Background
Striatal specific binding ratios (SBR) are widely used to support the interpretation of dopamine transporter SPECT scans. Automatic SBR computation often involves using affine transformations to map the individual SPECT images to an anatomical reference space for ROI analysis using predefined standard masks. This does not account for differences in volumetric scaling between brain structures since, by definition, affine transformations preserve volume ratios. However, striatal volume has been reported to scale proportional to (intracranial volume)
0.4
, indicating particularly pronounced “negative” allometric scaling. This study aimed to investigate the impact of disregarding allometric scaling on putamen SBR, and to propose an easy-to-implement method to avoid this issue. 656 [
123
I]FP-CIT SPECT (67.2 ± 11.4y, 44.2% females, 52.1/47.9% with reduced/normal striatal signal according to visual interpretation by an expert reader) were spatially normalized by affine transformation to the anatomical reference space of the Montreal Neurological Institute. Unilateral putamen SBR were estimated using hottest voxels analysis in large putamen masks predefined in the reference space. For conventional hottest voxels analysis, the number of hottest voxels was fixed at 1,250 (equivalent to 10 ml, the mean putamen volume in healthy adults). For allometry correction, the number of hottest voxels was determined separately for each subject as 1,250*DET
(1−0.4)
, where DET is the Jacobian determinant of the affine transformation. To characterize the impact of the allometry correction on diagnostic performance, a data-driven Gaussian mixture model was employed.
Results
Linear regression in the visually normal SPECT revealed a positive correlation between uncorrected putamen SBR and DET (Pearson’s
R
= 0.509, 95%-CI 0.422–0.587). The correlation was significantly (
p
< 0.00005) weaker when allometry-corrected ROI analysis was used (
R
= 0.285, 95%-CI 0.180–0.383). The effect size of the distance between reduced and normal putamen SBR as determined by the Gaussian mixture model was significantly larger with than without allometry correction (3.979 versus 3.335, one-sided
p
< 0.0001). When using the visual expert reading as diagnostic reference standard, the overall accuracy of the dichotomized SBR was significantly improved by allometry correction (from 94.4% to 96.2%, one-sided
p
= 0.026).
Conclusion
The diagnostic performance of semi-quantitative SBR analyses involving affine transformations to an anatomical reference space can be enhanced through the application of the proposed allometry correction.
Springer Science and Business Media LLC
Title: Allometry correction in semi-quantitative analysis of dopamine transporter SPECT
Description:
Abstract
Background
Striatal specific binding ratios (SBR) are widely used to support the interpretation of dopamine transporter SPECT scans.
Automatic SBR computation often involves using affine transformations to map the individual SPECT images to an anatomical reference space for ROI analysis using predefined standard masks.
This does not account for differences in volumetric scaling between brain structures since, by definition, affine transformations preserve volume ratios.
However, striatal volume has been reported to scale proportional to (intracranial volume)
0.
4
, indicating particularly pronounced “negative” allometric scaling.
This study aimed to investigate the impact of disregarding allometric scaling on putamen SBR, and to propose an easy-to-implement method to avoid this issue.
656 [
123
I]FP-CIT SPECT (67.
2 ± 11.
4y, 44.
2% females, 52.
1/47.
9% with reduced/normal striatal signal according to visual interpretation by an expert reader) were spatially normalized by affine transformation to the anatomical reference space of the Montreal Neurological Institute.
Unilateral putamen SBR were estimated using hottest voxels analysis in large putamen masks predefined in the reference space.
For conventional hottest voxels analysis, the number of hottest voxels was fixed at 1,250 (equivalent to 10 ml, the mean putamen volume in healthy adults).
For allometry correction, the number of hottest voxels was determined separately for each subject as 1,250*DET
(1−0.
4)
, where DET is the Jacobian determinant of the affine transformation.
To characterize the impact of the allometry correction on diagnostic performance, a data-driven Gaussian mixture model was employed.
Results
Linear regression in the visually normal SPECT revealed a positive correlation between uncorrected putamen SBR and DET (Pearson’s
R
= 0.
