Repeatability and Reproducibility of Microdosimetry With a Mini-TEPC
View through CrossRef
Experimental microdosimetry measures the energy deposited in a microscopic sensitive volume (
SV
) by single ionizing particles traversing the
SV
or passing by. The fundamental advantage of experimental microdosimetry over the computational approach is that the first allows to determine distributions of energy deposition when information on the energy and nature of the charged particles at the point of interest is incomplete or fragmentary. This is almost always the case in radiation protection applications, but discrepancies between the modelled and the actual scenarios should be considered also in radiation therapy. Models for physical reality are always imperfect and rely both on basic input data and on assumptions and simplifications that are necessarily implemented. Furthermore, unintended events due to human errors or machine/system failures can be minimized but cannot be completely avoided.
Though in proton radiation therapy (PRT) a constant relative biological effectiveness (RBE) of 1.1 is assumed, there is evidence of an increasing RBE towards the end of the proton penetration depth. Treatment Planning Systems (TPS) that take into account a variable linear energy transfer (LET) or RBE are already available and could be implemented in PRT in the near future. However, while the calculated dose distributions produced by the TPS are routinely verified with ionization chambers as part of the quality assurance program of every radiotherapy center, there is no commercial detector currently available to perform routine verification of the radiation quality, calculated by the TPS through LET or RBE distributions. Verification of calculated LET is required to make sure that a complex robustly optimized plan will be delivered as planned. The scientific community is coming to conclusion that a new domain of Quality Assurance additionally to the physical dose verification is required, and microdosimetry can be the right approach to address that. A first important prerequisite is the repeatability and reproducibility of microdosimetric measurements. This work aims at studying experimentally the repeatability and reproducibility of microdosimetric measurements performed with a miniaturized Tissue Equivalent Proportional Counter (mini-TEPC) in a 62 MeV proton beam. Experiments were carried out within 1 year and without propane gas recharging and by different operators. RBE was also calculated by applying the Loncol’s weighting function
r(y)
to microdosimetric distributions. Demonstration of reproducibility of measured microdosimetric quantities
y
¯
F
,
y
¯
D
and RBE
10
in 62 MeV proton beam makes this TEPC a possible metrological tool for LET verification in proton therapy. Future characterization will be performed in higher energy proton beams.
Frontiers Media SA
Title: Repeatability and Reproducibility of Microdosimetry With a Mini-TEPC
Description:
Experimental microdosimetry measures the energy deposited in a microscopic sensitive volume (
SV
) by single ionizing particles traversing the
SV
or passing by.
The fundamental advantage of experimental microdosimetry over the computational approach is that the first allows to determine distributions of energy deposition when information on the energy and nature of the charged particles at the point of interest is incomplete or fragmentary.
This is almost always the case in radiation protection applications, but discrepancies between the modelled and the actual scenarios should be considered also in radiation therapy.
Models for physical reality are always imperfect and rely both on basic input data and on assumptions and simplifications that are necessarily implemented.
Furthermore, unintended events due to human errors or machine/system failures can be minimized but cannot be completely avoided.
Though in proton radiation therapy (PRT) a constant relative biological effectiveness (RBE) of 1.
1 is assumed, there is evidence of an increasing RBE towards the end of the proton penetration depth.
Treatment Planning Systems (TPS) that take into account a variable linear energy transfer (LET) or RBE are already available and could be implemented in PRT in the near future.
However, while the calculated dose distributions produced by the TPS are routinely verified with ionization chambers as part of the quality assurance program of every radiotherapy center, there is no commercial detector currently available to perform routine verification of the radiation quality, calculated by the TPS through LET or RBE distributions.
Verification of calculated LET is required to make sure that a complex robustly optimized plan will be delivered as planned.
The scientific community is coming to conclusion that a new domain of Quality Assurance additionally to the physical dose verification is required, and microdosimetry can be the right approach to address that.
A first important prerequisite is the repeatability and reproducibility of microdosimetric measurements.
This work aims at studying experimentally the repeatability and reproducibility of microdosimetric measurements performed with a miniaturized Tissue Equivalent Proportional Counter (mini-TEPC) in a 62 MeV proton beam.
Experiments were carried out within 1 year and without propane gas recharging and by different operators.
RBE was also calculated by applying the Loncol’s weighting function
r(y)
to microdosimetric distributions.
Demonstration of reproducibility of measured microdosimetric quantities
y
¯
F
,
y
¯
D
and RBE
10
in 62 MeV proton beam makes this TEPC a possible metrological tool for LET verification in proton therapy.
Future characterization will be performed in higher energy proton beams.
Related Results
STRUKTUR DRAMATIK PERTUNJUKAN DRAMA KLASIK SANGGAR TEATER MINI LAKON DEWA RUCI KAJIAN BENTUK DAN FUNGSI
STRUKTUR DRAMATIK PERTUNJUKAN DRAMA KLASIK SANGGAR TEATER MINI LAKON DEWA RUCI KAJIAN BENTUK DAN FUNGSI
ABSTRAK
Pada dasarnya nilai pendidikan karakter mempunyai tiga bagian yang saling bekaitan, yaitu pengetahuan moral, penghayatan moral dan perilaku moral. Oleh karena...
Repeatability and reproducibility of Orbscan II
Repeatability and reproducibility of Orbscan II
This study aimed to determine the repeatability and reproducibility of Orbscan for anterior and posterior best fit sphere (BFS), simulated keratometry (Sim-K), and central (CCT) an...
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
In this paper, we apply Taguchi Methods in the framework of concurrent engineering design for robot-based manufacturing sys tems to obtain higher accuracy and repeatability for rob...
Understanding reproducibility of bioinformatics workflows
Understanding reproducibility of bioinformatics workflows
Reproducibility is an essential factor in establishing the reliability of results reported in the scientific literature and ultimately in the knowledge that is generated. This know...
Repeatability and reproducibility of Keratograph 5M corneal topography
Repeatability and reproducibility of Keratograph 5M corneal topography
Background
Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido ...
COMPARATIVE ANALYSIS OF MONOCORTICAL AND BICORTICAL METHODS OF INSTALLING MINI-IMPLANTS
COMPARATIVE ANALYSIS OF MONOCORTICAL AND BICORTICAL METHODS OF INSTALLING MINI-IMPLANTS
The aim of the study. Carry out a comparative analysis of the support ability of human jaw bone tissue in monocortical and bicortical installation of a mini-implant of own design O...
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
PurposeThis paper aims to propose a reasonable method to evaluate uncertainty of measurement of industrial robots’ orientation repeatability and solve the non-linear problem existi...
Oxidative stress markers have low repeatability: A meta-analysis and simulation study with implications for measuring physiological condition and fitness
Oxidative stress markers have low repeatability: A meta-analysis and simulation study with implications for measuring physiological condition and fitness
Abstract
Markers of oxidative stress are widely used as indicators of physiological state, but their utility for ecological and evolutionary inference remains uncer...