Patient Dose Estimation from Digital Radiography Repeat Rate
View through CrossRef
Purpose: This research aimed at estimating the patient effective dose from digital radiography repeat rateand evaluating the factors that contribute to the repetition of radiographic procedures in two referral hospitals in Sari, Iran.
Materials and Methods: Entrance skin dose and effective dose per common radiography procedures in digital radiography were measured. 1724 X-ray exposures were investigated prospectively. The annual number of each radiography procedure was obtained from the archives of the digital radiography department. The patient cumulative dose was the calculation from annual exposures Repeat Rate (RR) and effective dose per digital radiography procedures.
Results: The mean exposure RR per examination was 1.12% and the total percentage of a repeat of all examination was 8.9%. Annual cumulative dose (man-mSivert) resulted from radiographic RR was 449.2.The chest and lumbar spine radiography had the highest annual number and the highest radiation dose, respectively; therefore, these procedures transfer the largest annual dose from repeated radiographic images to the patients. The factors leading to the repetition included the radiographer error (69%), the X-ray tube and equipment error (10.4%), the patient related error (16.1%) and other cases (4.1%). The average effective dose for each examination was 0.36 mSv.
Conclusion: Digital radiographic repeat rate increases a 1.1 % annual patient effective dose from the base level that receives from the current radiographic examination. Inconsistency of the center of the digital panel and the central axis of the X-ray beam and error in the selection of the upright or table digital panel are among the most important factors in the repetition of digital radiography.
Title: Patient Dose Estimation from Digital Radiography Repeat Rate
Description:
Purpose: This research aimed at estimating the patient effective dose from digital radiography repeat rateand evaluating the factors that contribute to the repetition of radiographic procedures in two referral hospitals in Sari, Iran.
Materials and Methods: Entrance skin dose and effective dose per common radiography procedures in digital radiography were measured.
1724 X-ray exposures were investigated prospectively.
The annual number of each radiography procedure was obtained from the archives of the digital radiography department.
The patient cumulative dose was the calculation from annual exposures Repeat Rate (RR) and effective dose per digital radiography procedures.
Results: The mean exposure RR per examination was 1.
12% and the total percentage of a repeat of all examination was 8.
9%.
Annual cumulative dose (man-mSivert) resulted from radiographic RR was 449.
2.
The chest and lumbar spine radiography had the highest annual number and the highest radiation dose, respectively; therefore, these procedures transfer the largest annual dose from repeated radiographic images to the patients.
The factors leading to the repetition included the radiographer error (69%), the X-ray tube and equipment error (10.
4%), the patient related error (16.
1%) and other cases (4.
1%).
The average effective dose for each examination was 0.
36 mSv.
Conclusion: Digital radiographic repeat rate increases a 1.
1 % annual patient effective dose from the base level that receives from the current radiographic examination.
Inconsistency of the center of the digital panel and the central axis of the X-ray beam and error in the selection of the upright or table digital panel are among the most important factors in the repetition of digital radiography.
.
Related Results
Autonomy on Trial
Autonomy on Trial
Photo by CHUTTERSNAP on Unsplash
Abstract
This paper critically examines how US bioethics and health law conceptualize patient autonomy, contrasting the rights-based, individualist...
Macroeconomic and Social Precursors of Suicide Rates in the Philippines: A Quantitative Analysis (Preprint)
Macroeconomic and Social Precursors of Suicide Rates in the Philippines: A Quantitative Analysis (Preprint)
BACKGROUND
Suicide is a complex, serious and multifaceted public health issue that poses significant challenges to societies worldwide. In fact, it represen...
Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Orthopedic Injury Diagnosis Using Digital Radiography: A Retrospective Study
Background: Hundred patients were examined to determine the effectiveness of digital radiography in detecting Orthopedic injuries. The decisive findings showed that digital radiogr...
Access Denied
Access Denied
Introduction
As social-distancing mandates in response to COVID-19 restricted in-person data collection methods such as participant observation and interviews, researchers turned t...
Comparative Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Bone Fractures
Comparative Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Bone Fractures
Background: Bone fractures represent a substantial proportion of cases presenting to emergency and orthopedic departments worldwide, including in resource-limited settings like Nis...
The role of chest radiography in the diagnosis of bacteriologically confirmed pulmonary tuberculosis in hospitalised Xpert MTB/RIF-negative patients
The role of chest radiography in the diagnosis of bacteriologically confirmed pulmonary tuberculosis in hospitalised Xpert MTB/RIF-negative patients
The role of chest radiography to diagnose active tuberculosis in symptomatic patients who have a negative Xpert MTB/RIF (Xpert) test result is unclear. This study aimed to assess t...
Secondary use of hospital information system data for safe bedside radiography in terms of patient factors
Secondary use of hospital information system data for safe bedside radiography in terms of patient factors
Objective:
The purpose of this study was to use patient data gathered by a hospital information system (HIS) to improve the safe performance of bedside radiogra...
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
Background:
The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) ...