SU‐E‐T‐506: Dosimetric Study for Shallow‐Seated Tumor Using Passive/active Scanning Proton Beam
View through CrossRef
Purpose: To investigate the possibility of using a single spot scanning proton beam to treat superficial lesions. Methods: A cylindrical phantom with a simulated superficial target (it seated 0.5–4cm depth from the surface, volume: 270cm3) was created in Eclipse treatment planning system. Three proton plans were generated: (a) a single AP uniform scanning beam with aperture and range compensator; (b) a single AP spot scanning beam with a pre‐absorber. The location and thickness of the pre‐absorber were calculated using Geant4 to Monte Carlo code to make use of the available spot scanning beams to get a conformal plan. (c) a five‐beam spot scanning beam plan using multi‐field optimization. The prescription is 54 cobalt grey equivalent (CGE) which covers 95% of the target. The target coverage, lateral penumbra at 2 and 4cm depth in water, the doses to normal tissue (phantom‐target) and skin (2mm from the surface) were evaluated and compared for three plans. Results: The mean doses to the target are comparable within 2.4% for all three plans. The conformity indices (at 95%) are 1.36, 1.04 and 0.98 for plan (a), (b) and (c) respectively. The lateral penumbra (80% to 20%) for plan (a), (b) are both 0.73 cm, while it is 3.75 cm for plan (c). The skin dose which received more than 40 (CGE) from plan (a) is 10% higher than that of other two plans. Plan (c) has 70% higher mean doses to normal tissue than that of plan (a) and (b). Conclusions: Each plan provides good coverage of target. And in this study, it showed that, with a properly designed pre‐absorber, it is possible to use a single spot scanning beam to treat superficial lesion. The plan provides good target coverage and maintains normal tissue sparing in the mean time.
Title: SU‐E‐T‐506: Dosimetric Study for Shallow‐Seated Tumor Using Passive/active Scanning Proton Beam
Description:
Purpose: To investigate the possibility of using a single spot scanning proton beam to treat superficial lesions.
Methods: A cylindrical phantom with a simulated superficial target (it seated 0.
5–4cm depth from the surface, volume: 270cm3) was created in Eclipse treatment planning system.
Three proton plans were generated: (a) a single AP uniform scanning beam with aperture and range compensator; (b) a single AP spot scanning beam with a pre‐absorber.
The location and thickness of the pre‐absorber were calculated using Geant4 to Monte Carlo code to make use of the available spot scanning beams to get a conformal plan.
(c) a five‐beam spot scanning beam plan using multi‐field optimization.
The prescription is 54 cobalt grey equivalent (CGE) which covers 95% of the target.
The target coverage, lateral penumbra at 2 and 4cm depth in water, the doses to normal tissue (phantom‐target) and skin (2mm from the surface) were evaluated and compared for three plans.
Results: The mean doses to the target are comparable within 2.
4% for all three plans.
The conformity indices (at 95%) are 1.
36, 1.
04 and 0.
98 for plan (a), (b) and (c) respectively.
The lateral penumbra (80% to 20%) for plan (a), (b) are both 0.
73 cm, while it is 3.
75 cm for plan (c).
The skin dose which received more than 40 (CGE) from plan (a) is 10% higher than that of other two plans.
Plan (c) has 70% higher mean doses to normal tissue than that of plan (a) and (b).
Conclusions: Each plan provides good coverage of target.
And in this study, it showed that, with a properly designed pre‐absorber, it is possible to use a single spot scanning beam to treat superficial lesion.
The plan provides good target coverage and maintains normal tissue sparing in the mean time.
Related Results
Complex Collision Tumors: A Systematic Review
Complex Collision Tumors: A Systematic Review
Abstract
Introduction: A collision tumor consists of two distinct neoplastic components located within the same organ, separated by stromal tissue, without histological intermixing...
TU‐H‐CAMPUS‐JeP3‐04: Factors Predicting a Need for Treatment Replanning with Proton Radiotherapy for Lung Cancer
TU‐H‐CAMPUS‐JeP3‐04: Factors Predicting a Need for Treatment Replanning with Proton Radiotherapy for Lung Cancer
Purpose:Proton dose distribution is sensitive to tumor regression and tissue and normal anatomy changes. Replanning is sometimes necessary during treatment to ensure continue tumor...
MO‐SAM‐BRB‐02: Proton Physics and Technology
MO‐SAM‐BRB‐02: Proton Physics and Technology
The dose localization advantages of proton beams derive primarily from the Bragg peak in the proton stopping distribution. Therefore, the potential clinical gains expected form pro...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
Purpose:To build the model of a spot scanning proton beam for the dose calculation of a synchrotron proton therapy accelerator, which is capable of accelerating protons from 50 up ...
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses...
Giant Sacrococcygeal Teratoma in Infant: Systematic Review
Giant Sacrococcygeal Teratoma in Infant: Systematic Review
Abstract
Introduction
Sacrococcygeal teratoma (SCT) is a rare embryonal tumor that occurs in the sacrococcygeal region, with an incidence of about 1 in 35,000 to 40,000 live births...