Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
View through CrossRef
<b><i>Objective:</i></b> The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic. <b><i>Subjects and Methods:</i></b> A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria. A pretested structured questionnaire was used to collect sociodemographic data. The Tono-Pen and the GAT were used to measure the IOP in each patient. Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient. The Tono-Pen readings were compared with the uncorrected and corrected GAT readings. Descriptive and comparative analyses were performed. Values for p < 0.05 were considered statistically significant. <b><i>Results:</i></b> The mean age of the participants was 60.39 ± 16.71 years. The mean IOP using the Tono-Pen was 21.1 ± 6.8 mm Hg versus the mean uncorrected GAT value of 17.1 ± 6.9 mm Hg and the corrected GAT value of 18.9 ± 7.5 mm Hg. The mean CCT was 510.5 ± 29.6 µm. The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.9 ± 2.6 and 2.1 ± 3.5 mm Hg, respectively. Gender (Tono-Pen vs. GAT; p = 0.981 vs. 0.437) and corneal thickness (p = 0.057) did not significantly affect the IOP value. Of the 75 patients, 68 (90.7%) preferred the Tono-Pen to the GAT. <b><i>Conclusion:</i></b> In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values. Gender and corneal thickness did not significantly affect the measurements. Most patients found the Tono-Pen more acceptable than the GAT.
Title: Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
Description:
<b><i>Objective:</i></b> The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic.
<b><i>Subjects and Methods:</i></b> A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria.
A pretested structured questionnaire was used to collect sociodemographic data.
The Tono-Pen and the GAT were used to measure the IOP in each patient.
Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient.
The Tono-Pen readings were compared with the uncorrected and corrected GAT readings.
Descriptive and comparative analyses were performed.
Values for p < 0.
05 were considered statistically significant.
<b><i>Results:</i></b> The mean age of the participants was 60.
39 ± 16.
71 years.
The mean IOP using the Tono-Pen was 21.
1 ± 6.
8 mm Hg versus the mean uncorrected GAT value of 17.
1 ± 6.
9 mm Hg and the corrected GAT value of 18.
9 ± 7.
5 mm Hg.
The mean CCT was 510.
5 ± 29.
6 µm.
The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.
9 ± 2.
6 and 2.
1 ± 3.
5 mm Hg, respectively.
Gender (Tono-Pen vs.
GAT; p = 0.
981 vs.
0.
437) and corneal thickness (p = 0.
057) did not significantly affect the IOP value.
Of the 75 patients, 68 (90.
7%) preferred the Tono-Pen to the GAT.
<b><i>Conclusion:</i></b> In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values.
Gender and corneal thickness did not significantly affect the measurements.
Most patients found the Tono-Pen more acceptable than the GAT.
Related Results
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reli...
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Abstract
Background
Childhood glaucoma is a major cause of childhood blindness worldwide. The profile of childhood glaucoma has not been well characterized in sub-Saharan ...
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
ABSTRACT
BACKGROUND: Measured ocular tension is reported to be positively related to corneal thickness. It has been speculated that an edematous cornea will cause low mea...
The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
BACKGROUND፡ The aim of this study was to determine whether Central Corneal Thickness (CCT) is associated with intraocular pressure measurement (IOP) with a Non-contact tonometer an...
Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre
Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre
Purpose: The purpose of this study was to evaluate the role of air puff (AP) tonometer by comparing the measurements of intraocular pressure (IOP) made using this device with those...
Implantable arterial blood pressure sensor
Implantable arterial blood pressure sensor
Although blood pressure measurements are well developed, the existing technology in the field of Arterial Blood Pressure (ABP) measurements is not sufficient in terms of long-term ...
Cross-sectional Survey to Determine the Awareness and Uptake of Glaucoma Surgery
Cross-sectional Survey to Determine the Awareness and Uptake of Glaucoma Surgery
Introduction
: This study aimed to determine the awareness and uptake of surgery as a treatment option in patients with glaucoma on medical treatment.
...
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Abstract
Purpose: To report the profile of newly referred glaucoma patients to the largest tertiary eye care hospital in Saudi Arabia.
Patients and Methods: Medical record...