SU‐E‐T‐198: Evaluation of Dosimetric Characteristics of MOSFET Dosimeter for the Quality Assurance of Photon and Electron Beams
View through CrossRef
Purpose: To study the basic dosimetric characteristics of MOSFET dosimeter for its suitability on constructing the quality assurance phantom. Methods: For all the measurements, the high sensitivity MOSFET with standard bias settings (3mV/cGy) was used. Solid water phantoms were used for both photon and electron beams. Tests were performed for 60Co beams, 6 and 15MV photon beams and 9MeV electrons. The dose‐linearity of MOSFET was studied up to 500 cGy with small increment at lowerdoses. Results were compared with the farmer chamber for photons and PP chamber for electrons. Reproducibility of the MOSFET was studied for the absorbed dose of 100cGy. Beam symmetry was measured at 5‐cm off‐axis position for 15×15cm2 field and comparisons were made with the farmer chamber and EDR‐2 film for photons and PP chamber and Kodak Xomat‐V for electrons. Also, the beam quality index was derived by measuring the dose at 10 and 20cm depth for photon beams. Measurements were done for five different days and compared with ion chamber measurements. Results: The correlation co‐efficient (R2) for MOSFET and ion chamber linearity curve is 0.9999, 0.9999, 0.9998 and 0.9999 for 6MV, 15MV and 60Co beams and 9MeV electrons. High sensitivity MOSFET showed good linearity at lower doses even with standard bias settings. Reproducibility of the MOSFET is 1.680%, 1.815%, 1.7194% and 1.754% for 6MV, 15MV and 60Co beams and 9MeV electrons. Beam Symmetry measured with MOSFET showed maximum variation up to 4.2% where as ion chamber and EDR‐2 was within 2.7% and 3.1%. MOSFET over‐responded consistently at the off‐axis position. Quality index measured with MOSFET was within 0.701% and 1.355% for 6 and 15MV photons, whereas the ion chamber was within 0.450%. Conclusions: The high sensitivity MOSFET may be used as a dosimeter with proper cross‐calibration for the quality assurance phantom, which is under construction.
Title: SU‐E‐T‐198: Evaluation of Dosimetric Characteristics of MOSFET Dosimeter for the Quality Assurance of Photon and Electron Beams
Description:
Purpose: To study the basic dosimetric characteristics of MOSFET dosimeter for its suitability on constructing the quality assurance phantom.
Methods: For all the measurements, the high sensitivity MOSFET with standard bias settings (3mV/cGy) was used.
Solid water phantoms were used for both photon and electron beams.
Tests were performed for 60Co beams, 6 and 15MV photon beams and 9MeV electrons.
The dose‐linearity of MOSFET was studied up to 500 cGy with small increment at lowerdoses.
Results were compared with the farmer chamber for photons and PP chamber for electrons.
Reproducibility of the MOSFET was studied for the absorbed dose of 100cGy.
Beam symmetry was measured at 5‐cm off‐axis position for 15×15cm2 field and comparisons were made with the farmer chamber and EDR‐2 film for photons and PP chamber and Kodak Xomat‐V for electrons.
Also, the beam quality index was derived by measuring the dose at 10 and 20cm depth for photon beams.
Measurements were done for five different days and compared with ion chamber measurements.
Results: The correlation co‐efficient (R2) for MOSFET and ion chamber linearity curve is 0.
9999, 0.
9999, 0.
9998 and 0.
9999 for 6MV, 15MV and 60Co beams and 9MeV electrons.
High sensitivity MOSFET showed good linearity at lower doses even with standard bias settings.
Reproducibility of the MOSFET is 1.
680%, 1.
815%, 1.
7194% and 1.
754% for 6MV, 15MV and 60Co beams and 9MeV electrons.
Beam Symmetry measured with MOSFET showed maximum variation up to 4.
2% where as ion chamber and EDR‐2 was within 2.
7% and 3.
1%.
MOSFET over‐responded consistently at the off‐axis position.
Quality index measured with MOSFET was within 0.
701% and 1.
355% for 6 and 15MV photons, whereas the ion chamber was within 0.
450%.
Conclusions: The high sensitivity MOSFET may be used as a dosimeter with proper cross‐calibration for the quality assurance phantom, which is under construction.
Related Results
SU‐F‐T‐474: Evaluation of Dose Perturbation, Temperature and Sensitivity Variation With Accumulated Dose of MOSFET Detector
SU‐F‐T‐474: Evaluation of Dose Perturbation, Temperature and Sensitivity Variation With Accumulated Dose of MOSFET Detector
Purpose:The use of mega voltage gamma and x‐ray sources with their skin sparring qualities in radiation therapy has been a boon in relieving patient discomfort and allowing high tu...
METHODS FOR IMPROVING THE DRG-05M DOSIMETER IN BRACHYTHERAPY
METHODS FOR IMPROVING THE DRG-05M DOSIMETER IN BRACHYTHERAPY
Relevance: The article discusses one of the ways to improve the DRG-05M dosimeter in brachytherapy by using new components and technologies in an improved scheme, which includes mo...
METHODS FOR IMPROVING THE DRG-05M DOSIMETER IN BRACHYTHERAPY
METHODS FOR IMPROVING THE DRG-05M DOSIMETER IN BRACHYTHERAPY
Relevance: The article discusses one of the ways to improve the DRG-05M dosimeter in brachytherapy by using new components and
technologies in an improved scheme, which includes m...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...