectonic processes of the central Mariana Arc: Mariana Trough back‐arc basin formation and the West Mariana Ridge
We present new multichannel seismic profiles and bathymetric data from the central Marianas that image the West Mariana Ridge (WMR) remnant arc, both margins of the Mariana Trough ...
Origin of the Kunlun Mountains by arc‐arc and arc‐continent collisions
Origin of the Kunlun Mountains by arc‐arc and arc‐continent collisions
Abstract The Kunlun Mountains were formed by early Mesozoic arc‐arc and arc‐continent collisions. The Middle Kunlun Are was the outer volcanic arc of the Paleozoic Asiatic continen...
Pregnancy and Challenging Transient Anti-GAD65 Positivity: A Case Report with Literature Review
Pregnancy and Challenging Transient Anti-GAD65 Positivity: A Case Report with Literature Review
Abstract
Introduction
During pregnancy, women may develop blood glucose abnormalities like gestational diabetes mellitus (GDM) or, rarely, type 1 diabetes (T1D), which can lead to ...