SU‐GG‐T‐572: Are Constant Dose Rate VMAT Plans Clinically Acceptable? — A Site‐By‐Site Plan Comparison Study

View through CrossRef
Purpose: To evaluate the adequacy of constant dose rate (CDR) VMAT plans for clinical use of commonly treated sites. The clinical acceptability of CDR VMAT plans was evaluated based on plan quality comparisons against clinically used IMRT plans. Methods: Several patient treatments planned using IMRT were considered for the study and replanned using Pinnacle 9.0 (SmartArc) treatment planning system. Several clinical sites were considered for this study — prostate and prostate bed (4 patients), pelvis (1), breast (1) and head and neck (H&N) (2). The quality of CDR plans normalized to PTV mean dose was compared against static field IMRT plans based on dose conformality, planning target volume (PTV) coverage and dose volume histograms (DVHs) of organs at risk (OAR). Results: Plan qualities of prostate, prostate bed and pelvis plans were evaluated based on rectum and bladder DVHs, PTV coverage and dose conformality. Prostate and prostate bed: CDR plan was found to be clinically acceptable with better OAR sparing and delivery time. Pelvis: CDR plan was lacking coverage and conformality and didn't have a significant advantage over the IMRT plan. Breast: Plan quality was evaluated based on PTV coverage, isodose conformality and DVHs generated for OARs — lungs, heart and cord. The CDR plan was clinically unacceptable, primarily due to low doses (<10Gy) covering a large volume of the lungs. H&N plans were evaluated based on PTV coverage, dose conformality and dose to parotids, cord and submandibular glands. The CDR plans were clinically unacceptable and didn't show significant advantage over IMRT plans beyond the reduced treatment time. Conclusion: The CDR plans for prostate and prostate bed were clinically acceptable and found to be equivalent, if not superior to IMRT plans. For the other sites investigated ‐ pelvis, breast, and H&N, static field IMRT plans proved to be superior to CDR plans.
Title: SU‐GG‐T‐572: Are Constant Dose Rate VMAT Plans Clinically Acceptable? — A Site‐By‐Site Plan Comparison Study
Description:
Purpose: To evaluate the adequacy of constant dose rate (CDR) VMAT plans for clinical use of commonly treated sites.
The clinical acceptability of CDR VMAT plans was evaluated based on plan quality comparisons against clinically used IMRT plans.
Methods: Several patient treatments planned using IMRT were considered for the study and replanned using Pinnacle 9.
0 (SmartArc) treatment planning system.
Several clinical sites were considered for this study — prostate and prostate bed (4 patients), pelvis (1), breast (1) and head and neck (H&N) (2).
The quality of CDR plans normalized to PTV mean dose was compared against static field IMRT plans based on dose conformality, planning target volume (PTV) coverage and dose volume histograms (DVHs) of organs at risk (OAR).
Results: Plan qualities of prostate, prostate bed and pelvis plans were evaluated based on rectum and bladder DVHs, PTV coverage and dose conformality.
Prostate and prostate bed: CDR plan was found to be clinically acceptable with better OAR sparing and delivery time.
Pelvis: CDR plan was lacking coverage and conformality and didn't have a significant advantage over the IMRT plan.
Breast: Plan quality was evaluated based on PTV coverage, isodose conformality and DVHs generated for OARs — lungs, heart and cord.
The CDR plan was clinically unacceptable, primarily due to low doses (<10Gy) covering a large volume of the lungs.
H&N plans were evaluated based on PTV coverage, dose conformality and dose to parotids, cord and submandibular glands.
The CDR plans were clinically unacceptable and didn't show significant advantage over IMRT plans beyond the reduced treatment time.
Conclusion: The CDR plans for prostate and prostate bed were clinically acceptable and found to be equivalent, if not superior to IMRT plans.
For the other sites investigated ‐ pelvis, breast, and H&N, static field IMRT plans proved to be superior to CDR plans.

Related Results

Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Objectives: This study aimed to find the optimal radiotherapy VMAT plans, that achieved high conformity and homogeneity to the planned target volume (PTV), and minimize the dose to...
Feasibility of 4D VMAT-CT
Feasibility of 4D VMAT-CT
Abstract Objective. Feasibility of three-dimensional (3D) tracking of volumetric modulated arc therapy (VMAT) based on VMAT–computed tomography (V...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
Evaluation of Delta 4 system in patient specific QA for VMAT technique: Retrospective lung VMAT cases
Evaluation of Delta 4 system in patient specific QA for VMAT technique: Retrospective lung VMAT cases
Abstract The objective of this study was to evaluate the Delta 4 system for lung VMAT technique. The standard fields and VMAT plans were generated into the homogeneo...
SU‐E‐P‐44: The Characteristics and Clinical Application of the ArcCHECK for VMAT Verification
SU‐E‐P‐44: The Characteristics and Clinical Application of the ArcCHECK for VMAT Verification
Purpose:To evaluate the characteristics and clinical suitability of the ArcCHECK QA system for VMAT verification.Methods:The intrinsic sensitivity,short and long term reproducibili...