Measurements of photoneutron dose rate for 15 MV photon beam from medical linear accelerator using neutron survey meter
View through CrossRef
High-energy medical linear accelerators (>10MV) are increasingly used in the medical field to treat cancer patients depending on the treatment organ and patient physic. Medical linear accelerators (Linacs) operating above 10 MV photon beam produce unwanted neutrons (photoneutron) by means of photonuclear reaction (γ, n) between bremsstrahlung photon beams and constituent materials of Linac head. This study involved the measurement of the neutron dose rate from photon beam of medical linear accelerator (Elekta Synergy) operated at 15 MV photon. For the measurement of dose rate, a neutron survey meter LB 6411 probe was used. Neutron dose rate was obtained as a function of delivered dose, field size and detector position. In the study neutron dose rate was found 13.14 mSv/h at a field size of 10 × 10 cm2 at position (0,90,0) cm for 350 cGy and 9.26 mSv/h at 5 × 5 cm2 field size at position (0,80,0) cm for 200 cGy photon dose delivery. The dose rate varied with respect to detector position in an irregular manner; 2.95 mSv/h at position (-100,0,0) cm -on the side of maze entrance and 4.85 mSv/h at position (100,0,0) cm -off side of maze entrance.
Luminescience Press Limited
Title: Measurements of photoneutron dose rate for 15 MV photon beam from medical linear accelerator using neutron survey meter
Description:
High-energy medical linear accelerators (>10MV) are increasingly used in the medical field to treat cancer patients depending on the treatment organ and patient physic.
Medical linear accelerators (Linacs) operating above 10 MV photon beam produce unwanted neutrons (photoneutron) by means of photonuclear reaction (γ, n) between bremsstrahlung photon beams and constituent materials of Linac head.
This study involved the measurement of the neutron dose rate from photon beam of medical linear accelerator (Elekta Synergy) operated at 15 MV photon.
For the measurement of dose rate, a neutron survey meter LB 6411 probe was used.
Neutron dose rate was obtained as a function of delivered dose, field size and detector position.
In the study neutron dose rate was found 13.
14 mSv/h at a field size of 10 × 10 cm2 at position (0,90,0) cm for 350 cGy and 9.
26 mSv/h at 5 × 5 cm2 field size at position (0,80,0) cm for 200 cGy photon dose delivery.
The dose rate varied with respect to detector position in an irregular manner; 2.
95 mSv/h at position (-100,0,0) cm -on the side of maze entrance and 4.
85 mSv/h at position (100,0,0) cm -off side of maze entrance.
Related Results
Comparative study of the effect of neutrons emitted from neutron source 241Am-Be and curcumin on MCF-7 breast cancer cells in 3D culture medium
Comparative study of the effect of neutrons emitted from neutron source 241Am-Be and curcumin on MCF-7 breast cancer cells in 3D culture medium
AbstractIntroductionCancer is one of the major medical problems threatening human health. Breast cancer is the most prevalent type of cancer in women. Reports indicate that treatme...
Study of coded source neutron imaging based on a compact accelerator
Study of coded source neutron imaging based on a compact accelerator
Compact accelerator based neutron source has lower cost and better flexibility than nuclear reactor. Neutron imaging using such a neutron source has attracted more and more attenti...
Update of the radiation environment measurement results aboard ExoMars TGO in May 2018-June 202
Update of the radiation environment measurement results aboard ExoMars TGO in May 2018-June 202
AbstractThe dosimetric telescope Liulin-MO [1] for measuring the radiation environment onboard the ExoMars TGO is a module of the Fine Resolution Epithermal Neutron Detector (FREND...
Development of neutron survey meter using prescila neutron probe
Development of neutron survey meter using prescila neutron probe
This paper presents the design and validation of a neutron survey meter. The meter consists of a PRESCILA neutron probe (with good sensitivity, directional response, gamma rejectio...
serpent 2 Validation for Radiation Shielding Applications
serpent 2 Validation for Radiation Shielding Applications
Abstract
This paper contributes to the validation of serpent's photon transport and coupled neutron–photon transport routines. Two benchmarks presenting measurements...
Neutron-induced fission cross section of 240,242Pu
Neutron-induced fission cross section of 240,242Pu
A recent sensitivity analysis done for the new generation of fast reactors [1] has shown the importance of improved cross section data for several actinides. Among them, the neutro...
Domestic water meter optimal replacement period to minimize water revenue loss
Domestic water meter optimal replacement period to minimize water revenue loss
Water meter under-registration results in apparent losses and lost revenue for municipalities. Municipalities should scientifically determine the optimal replacement periods for th...
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...