Doppler Echocardiographic Assessment of Normally Functioning Starr-Edwards, Carbomedics and Carpentier-Edwards Valves in Aortic Position
View through CrossRef
Doppler echocardiography was performed in 168 normally functioning aortic prostheses to determine acceptable pressure gradients across the commonly used valves and to establish the relationship between valve size and gradients. There were 82 Carbomedics (C), 63 Starr-Edwards (SE), and 23 Carpentier-Edwards (CE) valves. Peak and mean gradients across the prostheses were measured by use of the simplified Bernoulli equation. CarboMedics valve had a lower peak and mean gradient than Starr-Edwards and Carpentier-Edwards valve (P < 0.05 when compared with Starr-Edwards). The authors observed a weak inverse correlation between valve size and peak and mean gradients in CarboMedics and Carpentier-Edwards valves but not in the Starr-Edwards valve. For the CarboMedics valve the peak pressure gradient (PPG) was 26.1 ±8.2 mm Hg and the mean pressure gradient (MPG) was 14.7 ± 5.1 mm Hg; in Starr-Edwards valve the PPG was 32.8 ±9.1 mm Hg and the MPG was 19.5 ±5.6 mm Hg; in the Carpentier-Edwards valve the PPG was 28.7 ± 10.1 mm Hg and the MPG was 16.1 ± 5.2 mm Hg when size was not specified. The CarboMedics valves were noted to have a better hemodynamic profile in comparison with Starr-Edwards and Carpentier-Edwards prostheses.
SAGE Publications
Title: Doppler Echocardiographic Assessment of Normally Functioning Starr-Edwards, Carbomedics and Carpentier-Edwards Valves in Aortic Position
Description:
Doppler echocardiography was performed in 168 normally functioning aortic prostheses to determine acceptable pressure gradients across the commonly used valves and to establish the relationship between valve size and gradients.
There were 82 Carbomedics (C), 63 Starr-Edwards (SE), and 23 Carpentier-Edwards (CE) valves.
Peak and mean gradients across the prostheses were measured by use of the simplified Bernoulli equation.
CarboMedics valve had a lower peak and mean gradient than Starr-Edwards and Carpentier-Edwards valve (P < 0.
05 when compared with Starr-Edwards).
The authors observed a weak inverse correlation between valve size and peak and mean gradients in CarboMedics and Carpentier-Edwards valves but not in the Starr-Edwards valve.
For the CarboMedics valve the peak pressure gradient (PPG) was 26.
1 ±8.
2 mm Hg and the mean pressure gradient (MPG) was 14.
7 ± 5.
1 mm Hg; in Starr-Edwards valve the PPG was 32.
8 ±9.
1 mm Hg and the MPG was 19.
5 ±5.
6 mm Hg; in the Carpentier-Edwards valve the PPG was 28.
7 ± 10.
1 mm Hg and the MPG was 16.
1 ± 5.
2 mm Hg when size was not specified.
The CarboMedics valves were noted to have a better hemodynamic profile in comparison with Starr-Edwards and Carpentier-Edwards prostheses.
Related Results
P1820New parameters for the evaluation of mechanic and elastic properties of the aortic root in Marfan Syndrome
P1820New parameters for the evaluation of mechanic and elastic properties of the aortic root in Marfan Syndrome
Abstract
Background
Elastic properties of the thoracic aorta in patients with Marfan Syndrome (MS) have already been evaluated w...
new insights into the pathophysiology of rheumatic valve disease based on different genetic and proteomics profiles
new insights into the pathophysiology of rheumatic valve disease based on different genetic and proteomics profiles
Abstract
Introduction
Rheumatic heart disease (RHD) pathogenesis is not fully understood. Purpose The protein profile in rheumat...
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Abstract
Background
Although an association between elevated blood pressure and risk of aortic aneurysm is established, f...
Bicuspid aortic valve flow dynamics using blood speckle tracking in children
Bicuspid aortic valve flow dynamics using blood speckle tracking in children
Abstract
Funding Acknowledgements
Type of funding sources: Foundation. Main funding source(s): Labatt Family Heart Centre
...
Doppler Echocardiographic Assessment of the Bjork‐Shiley Monostrut Valve Prosthesis in the Aortic Position
Doppler Echocardiographic Assessment of the Bjork‐Shiley Monostrut Valve Prosthesis in the Aortic Position
In order to determine the standard Doppler hemodynamic characteristics of the Bjork‐Shiley monostrut aortic prosthesis and the value of the continuity equation to calculate the pro...
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
Objectives
To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia.
...
Aberrant right coronary artery in a grown up congenital cardiac patient, successfully treated 46 years earlier with a double Starr-Edwards silastic ball valve replacement: a case report
Aberrant right coronary artery in a grown up congenital cardiac patient, successfully treated 46 years earlier with a double Starr-Edwards silastic ball valve replacement: a case report
AbstractBackgroundThe Starr-Edwards ball valve prosthesis was successfully introduced in 1961–62 and largely used for aortic and mitral valve replacement. Even if Starr-Edwards val...
Aortic root wrapped procedure
Aortic root wrapped procedure
Objective
To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia.
...