phrenic-safe pulsed field ablation: acute and long-term outcomes from a post hoc analysis of the Volt CE Mark and IDE Studies
View through CrossRef
Abstract
Background
Pulsed field ablation (PFA) has emerged as a tissue-selective energy source for atrial fibrillation (AF) ablation due to its preferential effect on myocardial tissue. This selectivity is hypothesized to reduce the risk of phrenic nerve injury (PNI), a known complication of pulmonary vein isolation. However, clinical reports indicate that PNI can still occur [1]. A novel CE-marked balloon-in-basket PFA system utilizing a lower voltage (LV) setting was developed for use in regions where phrenic nerve capture is unavoidable during ablation [2].
Purpose
This post hoc analysis aimed to evaluate the acute and long-term effectiveness and safety of any low voltage (LV) setting use compared to nominal voltage (NV) only use in subjects undergoing PFA for paroxysmal (PAF) or persistent atrial fibrillation (PersAF).
Methods
Subjects were enrolled in the Volt CE Mark and VOLT-AF IDE studies, comprising 150 and 394 subjects, respectively. This analysis included subjects with PAF or PersAF who were categorized into four groups: PAF-LV, PAF-NV, PersAF-LV, and PersAF-NV. Baseline characteristics, acute procedural effectiveness (APE), long-term effectiveness (LTE) at 12 months, and primary safety outcomes were compared between LV and NV groups.
Results
A total of 166 subjects (108 PAF, 58 PersAF) received some ablation with LV, targeting the right inferior pulmonary vein (RIPV) or right superior pulmonary vein (RSPV). The LV cohort was 33.7% female and 66.3% male, with a mean age of 62.2 ± 11.1 years, and mean LAD of 39.2 ± 7.3 mm. Subjects were geographically distributed across the USA (57.2%), Australia (14.5%), and Europe (28.3%).
Mean LV applications per subject were 3.3 ± 1.7 PAF and 3.1 ± 2.1 PersAF (RSPV) and 1.3 ± 1.8 PAF and 1.2 ± 1.7 PersAF (RIPV). APE was achieved in 97.2% of PAF-LV and 99.5% of PAF-NV subjects (p = 0.1312), and in 98.3% of PersAF-LV and 98.7% of PersAF-NV subjects (p = 1.000). At 12-month follow-up, LTE was 81.3% in PAF-LV and 79.9% in PAF-NV (p = 0.7854), and 72.4% in PersAF-LV versus 58.9% in PersAF-NV (p = 0.0642). A primary safety event through 7 days post-procedure occurred in 0.9% of PAF-LV and 1.5% of PAF-NV subjects (p = 1.0000), and in 1.4% of PersAF-LV and 4.4% of PersAF-NV subjects (p = 0.6852). There were no phrenic nerve related primary safety events reported in any group. In a pooled analysis of treated pulmonary veins (PVs) assessed during repeat procedures, reconnection was observed in 3/5 right-sided PVs treated with LV and 10/33 right-sided PVs treated with NV.
Conclusion
Use of LV in areas of unavoidable phrenic nerve capture during PFA demonstrated comparable APE and LTE to use of NV in both PAF and PersAF. Safety outcomes were consistent across all groups. These findings support the use of phrenic-safe LV to mitigate PNI risk without compromising procedural effectiveness.PAF No Significant Difference p=0.7854PersAF No Significant Difference p=0.064
Oxford University Press (OUP)
Title: phrenic-safe pulsed field ablation: acute and long-term outcomes from a post hoc analysis of the Volt CE Mark and IDE Studies
Description:
Abstract
Background
Pulsed field ablation (PFA) has emerged as a tissue-selective energy source for atrial fibrillation (AF) ablation due to its preferential effect on myocardial tissue.
This selectivity is hypothesized to reduce the risk of phrenic nerve injury (PNI), a known complication of pulmonary vein isolation.
However, clinical reports indicate that PNI can still occur [1].
A novel CE-marked balloon-in-basket PFA system utilizing a lower voltage (LV) setting was developed for use in regions where phrenic nerve capture is unavoidable during ablation [2].
Purpose
This post hoc analysis aimed to evaluate the acute and long-term effectiveness and safety of any low voltage (LV) setting use compared to nominal voltage (NV) only use in subjects undergoing PFA for paroxysmal (PAF) or persistent atrial fibrillation (PersAF).
Methods
Subjects were enrolled in the Volt CE Mark and VOLT-AF IDE studies, comprising 150 and 394 subjects, respectively.
This analysis included subjects with PAF or PersAF who were categorized into four groups: PAF-LV, PAF-NV, PersAF-LV, and PersAF-NV.
Baseline characteristics, acute procedural effectiveness (APE), long-term effectiveness (LTE) at 12 months, and primary safety outcomes were compared between LV and NV groups.
