Javascript must be enabled to continue!
27-Gauge trocar-assisted sutureless intraocular lens fixation.
View through CrossRef
Abstract
Backgrounds: However there have been numerous investigations of intrascleral intraocular lens (IOL) fixation techniques, there is room for improvement in terms of simplifying complicated techniques and reducing the high levels of skill required. This study aimed to report a novel technique for sutureless intrascleral fixation of the IOL using retinal forceps with a 27-gauge trocar.Methods: Nineteen eyes of 18 patients underwent intrascleral fixation of the IOL from July 2018 to September 2019 were enrolled in this study. A 27-gauge trocar formed 3-mm scleral tunnels positioned at 4 and 10 o’clock, 2 mm from the corneal limbus. We used a 3-piece IOL haptic grasped by a 27-gauge retinal forceps and pulled from the 27-gauge trocar. The IOL was fixed by making a flange. Main outcome measures were visual acuity, corneal endothelial cell density, IOL tilt, decentration, predicted error of refraction and complications.Results: The 19 eyes were followed up for 1 month. The mean pre- and postoperative logMAR uncorrected visual acuity (UCVA) was 1.06 ± 0.63 and 0.40 ± 0.26, respectively (p<0.01), while the mean pre- and postoperative logMAR best corrected visual acuity (BCVA) was 0.27 ± 0.51 and 0.06 ± 0.15, respectively (p=0.09). The mean corneal endothelial cell density was 2406 ± 625 to 2004 ± 759 cells/mm2 at 1 month (p=0.13). The mean IOL tilt was 3.52 ± 3.00°, and the mean IOL decentration was 0.39 ± 0.39 mm. There was no correlation among IOL tilt, decentration and BCVA (p>0.05). The mean prediction error of the target refraction was –0.03 ± 0.93 D. The complications were vitreous hemorrhage (3 eyes), hyphema (1 eye), IOP elevation (1 eye), iris capture of the IOL (1 eye) and hypotony (2 eyes). No IOL dislocation occurred.Conclusions: IOL intrascleral fixation with a flange achieved good IOL fixation and visual outcome in the scleral tunnels created with the 27-gauge trocar.
Springer Science and Business Media LLC
Title: 27-Gauge trocar-assisted sutureless intraocular lens fixation.
Description:
Abstract
Backgrounds: However there have been numerous investigations of intrascleral intraocular lens (IOL) fixation techniques, there is room for improvement in terms of simplifying complicated techniques and reducing the high levels of skill required.
This study aimed to report a novel technique for sutureless intrascleral fixation of the IOL using retinal forceps with a 27-gauge trocar.
Methods: Nineteen eyes of 18 patients underwent intrascleral fixation of the IOL from July 2018 to September 2019 were enrolled in this study.
A 27-gauge trocar formed 3-mm scleral tunnels positioned at 4 and 10 o’clock, 2 mm from the corneal limbus.
We used a 3-piece IOL haptic grasped by a 27-gauge retinal forceps and pulled from the 27-gauge trocar.
The IOL was fixed by making a flange.
Main outcome measures were visual acuity, corneal endothelial cell density, IOL tilt, decentration, predicted error of refraction and complications.
Results: The 19 eyes were followed up for 1 month.
The mean pre- and postoperative logMAR uncorrected visual acuity (UCVA) was 1.
06 ± 0.
63 and 0.
40 ± 0.
26, respectively (p<0.
01), while the mean pre- and postoperative logMAR best corrected visual acuity (BCVA) was 0.
27 ± 0.
51 and 0.
06 ± 0.
15, respectively (p=0.
09).
The mean corneal endothelial cell density was 2406 ± 625 to 2004 ± 759 cells/mm2 at 1 month (p=0.
13).
The mean IOL tilt was 3.
52 ± 3.
00°, and the mean IOL decentration was 0.
39 ± 0.
39 mm.
There was no correlation among IOL tilt, decentration and BCVA (p>0.
05).
The mean prediction error of the target refraction was –0.
03 ± 0.
93 D.
The complications were vitreous hemorrhage (3 eyes), hyphema (1 eye), IOP elevation (1 eye), iris capture of the IOL (1 eye) and hypotony (2 eyes).
No IOL dislocation occurred.
Conclusions: IOL intrascleral fixation with a flange achieved good IOL fixation and visual outcome in the scleral tunnels created with the 27-gauge trocar.
Related Results
Quantum gauge theory simulation with ultracold atoms
Quantum gauge theory simulation with ultracold atoms
The study of ultracold atoms constitutes one of the hottest areas of atomic, molecular, and optical physics and quantum optics. The experimental and theoretical achievements in the...
Trocar localisation for robot-assisted vitreoretinal surgery
Trocar localisation for robot-assisted vitreoretinal surgery
Abstract
Purpose
Robot-assisted vitreoretinal surgery provides precise and consistent operations on the back of the eye. To perform this safely, kno...
Transconjunctival sutureless 20G
Transconjunctival sutureless 20G
Abstract Purpose To evaluate a trocar system that allows the use of the regular 20‐gauge vitrectomy instruments for a transconjunctival sutureless surgery for the treatment of vari...
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Purpose: To study the results of sutureless scleral fixation of intraocular lens (IOL) in cases of complicated cataract surgery.
Study Design: Interventional case series....
Novel trocar assisted intraocular lens and capsular bag complex fixation
Novel trocar assisted intraocular lens and capsular bag complex fixation
Purpose
To describe the innovative technique of trocar-assisted intraocular lens (IOL) and capsular bag complex fixation.
...
Trocar Site Hernia after Laparoscopic Colectomy: A Case Report and Literature Review
Trocar Site Hernia after Laparoscopic Colectomy: A Case Report and Literature Review
Background. Trocar Site Hernia (TSH) is defined as an incisional hernia which occurs after minimally invasive surgery on the trocar incision site.In 2004 Tonouchi classified trocar...
Comparing 3 Different Techniques of Patella Fracture Fixation and Their Complications
Comparing 3 Different Techniques of Patella Fracture Fixation and Their Complications
Introduction: Patella fractures managed by fixation with metal implants often cause local soft tissue irritation and necessitate implant removal. An alternative is to utilize sutur...
Secondary Sutureless Posterior Chamber Lens Implantation with Two Specifically Designed IOLs: Iris Claw Lens versus Sutureless Trans-Scleral Plugs Fixated Lens
Secondary Sutureless Posterior Chamber Lens Implantation with Two Specifically Designed IOLs: Iris Claw Lens versus Sutureless Trans-Scleral Plugs Fixated Lens
Background. The management of patients with aphakia and/or lack of capsular support remains debated. The sutureless posterior chamber IOL (PCIOL) fixation is a very useful surgical...