Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

Suspended Lead Suits and Radiation Exposure in Interventional Echocardiographers

View through CrossRef
Importance Percutaneous left atrial appendage occlusion (LAAO) is frequently performed with an interventional echocardiographer (IE) in close proximity to the patient and radiation source, increasing the risk for occupational radiation exposure. Yet optimal radiation shielding methods for IEs remain unknown. Objective To assess whether using a suspended lead suit reduces IE radiation dose compared with traditional lead apron shielding during LAAO procedures. Design, Setting, and Participants In this blinded cross-sectional study, IE radiation exposure data were prospectively collected during all LAAO procedures conducted for adults at a single quaternary care center from February 21 to August 22, 2023. Head-level dosimeters were worn by IEs who were using suspended lead suits during sequential LAAO procedures conducted at the center. Exposures Use of a suspended lead suit or traditional lead apron radiation protection device. Main Outcomes and Measures The primary outcome was measured personal dose equivalents at head level to IEs per case using real-time radiation dosimeters. Dosimeter-derived physician radiation doses and procedural radiation (dose area product) were compared with a control group of 30 sequential historical LAAO cases performed with IEs wearing traditional lead aprons from July 1, 2016, to January 31, 2018. Results Overall, 125 patients (mean [SD] age, 78 [8] years; 77 [61.6%] male) were included. Among 95 cases in which IEs used suspended lead suits, the median (IQR) radiation dose was 0.0 (0.0-0.3) μSv, which was significantly lower than 30 cases in which IEs wore traditional lead aprons (median [IQR], 10.6 [5.8-24.1] μSv; P  < .001). Radiation doses were undetectable in 60% of IEs (57of 95) using suspended lead suits compared with 0% of IEs (0 of 30) using traditional lead aprons ( P  < .001). Radiation doses of at least 20 μSv were observed in 0 of 95 cases (0%) with suspended lead suits vs 9 of 30 cases (30%) with traditional lead aprons ( P  < .001). Similar associations were noted in exploratory analyses adjusting IE radiation doses for procedural dose area product (mean [IQR], 0.0 [0.0-0.0] μSv/Gy × cm 2 ) for suspended lead suits vs traditional lead aprons (mean [IQR], 0.6 [0.3-1.0] μSv/Gy × cm 2 ; P  < .001). Conclusions and Relevance In this cross-sectional study, the use of suspended lead suits was associated with large reductions in head-level radiation doses to IEs compared with the use of traditional lead aprons during LAAO cases. These findings have important ramifications for the risk of occupational radiation exposure to IEs and potential shielding mechanisms during structural heart interventions.
Title: Suspended Lead Suits and Radiation Exposure in Interventional Echocardiographers
Description:
Importance Percutaneous left atrial appendage occlusion (LAAO) is frequently performed with an interventional echocardiographer (IE) in close proximity to the patient and radiation source, increasing the risk for occupational radiation exposure.
Yet optimal radiation shielding methods for IEs remain unknown.
Objective To assess whether using a suspended lead suit reduces IE radiation dose compared with traditional lead apron shielding during LAAO procedures.
Design, Setting, and Participants In this blinded cross-sectional study, IE radiation exposure data were prospectively collected during all LAAO procedures conducted for adults at a single quaternary care center from February 21 to August 22, 2023.
Head-level dosimeters were worn by IEs who were using suspended lead suits during sequential LAAO procedures conducted at the center.
Exposures Use of a suspended lead suit or traditional lead apron radiation protection device.
Main Outcomes and Measures The primary outcome was measured personal dose equivalents at head level to IEs per case using real-time radiation dosimeters.
Dosimeter-derived physician radiation doses and procedural radiation (dose area product) were compared with a control group of 30 sequential historical LAAO cases performed with IEs wearing traditional lead aprons from July 1, 2016, to January 31, 2018.
Results Overall, 125 patients (mean [SD] age, 78 [8] years; 77 [61.
6%] male) were included.
Among 95 cases in which IEs used suspended lead suits, the median (IQR) radiation dose was 0.
0 (0.
0-0.
3) μSv, which was significantly lower than 30 cases in which IEs wore traditional lead aprons (median [IQR], 10.
6 [5.
8-24.
1] μSv; P  < .
001).
Radiation doses were undetectable in 60% of IEs (57of 95) using suspended lead suits compared with 0% of IEs (0 of 30) using traditional lead aprons ( P  < .
001).
Radiation doses of at least 20 μSv were observed in 0 of 95 cases (0%) with suspended lead suits vs 9 of 30 cases (30%) with traditional lead aprons ( P  < .
001).
Similar associations were noted in exploratory analyses adjusting IE radiation doses for procedural dose area product (mean [IQR], 0.
0 [0.
0-0.
0] μSv/Gy × cm 2 ) for suspended lead suits vs traditional lead aprons (mean [IQR], 0.
6 [0.
3-1.
0] μSv/Gy × cm 2 ; P  < .
001).
Conclusions and Relevance In this cross-sectional study, the use of suspended lead suits was associated with large reductions in head-level radiation doses to IEs compared with the use of traditional lead aprons during LAAO cases.
These findings have important ramifications for the risk of occupational radiation exposure to IEs and potential shielding mechanisms during structural heart interventions.

Related Results

Quantification of Radiation Exposure in Canadian Orthopaedic Surgery Residents
Quantification of Radiation Exposure in Canadian Orthopaedic Surgery Residents
Introduction: Natural radiation exposure in the general population averages 3 milliSieverts (mSv) annually; however, radiation exposure in orthopaedic residents is not ...
Hidden Radiation Exposure: Daily Life vs Medical Imaging
Hidden Radiation Exposure: Daily Life vs Medical Imaging
Radiation is a natural and unavoidable part of human life, present in the environment as well as in medical imaging procedures. While many people fear radiation from diagnostic ima...
Social media in interventional radiology
Social media in interventional radiology
AbstractIntroductionSocial media provides opportunities for Interventional Radiologists to share research, communicate with colleagues, provide health information and advertise the...
Radiation Injury
Radiation Injury
Since the development of x-rays in the late 19th century, radiation has increasingly been used for diagnostic imaging and various industrial purposes. The human health consequences...
Assessing and improving radiation safety in cardiac catheterization: a study from Cairo University Hospital
Assessing and improving radiation safety in cardiac catheterization: a study from Cairo University Hospital
Abstract Background Catheter laboratories are high-radiation exposure environments, especially during X-ray procedures like percutaneous translumina...

Back to Top