509, 95%-CI 0.
422–0.
587).
The correlation was significantly (
p
< 0.
00005) weaker when allometry-corrected ROI analysis was used (
R
= 0.
285, 95%-CI 0.
180–0.
383).
The effect size of the distance between reduced and normal putamen SBR as determined by the Gaussian mixture model was significantly larger with than without allometry correction (3.
979 versus 3.
335, one-sided
p
< 0.
0001).
When using the visual expert reading as diagnostic reference standard, the overall accuracy of the dichotomized SBR was significantly improved by allometry correction (from 94.
4% to 96.
2%, one-sided
p
= 0.
026).
Conclusion
The diagnostic performance of semi-quantitative SBR analyses involving affine transformations to an anatomical reference space can be enhanced through the application of the proposed allometry correction.
Related Results
Absolute quantification in brain SPECT imaging
Absolute quantification in brain SPECT imaging
Certes malalties neurològiques estan associades amb problemes en els sistemes de neurotransmissió. Una aproximació a l'estudi d'aquests sistemes és la tomografia d'emissió SPECT (S...
99mTc-DMSA renal cortical scanning: a comparison of planar, SPECT, and SPECT/CT imaging for the detection of renal cortical scarring
99mTc-DMSA renal cortical scanning: a comparison of planar, SPECT, and SPECT/CT imaging for the detection of renal cortical scarring
Abstract
Background
The best way to assess renal cortical scarring is planar scintigraphy with 99mTc-dimercaptosuccinic acid (DMSA), while the value...
Exploring the in vivo subthreshold membrane activity of phasic firing in midbrain dopamine neurons
Exploring the in vivo subthreshold membrane activity of phasic firing in midbrain dopamine neurons
Dopamine is a key neurotransmitter that serves several essential functions in daily behaviors such as locomotion, motivation, stimulus coding, and learning. Disrupted dopamine circ...
Evaluation of the sensitivity of SPECT/CT systems for 225Ac
Evaluation of the sensitivity of SPECT/CT systems for 225Ac
Radiopharmaceuticals labelled with alpha-emitting radionuclides, particularly 225Ac, are the most promising and interesting radiopharmaceuticals for the cancer treatment. Evaluatio...
SU‐E‐I‐154: Cardiac SPECT/CT: Comparing Slow‐Rotation and Fast‐Rotation CT for Attenuation Correction in Cardiac Imaging
SU‐E‐I‐154: Cardiac SPECT/CT: Comparing Slow‐Rotation and Fast‐Rotation CT for Attenuation Correction in Cardiac Imaging
Purpose: The hybridisation of SPECT with X‐ray CT is expanding the utility of SPECT. In addition to image fusion, CT enables improved attenuation maps (AM) for SPECT. Two CT design...
Diagnostic Performance of Quantitative Ga-SPECT/CT for Patients with Lower-limb Osteomyelitis
Diagnostic Performance of Quantitative Ga-SPECT/CT for Patients with Lower-limb Osteomyelitis
Abstract
Background
Patients with lower-limb osteomyelitis (LLOM) may experience major adverse events, such as lower-leg amputations or death; therefore, early diagnosis a...
Indeterminate solitary vertebral lesions on planar scintigraphy
Indeterminate solitary vertebral lesions on planar scintigraphy
Summary
Objective: This study aims to evaluate the added value of hybrid SPECT-CT in differential diagnosis of indeterminate solitary vertebral lesion (SVL) on planar sci...
Resumo Distribuição dos Equipamentos PET-CT e SPECT em Minas Gerais Uma Análise da Disponibilidade e Acessibilidade pelo SUS.
Resumo Distribuição dos Equipamentos PET-CT e SPECT em Minas Gerais Uma Análise da Disponibilidade e Acessibilidade pelo SUS.
A Medicina Nuclear utiliza modalidades de diagnóstico por imagem que investigam aspectos funcionais e morfológicos do organismo. Entre as principais tecnologias, destacam-se a Tomo...