A new continuous noninvasive finger cuff device (Vitalstream) for cardiac output that communicates wirelessly via bluetooth or Wi-Fi

View through CrossRef
Abstract Background The new noninvasive Vitalstream (VS) continuous physiological monitor (Caretaker Medical LLC, Charlottesville, Virginia), allows continuous cardiac output by a low pump-inflated, finger cuff that pneumatically couples arterial pulsations via a pressure line to a pressure sensor for detection and analysis. Physiological data are communicated wirelessly to a tablet-based user interface via Bluetooth or Wi-Fi. We evaluated its performance against thermodilution cardiac output in patients undergoing cardiac surgery. Methods We compared the agreement between thermodilution cardiac output to that obtained by the continuous noninvasive system during cardiac surgery pre and post-cardiac bypass. Thermodilution cardiac output was performed routinely when clinically indicated by an iced saline cold injectate system. All comparisons between VS and TD/CCO data were post-processed. In order to match the VS CO readings to the averaged discrete TD bolus data, the averaged CO readings of the ten seconds of VS CO data points prior to a sequence of TD bolus injections was matched. Time alignment was based on the medical record time and the VS time-stamped data points. The accuracy against reference TD measurements was assessed via Bland–Altman analysis of the CO values and standard concordance analysis of the ΔCO values (with a 15% exclusion zone). Results Analysis of the data compared the accuracy of the matched measurement pairs of VS and TD/CCO VS absolute CO values with and without initial calibration to the discrete TD CO values, as well as the trending ability, i.e., ΔCO values of the VS physiological monitor compared to those of the reference. The results were comparable with other non-invasive as well as invasive technologies and Bland-Altman analyses showed high agreement between devices in a diverse patient population. The results are significant regarding the goal of expanding access to effective, wireless and readily implemented fluid management monitoring tools to hospital sections previously not covered because of the limitations of traditional technologies. Conclusion This study demonstrated that the agreement between the VS CO and TD CO was clinically acceptable with a percent error (PE) of 34.5 to 38% with and without external calibration. The threshold for an acceptable agreement between the VS and TD was considered to be below 40% which is below the threshold recommended by others.
Title: A new continuous noninvasive finger cuff device (Vitalstream) for cardiac output that communicates wirelessly via bluetooth or Wi-Fi
Description:
Abstract Background The new noninvasive Vitalstream (VS) continuous physiological monitor (Caretaker Medical LLC, Charlottesville, Virginia), allows continuous cardiac output by a low pump-inflated, finger cuff that pneumatically couples arterial pulsations via a pressure line to a pressure sensor for detection and analysis.
Physiological data are communicated wirelessly to a tablet-based user interface via Bluetooth or Wi-Fi.
We evaluated its performance against thermodilution cardiac output in patients undergoing cardiac surgery.
Methods We compared the agreement between thermodilution cardiac output to that obtained by the continuous noninvasive system during cardiac surgery pre and post-cardiac bypass.
Thermodilution cardiac output was performed routinely when clinically indicated by an iced saline cold injectate system.
All comparisons between VS and TD/CCO data were post-processed.
In order to match the VS CO readings to the averaged discrete TD bolus data, the averaged CO readings of the ten seconds of VS CO data points prior to a sequence of TD bolus injections was matched.
Time alignment was based on the medical record time and the VS time-stamped data points.
The accuracy against reference TD measurements was assessed via Bland–Altman analysis of the CO values and standard concordance analysis of the ΔCO values (with a 15% exclusion zone).
Results Analysis of the data compared the accuracy of the matched measurement pairs of VS and TD/CCO VS absolute CO values with and without initial calibration to the discrete TD CO values, as well as the trending ability, i.
e.
, ΔCO values of the VS physiological monitor compared to those of the reference.
The results were comparable with other non-invasive as well as invasive technologies and Bland-Altman analyses showed high agreement between devices in a diverse patient population.
The results are significant regarding the goal of expanding access to effective, wireless and readily implemented fluid management monitoring tools to hospital sections previously not covered because of the limitations of traditional technologies.
Conclusion This study demonstrated that the agreement between the VS CO and TD CO was clinically acceptable with a percent error (PE) of 34.
5 to 38% with and without external calibration.
The threshold for an acceptable agreement between the VS and TD was considered to be below 40% which is below the threshold recommended by others.

Related Results

Contributions to bluetooth low energy mesh networks
Contributions to bluetooth low energy mesh networks
Bluetooth Low Energy (BLE) has become a popular Internet of Things (IoT) technology. However, it was originally designed to only support the star topology. This PhD thesis investig...
“ENDOTRACHEAL TUBE CUFF PRESSURE MONITORING USING PREFIXED VOLUME AIR VERSUS MANOMETER IN GENERAL ANAESTHESIA”
“ENDOTRACHEAL TUBE CUFF PRESSURE MONITORING USING PREFIXED VOLUME AIR VERSUS MANOMETER IN GENERAL ANAESTHESIA”
Background And Objective: The endotracheal tube (ETT) cuff seals the airway, preventing leakage and aspiration of pharyngeal contents into the trachea during ventilation. The Endot...
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Abstract Thoracic outlet syndrome (TOS) is a complex and often overlooked condition caused by the compression of neurovascular structures as they pass through the thoracic outlet. ...
Cuff Shaving Procedure for Distal Cuff Erosion in Peritoneal Dialysis Patients – Three-Year Follow-up
Cuff Shaving Procedure for Distal Cuff Erosion in Peritoneal Dialysis Patients – Three-Year Follow-up
Abstract Background Erosion of the distal cuff of peritoneal dialysis (PD) catheters can lead to exit site infections, tunnel infections, peritonitis, and often culminates...

Back to Top