Comparison of DCT, GAT and ORA values in patients with glaucoma
View through CrossRef
Abstract Purpose: The Ocular Response Analyser (ORA) is a novel device for the measurement of intraocular pressure (IOP), corneal hysteresis (CH) and the corneal resistance factor (CRF). We compared the ORA with Dynamic Contour Tonometry (DCT) and Goldmann Applanation Tonometry (GAT). Methods: We examined 63 eyes of 63 glaucoma patients (primary open angle glaucoma (POAG) n=26, normal tension glaucoma (NTG) n=10, ocular hypertension (OHT) n=15, pseudoexfoliation glaucoma (PEXG) n=12). After assessment of corneal thickness (CCT) by Orbscan, IOP was measured in a randomized order with the ORA, DCT and GAT. For each device mean IOP and for the ORA mean IOP, CH and CRF were calculated and assessed for potential dependency on CCT. Results: Mean IOP in mmHg was 20.6 (±6.1 SD) for the ORA, 18.2 (±4.2 SD) for DCT and 16.5 (±4.4 SD) for GAT. Bland‐Altman‐Analysis showed a good agreement between DCT and GAT, while the ORA showed a tendency to higher values in patients with higher IOP. None of the devices showed a dependency on CCT (GAT: r²=0.12, DCT: r²=0.07, ORA: r²=0.08). There was no correlation between CH and CCT (r²=0.01), but a weak correlation between CRF and CCT (r²=0.2). Patients with NTG had a significantly (p=0.005) thinner CCT (506±55μm) than patients with POAG (567±53μm). As could be expected, OHT‐patients showed higher CCT values (592±41μm). CRF was significantly higher in the OHT (11.6±2.2) than in the POAG group (9.6±2.0) (p=0.006). For CH, we observed no difference between groups. Conclusions: The ORA showed a good agreement with DCT and GAT for “normal” IOP, but we observed higher results compared to the other methods for high IOP‐values. We observed no correlation between CCT and IOP, but a weak positive correlation between CRF and CCT.
Title: Comparison of DCT, GAT and ORA values in patients with glaucoma
Description:
Abstract Purpose: The Ocular Response Analyser (ORA) is a novel device for the measurement of intraocular pressure (IOP), corneal hysteresis (CH) and the corneal resistance factor (CRF).
We compared the ORA with Dynamic Contour Tonometry (DCT) and Goldmann Applanation Tonometry (GAT).
Methods: We examined 63 eyes of 63 glaucoma patients (primary open angle glaucoma (POAG) n=26, normal tension glaucoma (NTG) n=10, ocular hypertension (OHT) n=15, pseudoexfoliation glaucoma (PEXG) n=12).
After assessment of corneal thickness (CCT) by Orbscan, IOP was measured in a randomized order with the ORA, DCT and GAT.
For each device mean IOP and for the ORA mean IOP, CH and CRF were calculated and assessed for potential dependency on CCT.
Results: Mean IOP in mmHg was 20.
6 (±6.
1 SD) for the ORA, 18.
2 (±4.
2 SD) for DCT and 16.
5 (±4.
4 SD) for GAT.
Bland‐Altman‐Analysis showed a good agreement between DCT and GAT, while the ORA showed a tendency to higher values in patients with higher IOP.
None of the devices showed a dependency on CCT (GAT: r²=0.
12, DCT: r²=0.
07, ORA: r²=0.
08).
There was no correlation between CH and CCT (r²=0.
01), but a weak correlation between CRF and CCT (r²=0.
2).
Patients with NTG had a significantly (p=0.
005) thinner CCT (506±55μm) than patients with POAG (567±53μm).
As could be expected, OHT‐patients showed higher CCT values (592±41μm).
CRF was significantly higher in the OHT (11.
6±2.
2) than in the POAG group (9.
6±2.
0) (p=0.
006).
For CH, we observed no difference between groups.
Conclusions: The ORA showed a good agreement with DCT and GAT for “normal” IOP, but we observed higher results compared to the other methods for high IOP‐values.
We observed no correlation between CCT and IOP, but a weak positive correlation between CRF and CCT.
Related Results
Immune response to the p-azobenzenearsonate (ABA)-GAT conjugate. II. Hapten-specific T cells induced with ABA-GAT in GAT responder X nonresponder F1 hybrids are restricted to the nonresponder haplotype.
Immune response to the p-azobenzenearsonate (ABA)-GAT conjugate. II. Hapten-specific T cells induced with ABA-GAT in GAT responder X nonresponder F1 hybrids are restricted to the nonresponder haplotype.
Abstract
Immunization of mice with the ABA-GAT conjugate stimulates GAT-specific T helper cells in GAT-responder animals and ABA-specific helpers in nonresponders. U...
Antibody Responses to GAT by (Responder × Nonresponder)F1 Spleen Cells Stimulated by Parental GAT-Macrophages
Antibody Responses to GAT by (Responder × Nonresponder)F1 Spleen Cells Stimulated by Parental GAT-Macrophages
Abstract
The development of IgG PFC responses specific for the random terpolymer of l-glutamic acid60-l-alanine30-L-tyrosine10 (GAT) by virgin and immune (responder ...
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Abstract
Background
Childhood glaucoma is a major cause of childhood blindness worldwide. The profile of childhood glaucoma has not been well characterized in sub-Saharan ...
Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
<b><i>Objective:</i></b> The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann app...
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Abstract
Purpose: To report the profile of newly referred glaucoma patients to the largest tertiary eye care hospital in Saudi Arabia.
Patients and Methods: Medical record...
Cross-sectional Survey to Determine the Awareness and Uptake of Glaucoma Surgery
Cross-sectional Survey to Determine the Awareness and Uptake of Glaucoma Surgery
Introduction
: This study aimed to determine the awareness and uptake of surgery as a treatment option in patients with glaucoma on medical treatment.
...
Undiagnosed Glaucoma and Glaucoma Awareness at a Rural Community in Southeast Nigeria
Undiagnosed Glaucoma and Glaucoma Awareness at a Rural Community in Southeast Nigeria
A
bstract
Background:
Undiagnosed glaucoma is a risk for visual impairment in reso...
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
AIM: To evaluate the trend of glaucoma internal filtration surgeries for inpatients between 2015 and 2021 at the Eye Hospital of Wenzhou Medical University.
METHODS: A review of th...