Results
A total of 166 subjects (108 PAF, 58 PersAF) received some ablation with LV, targeting the right inferior pulmonary vein (RIPV) or right superior pulmonary vein (RSPV).
The LV cohort was 33.
7% female and 66.
3% male, with a mean age of 62.
2 ± 11.
1 years, and mean LAD of 39.
2 ± 7.
3 mm.
Subjects were geographically distributed across the USA (57.
2%), Australia (14.
5%), and Europe (28.
3%).
Mean LV applications per subject were 3.
3 ± 1.
7 PAF and 3.
1 ± 2.
1 PersAF (RSPV) and 1.
3 ± 1.
8 PAF and 1.
2 ± 1.
7 PersAF (RIPV).
APE was achieved in 97.
2% of PAF-LV and 99.
5% of PAF-NV subjects (p = 0.
1312), and in 98.
3% of PersAF-LV and 98.
7% of PersAF-NV subjects (p = 1.
000).
At 12-month follow-up, LTE was 81.
3% in PAF-LV and 79.
9% in PAF-NV (p = 0.
7854), and 72.
4% in PersAF-LV versus 58.
9% in PersAF-NV (p = 0.
0642).
A primary safety event through 7 days post-procedure occurred in 0.
9% of PAF-LV and 1.
5% of PAF-NV subjects (p = 1.
0000), and in 1.
4% of PersAF-LV and 4.
4% of PersAF-NV subjects (p = 0.
6852).
There were no phrenic nerve related primary safety events reported in any group.
In a pooled analysis of treated pulmonary veins (PVs) assessed during repeat procedures, reconnection was observed in 3/5 right-sided PVs treated with LV and 10/33 right-sided PVs treated with NV.
Conclusion
Use of LV in areas of unavoidable phrenic nerve capture during PFA demonstrated comparable APE and LTE to use of NV in both PAF and PersAF.
Safety outcomes were consistent across all groups.
These findings support the use of phrenic-safe LV to mitigate PNI risk without compromising procedural effectiveness.
PAF No Significant Difference p=0.
7854PersAF No Significant Difference p=0.
064.
Related Results
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Abstract
Introduction
Thoracic outlet syndrome is a group of disorders arising from compressive forces on the neurovascular bundle in that region due to different etiologies. This...
[RETRACTED] Keanu Reeves CBD Gummies v1
[RETRACTED] Keanu Reeves CBD Gummies v1
[RETRACTED]Keanu Reeves CBD Gummies ==❱❱ Huge Discounts:[HURRY UP ] Absolute Keanu Reeves CBD Gummies (Available)Order Online Only!! ❰❰= https://www.facebook.com/Keanu-Reeves-CBD-G...
Phrenic Nerve Transfer to Musculocutaneous Nerve: An Anatomical and Histological Study
Phrenic Nerve Transfer to Musculocutaneous Nerve: An Anatomical and Histological Study
Background: To restore elbow flexor muscle function in case of traumatic brachial plexus avulsion, the phrenic nerve transfer to the musculocutaneous nerve has become part of clini...
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Abstract
Introduction
Microwave ablation (MWA) has emerged as a minimally invasive treatment for patients with inoperable non-small cell lung cancer (NSCLC). However, whether it i...
Entirely anatomical high-power ablation for refractory left ventricular summit arrhythmias
Entirely anatomical high-power ablation for refractory left ventricular summit arrhythmias
Abstract
Introduction
Ventricular arrhythmias (VA) originating from the left ventricular summit (LVS) can be difficult to...
Influence of pulmonary inflations on discharge patterns of phrenic motoneurons
Influence of pulmonary inflations on discharge patterns of phrenic motoneurons
The purpose of this study was to assess the influence of pulmonary inflations on activities of single phrenic motoneurons. Studies were performed in decerebrate and paralyzed cats;...
XÂY DỰNG VÀ SỬ DỤNG HỌC LIỆU SỐ TRONG DẠY HỌC “SINH HỌC VI SINH VẬT VÀ VIRUS” (SINH HỌC 10)
XÂY DỰNG VÀ SỬ DỤNG HỌC LIỆU SỐ TRONG DẠY HỌC “SINH HỌC VI SINH VẬT VÀ VIRUS” (SINH HỌC 10)
Bài viết đề cập đến những nghiên cứu về tầm quan trọng của học liệu số, các dạng học liệu số trong dạy học Sinh học nói chung và dạy học Sinh học vi sinh vật và virus – Sinh học 10...
Acute safety and procedural characteristics of conscious and deep sedation to general anesthesia workflows with novel balloon-based PFA system
Acute safety and procedural characteristics of conscious and deep sedation to general anesthesia workflows with novel balloon-based PFA system
Abstract
Background
There is an increasing desire to perform cardiac ablation procedures with decreased general anesthesia requi...