Javascript must be enabled to continue!
Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients
View through CrossRef
Objective: The interchangeability of Goldmann Applanation Tonometry (GAT) with or without fluorescein requires clarification among African populations. The present study aimed to determine the agreement between IOP measurement using GAT with and without fluorescein in consecutive patients in Ghana.Design: A prospective cross-sectional clinic-based study design.Setting: The study was conducted at the Ophthalmology and Optometry units of the UCC Eye Clinic, Cape Coast, Ghana.Participants: A convenient sample of 104 consecutive patients who visited the clinic. The participants were randomised into two groups, where those in the first had their IOPs first measured without fluorescein and later with fluorescein on GAT; participants in the second group had their IOPs first measured with fluorescein and later without fluorescein after a 10-minute washout period. The central corneal thickness was measured using the Anterior Segment model of Zeiss Cirrus HD-OCT.Results: There was no agreement between the unadjusted IOP measures with and without fluorescein with GAT using the Bland Altman analysis (p < 0.0001, Lower limit: -0.9433, Upper limit: 6.8914; 95% CI = -1.6097 to -.02769). There were no statistically significant differences in gender between unadjusted IOP measurements with (Mann-Whitney U test, p = 0.110) and without (Mann-Whitney U test, p = 0.083) fluorescein.Conclusion: Unadjusted IOP measurement without fluorescein was approximately 3 mmHg significantly lower than with fluorescein. We recommend practitioners maintain the gold standard GAT with fluorescein to guide accurate glaucoma diagnosis and management decision-making.
African Journals Online (AJOL)
Title: Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients
Description:
Objective: The interchangeability of Goldmann Applanation Tonometry (GAT) with or without fluorescein requires clarification among African populations.
The present study aimed to determine the agreement between IOP measurement using GAT with and without fluorescein in consecutive patients in Ghana.
Design: A prospective cross-sectional clinic-based study design.
Setting: The study was conducted at the Ophthalmology and Optometry units of the UCC Eye Clinic, Cape Coast, Ghana.
Participants: A convenient sample of 104 consecutive patients who visited the clinic.
The participants were randomised into two groups, where those in the first had their IOPs first measured without fluorescein and later with fluorescein on GAT; participants in the second group had their IOPs first measured with fluorescein and later without fluorescein after a 10-minute washout period.
The central corneal thickness was measured using the Anterior Segment model of Zeiss Cirrus HD-OCT.
Results: There was no agreement between the unadjusted IOP measures with and without fluorescein with GAT using the Bland Altman analysis (p < 0.
0001, Lower limit: -0.
9433, Upper limit: 6.
8914; 95% CI = -1.
6097 to -.
02769).
There were no statistically significant differences in gender between unadjusted IOP measurements with (Mann-Whitney U test, p = 0.
110) and without (Mann-Whitney U test, p = 0.
083) fluorescein.
Conclusion: Unadjusted IOP measurement without fluorescein was approximately 3 mmHg significantly lower than with fluorescein.
We recommend practitioners maintain the gold standard GAT with fluorescein to guide accurate glaucoma diagnosis and management decision-making.
Related Results
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Abstract
Introduction
The exact manner in which large language models (LLMs) will be integrated into pathology is not yet fully comprehended. This study examines the accuracy, bene...
Changes in Intraocular Pressure After Laser in situ Keratomileusis
Changes in Intraocular Pressure After Laser in situ Keratomileusis
ABSTRACT
PURPOSE: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia a...
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
ABSTRACT
BACKGROUND: Measured ocular tension is reported to be positively related to corneal thickness. It has been speculated that an edematous cornea will cause low mea...
Development of Fluorescently Labeled Self-Emulsifying Drug Delivery
Systems (SEDDS) for Prolonged Stability, In Vitro Sustained Release, and
Cellular Uptake
Development of Fluorescently Labeled Self-Emulsifying Drug Delivery
Systems (SEDDS) for Prolonged Stability, In Vitro Sustained Release, and
Cellular Uptake
Aim:
In this study, four fluorescein hydrophobic ionic complexes were formed with the
cationic polymers Eudragit RS, Eudragit RL, Eudragit E, and polyethyleneimine (PEI) to provide...
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reli...
Tonometry in three herbivorous wildlife species
Tonometry in three herbivorous wildlife species
Tonometry was performed to estimate intraocular pressure (IOP) in 12 Nubian ibexes (Capra ibex nubiana), 10 Grant zebras (Equus burchelli ) and five Arabian oryxes (Oryx leucoryx),...
Effect of a tight necktie on intraocular pressure
Effect of a tight necktie on intraocular pressure
Introduction. Any factor causing constriction of the neck may lead to an
increase in intraocular pressure. A tight necktie may result in increasing
intraocular pressure, whic...
The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
BACKGROUND፡ The aim of this study was to determine whether Central Corneal Thickness (CCT) is associated with intraocular pressure measurement (IOP) with a Non-contact tonometer an...

