Accuracy of the Postoperative Refractive Prediction in Optical Biometry Versus Ultrasound Biometry in Preoperative Intraocular Lens Power Calculation
View through CrossRef
Background: Contact A-scan ultrasonic biometry is the most frequently utilized technique for determining axial length (AXL). Optical biometry is a technique that utilises partial coherence interferometry to determine ocular biometry parameters. The aim of this research was to compare the efficacy in estimating postoperative refraction by measuring intraocular lens (IOL) power by the optical biometry (IOL Master) and contact A-scan ultrasound plus Bausch & Lomb (B&L) keratometry after phacoemulsification surgery.
Methods: This prospective, comparative, interventional and randomized study involved 40 eyes with uneventful phacoemulsification surgery and in the bag IOL of adult patients with cataract. Subjects were allocated into 2 equal groups: group A: implanted the IOL power estimated by the automated method and group B: implanted the IOL power estimated by the manual method. All patients were subjected to complete ophthalmic examination, best-corrected visual acuity (BCVA), uncorrected visual acuity (UCVA), snellen visual acuity, cycloplegics refraction and slit lamp examination.
Results: The mean predicted error of IOLM was -0.108 ± 0.462 (P= 0.311) in group A and was 0.053 ±1.092 (P= 0.830) in group B, and the Predicted error of A scan was -0.661 ± 0.686 (P= <0.001) in group A which is statistically significant and -0.340 ± 0.972 (P= <0.134) in group B. IOL power measured by IOLM in group A was significantly lower than IOL power measured by A-Scan (p =0.004). While the mean IOL power measured by IOLM in group B was significantly lower than IOL power measured by A-Scan (p =0.008).
Conclusions: The IOL Master accurately assesses the axial length of the lens and thus provides a significantly better prediction of IOL power and thus refractive outcome in cataract surgery than US biometry.
Title: Accuracy of the Postoperative Refractive Prediction in Optical Biometry Versus Ultrasound Biometry in Preoperative Intraocular Lens Power Calculation
Description:
Background: Contact A-scan ultrasonic biometry is the most frequently utilized technique for determining axial length (AXL).
Optical biometry is a technique that utilises partial coherence interferometry to determine ocular biometry parameters.
The aim of this research was to compare the efficacy in estimating postoperative refraction by measuring intraocular lens (IOL) power by the optical biometry (IOL Master) and contact A-scan ultrasound plus Bausch & Lomb (B&L) keratometry after phacoemulsification surgery.
Methods: This prospective, comparative, interventional and randomized study involved 40 eyes with uneventful phacoemulsification surgery and in the bag IOL of adult patients with cataract.
Subjects were allocated into 2 equal groups: group A: implanted the IOL power estimated by the automated method and group B: implanted the IOL power estimated by the manual method.
All patients were subjected to complete ophthalmic examination, best-corrected visual acuity (BCVA), uncorrected visual acuity (UCVA), snellen visual acuity, cycloplegics refraction and slit lamp examination.
Results: The mean predicted error of IOLM was -0.
108 ± 0.
462 (P= 0.
311) in group A and was 0.
053 ±1.
092 (P= 0.
830) in group B, and the Predicted error of A scan was -0.
661 ± 0.
686 (P= <0.
001) in group A which is statistically significant and -0.
340 ± 0.
972 (P= <0.
134) in group B.
IOL power measured by IOLM in group A was significantly lower than IOL power measured by A-Scan (p =0.
004).
While the mean IOL power measured by IOLM in group B was significantly lower than IOL power measured by A-Scan (p =0.
008).
Conclusions: The IOL Master accurately assesses the axial length of the lens and thus provides a significantly better prediction of IOL power and thus refractive outcome in cataract surgery than US biometry.
Related Results
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
Evaluation of different techniques of measurement of the axial length in silicone oil-filled eyes
Evaluation of different techniques of measurement of the axial length in silicone oil-filled eyes
Purpose
To evaluate different techniques of axial length measurement in silicone oil-filled eyes and compare the results of each with the original axial length ...
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Background: Accurate biometry is critical for optimal visual outcomes in cataract surgery. Both optical and ultrasound biometry are widely used, but their comparative effectiveness...
Comparison of Visual Outcomes of Extended Depth of Focus Lenses in Patients With and Without Previous Laser Refractive Surgery
Comparison of Visual Outcomes of Extended Depth of Focus Lenses in Patients With and Without Previous Laser Refractive Surgery
PURPOSE:
To compare visual outcomes of eyes with and without previous laser refractive surgery that received an extended depth of focus intraocular lens during cataract...
Prediction and Estimation of Postoperative Refractive Error in Phacoemulsification: Using Ultrasound A-Scan and Intra Ocular Lens Master
Prediction and Estimation of Postoperative Refractive Error in Phacoemulsification: Using Ultrasound A-Scan and Intra Ocular Lens Master
Background: This study aims to predict and estimate the postoperative refractive outcome in participants undergoing phacoemulsification using IOL Master and A-scan biometry.Methods...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background: To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefront...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background : To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefron...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background: To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefront...