Doppler Echocardiographic Assessment of Normally Functioning Starr-Edwards, Carbomedics and Carpentier-Edwards Valves in Aortic Position

View through CrossRef
Doppler echocardiography was performed in 168 normally functioning aortic prostheses to determine acceptable pressure gradients across the commonly used valves and to establish the relationship between valve size and gradients. There were 82 Carbomedics (C), 63 Starr-Edwards (SE), and 23 Carpentier-Edwards (CE) valves. Peak and mean gradients across the prostheses were measured by use of the simplified Bernoulli equation. CarboMedics valve had a lower peak and mean gradient than Starr-Edwards and Carpentier-Edwards valve (P < 0.05 when compared with Starr-Edwards). The authors observed a weak inverse correlation between valve size and peak and mean gradients in CarboMedics and Carpentier-Edwards valves but not in the Starr-Edwards valve. For the CarboMedics valve the peak pressure gradient (PPG) was 26.1 ±8.2 mm Hg and the mean pressure gradient (MPG) was 14.7 ± 5.1 mm Hg; in Starr-Edwards valve the PPG was 32.8 ±9.1 mm Hg and the MPG was 19.5 ±5.6 mm Hg; in the Carpentier-Edwards valve the PPG was 28.7 ± 10.1 mm Hg and the MPG was 16.1 ± 5.2 mm Hg when size was not specified. The CarboMedics valves were noted to have a better hemodynamic profile in comparison with Starr-Edwards and Carpentier-Edwards prostheses.
Title: Doppler Echocardiographic Assessment of Normally Functioning Starr-Edwards, Carbomedics and Carpentier-Edwards Valves in Aortic Position
Description:
Doppler echocardiography was performed in 168 normally functioning aortic prostheses to determine acceptable pressure gradients across the commonly used valves and to establish the relationship between valve size and gradients.
There were 82 Carbomedics (C), 63 Starr-Edwards (SE), and 23 Carpentier-Edwards (CE) valves.
Peak and mean gradients across the prostheses were measured by use of the simplified Bernoulli equation.
CarboMedics valve had a lower peak and mean gradient than Starr-Edwards and Carpentier-Edwards valve (P < 0.
05 when compared with Starr-Edwards).
The authors observed a weak inverse correlation between valve size and peak and mean gradients in CarboMedics and Carpentier-Edwards valves but not in the Starr-Edwards valve.
For the CarboMedics valve the peak pressure gradient (PPG) was 26.
1 ±8.
2 mm Hg and the mean pressure gradient (MPG) was 14.
7 ± 5.
1 mm Hg; in Starr-Edwards valve the PPG was 32.
8 ±9.
1 mm Hg and the MPG was 19.
5 ±5.
6 mm Hg; in the Carpentier-Edwards valve the PPG was 28.
7 ± 10.
1 mm Hg and the MPG was 16.
1 ± 5.
2 mm Hg when size was not specified.
The CarboMedics valves were noted to have a better hemodynamic profile in comparison with Starr-Edwards and Carpentier-Edwards prostheses.

Related Results

Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Abstract Background Although an association between elevated blood pressure and risk of aortic aneurysm is established, f...
Bicuspid aortic valve flow dynamics using blood speckle tracking in children
Bicuspid aortic valve flow dynamics using blood speckle tracking in children
Abstract Funding Acknowledgements Type of funding sources: Foundation. Main funding source(s): Labatt Family Heart Centre ...
Doppler Echocardiographic Assessment of the Bjork‐Shiley Monostrut Valve Prosthesis in the Aortic Position
Doppler Echocardiographic Assessment of the Bjork‐Shiley Monostrut Valve Prosthesis in the Aortic Position
In order to determine the standard Doppler hemodynamic characteristics of the Bjork‐Shiley monostrut aortic prosthesis and the value of the continuity equation to calculate the pro...
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
Objectives To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia. ...
Aortic root wrapped procedure
Aortic root wrapped procedure
Objective To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia. ...

Back to Top