Size variability and fitting of 30-gauge thin-wall needles and 3-piece intraocular lens haptics for the flanged intrascleral fixation technique
View through CrossRef
Purpose:
To assess the diameter of different 30-gauge thin-wall needles and 3-piece intraocular lens (IOL) haptics readily used for the flanged-haptic intrascleral fixation technique.
Setting:
Hanusch Hospital, Vienna, Austria.
Design:
Laboratory investigation.
Methods:
5 30-gauge thin-wall needles and 5 3-piece IOLs were assessed. An upright light microscopy was used for measurements. The inner and outer diameters of the needles and the end thickness of the haptics were analyzed and compared for haptic fitting into the needle.
Results:
Among the needles, the inner diameter of the T-lab needle was significantly wider compared with all the others (mean 209.3 ± 8.0 μm, P < .001), followed by TSK (194.8 ± 5.0 μm), MST (194.7 ± 5.8 μm), Sterimedix (187.5 ± 9.0 μm) and significantly narrower Meso-relle (mean 178.7 ± 7.0 μm, P < .05). The outer diameter of the T-lab needle was significantly larger of all (mean 316.0 ± 2.0 μm, P < .001). Concerning the IOLs, the AvanseePreset Kowa's haptic was significantly thinner (mean 127.2 ± 0.7 μm) than all the others, such as the TecnisZA900 Johnson & Johnson (143.5 ± 3.1 μm), the CTLucia202 Zeiss (143.8 ± 1.3 μm), and the AcrysofMA60AC Alcon (143.9 ± 1.4 μm). The only haptic that was thicker than all the others assessed was that of SensarAR40 Johnson & Johnson (170.7 ± 1.7 μm, P < .001).
Conclusions:
Most of the analyzed haptics would fit into most of the measured needles, with the exception of the Sensar AR40 in combination with the Meso-relle or Sterimedix needles. The combination of a larger needle lumen and a thinner haptic could result in more ease of insertion during surgery. If the dimensions of the needle and IOL haptics used are unknown, we recommend trying insertion before beginning surgery.
Ovid Technologies (Wolters Kluwer Health)
Title: Size variability and fitting of 30-gauge thin-wall needles and 3-piece intraocular lens haptics for the flanged intrascleral fixation technique
Description:
Purpose:
To assess the diameter of different 30-gauge thin-wall needles and 3-piece intraocular lens (IOL) haptics readily used for the flanged-haptic intrascleral fixation technique.
Setting:
Hanusch Hospital, Vienna, Austria.
Design:
Laboratory investigation.
Methods:
5 30-gauge thin-wall needles and 5 3-piece IOLs were assessed.
An upright light microscopy was used for measurements.
The inner and outer diameters of the needles and the end thickness of the haptics were analyzed and compared for haptic fitting into the needle.
Results:
Among the needles, the inner diameter of the T-lab needle was significantly wider compared with all the others (mean 209.
3 ± 8.
0 μm, P < .
001), followed by TSK (194.
8 ± 5.
0 μm), MST (194.
7 ± 5.
8 μm), Sterimedix (187.
5 ± 9.
0 μm) and significantly narrower Meso-relle (mean 178.
7 ± 7.
0 μm, P < .
05).
The outer diameter of the T-lab needle was significantly larger of all (mean 316.
0 ± 2.
0 μm, P < .
001).
Concerning the IOLs, the AvanseePreset Kowa's haptic was significantly thinner (mean 127.
2 ± 0.
7 μm) than all the others, such as the TecnisZA900 Johnson & Johnson (143.
5 ± 3.
1 μm), the CTLucia202 Zeiss (143.
8 ± 1.
3 μm), and the AcrysofMA60AC Alcon (143.
9 ± 1.
4 μm).
The only haptic that was thicker than all the others assessed was that of SensarAR40 Johnson & Johnson (170.
7 ± 1.
7 μm, P < .
001).
Conclusions:
Most of the analyzed haptics would fit into most of the measured needles, with the exception of the Sensar AR40 in combination with the Meso-relle or Sterimedix needles.
The combination of a larger needle lumen and a thinner haptic could result in more ease of insertion during surgery.
If the dimensions of the needle and IOL haptics used are unknown, we recommend trying insertion before beginning surgery.
Related Results
Quantum gauge theory simulation with ultracold atoms
Quantum gauge theory simulation with ultracold atoms
The study of ultracold atoms constitutes one of the hottest areas of atomic, molecular, and optical physics and quantum optics. The experimental and theoretical achievements in the...
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
<span style="font-size:11pt"><span style="background:#f9f9f4"><span style="line-height:normal"><span style="font-family:Calibri,sans-serif"><b><spa...
Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
<p><em><span style="font-size: 11.0pt; font-family: 'Times New Roman',serif; mso-fareast-font-family: 'Times New Roman'; mso-ansi-language: EN-US; mso-fareast-langua...
SIMPLE REFIXATION OF A DISLOCATED HAPTIC OF FLANGED INTRAOCULAR LENS USING A 25-GAUGE NEEDLE
SIMPLE REFIXATION OF A DISLOCATED HAPTIC OF FLANGED INTRAOCULAR LENS USING A 25-GAUGE NEEDLE
Purpose:
To report the development of a novel, simple-to-use technique for the salvage of a haptic of dislocated flanged intraocular lens during a retinal detachment su...
FLANGED INTRASCLERAL INTRAOCULAR LENS FIXATION USING 31G NEEDLES
FLANGED INTRASCLERAL INTRAOCULAR LENS FIXATION USING 31G NEEDLES
Purpose:
To demonstrate the novel use of a 31-gauge needle for flanged intrascleral intraocular lens (IOL) fixation using the double-needle technique.
...
Successional patterns of fungal communities in needles of European silver fir (Abies alba Mill.)
Successional patterns of fungal communities in needles of European silver fir (Abies alba Mill.)
SUMMARYSuccessions of fungal communities in mature, senescing and decomposing needles of silver fir (Abies alba Mill.) were studied at two sites in Switzerland. These were in natur...
27-Gauge trocar-assisted sutureless intraocular lens fixation.
27-Gauge trocar-assisted sutureless intraocular lens fixation.
Abstract
Backgrounds: However there have been numerous investigations of intrascleral intraocular lens (IOL) fixation techniques, there is room for improvement in terms of ...
Secondary Intraocular Lens Implantation with Flanged Intrascleral Fixation in Aphakia Patient
Secondary Intraocular Lens Implantation with Flanged Intrascleral Fixation in Aphakia Patient
Cataracts are a leading cause of blindness globally, with surgery being the most effective treatment. While phacoemulsification is the standard procedure in developed nations, case...