Abstract 4142243: Efficacy of Radiation Protective Shields on Operator Radiation Exposure During Cardiac Interventions: A Systematic Review and Meta-analysis of Randomized Controlled Trials.
View through CrossRef
Background:
The cumulative exposure to X-ray radiation during cardiac intervention can indeed pose various health risks. This meta-analysis aims to compare radiation protective shields (drapes and X-ray shields) versus conventional safety measures (lead apron) on the operator’s procedural radiation exposure during cardiac interventions.
Methods:
A systematic review and meta-analysis of randomized controlled trials (RCTs) from PubMed, Embase Cochrane, Scopus, and WOS until February 2024. We used the random-effects model to report continuous outcomes using mean difference (MD) with a 95% confidence interval (CI).
Results:
We included 16 RCTs with 3370 patients. Radiation shields were significantly associated with low total operator radiation dose (MD: -7.3, 95%CI [-11.9, -2.7], P< 0.01) with no significant difference between both arms regarding chest radiation dose (MD: -20.7, 95%CI [-48.9, 7.6], P= 0.15) and thyroid radiation dose (MD: -15.4, 95%CI [-32.4, 1.7], P= 0.08).
Also, shields were significantly associated with low air kerma (MD: -46.4, 95%CI [-87.3, -5.5], P= 0.03) and low fluoroscopy duration (MD: -0.3, 95%CI [-0.6, -0.04], P= 0.02). However, there was no difference between both arms regarding the total procedure time (MD: -0.7, 95%CI [-3.1, 1.6], P= 0.54), contrast volume (MD: -3.2, 95%CI [-10.2, 3.7], P= 0.36), and dose area product (MD: 628.4, 95% CI [-3466.9, 4723.8], P= 0.76). Also, we found no differences between drape and shields subgroups in all outcomes.
Conclusion:
Radiation protection shields significantly reduced the total radiation dose exposure and air kerma. Also, shields are associated with lower fluoroscopy duration, insignificant lower procedure time, and contrast volume.
Ovid Technologies (Wolters Kluwer Health)
Title: Abstract 4142243: Efficacy of Radiation Protective Shields on Operator Radiation Exposure During Cardiac Interventions: A Systematic Review and Meta-analysis of Randomized Controlled Trials.
Description:
Background:
The cumulative exposure to X-ray radiation during cardiac intervention can indeed pose various health risks.
This meta-analysis aims to compare radiation protective shields (drapes and X-ray shields) versus conventional safety measures (lead apron) on the operator’s procedural radiation exposure during cardiac interventions.
Methods:
A systematic review and meta-analysis of randomized controlled trials (RCTs) from PubMed, Embase Cochrane, Scopus, and WOS until February 2024.
We used the random-effects model to report continuous outcomes using mean difference (MD) with a 95% confidence interval (CI).
Results:
We included 16 RCTs with 3370 patients.
Radiation shields were significantly associated with low total operator radiation dose (MD: -7.
3, 95%CI [-11.
9, -2.
7], P< 0.
01) with no significant difference between both arms regarding chest radiation dose (MD: -20.
7, 95%CI [-48.
9, 7.
6], P= 0.
15) and thyroid radiation dose (MD: -15.
4, 95%CI [-32.
4, 1.
7], P= 0.
08).
Also, shields were significantly associated with low air kerma (MD: -46.
4, 95%CI [-87.
3, -5.
5], P= 0.
03) and low fluoroscopy duration (MD: -0.
3, 95%CI [-0.
6, -0.
04], P= 0.
02).
However, there was no difference between both arms regarding the total procedure time (MD: -0.
7, 95%CI [-3.
1, 1.
6], P= 0.
54), contrast volume (MD: -3.
2, 95%CI [-10.
2, 3.
7], P= 0.
36), and dose area product (MD: 628.
4, 95% CI [-3466.
9, 4723.
8], P= 0.
76).
Also, we found no differences between drape and shields subgroups in all outcomes.
Conclusion:
Radiation protection shields significantly reduced the total radiation dose exposure and air kerma.
Also, shields are associated with lower fluoroscopy duration, insignificant lower procedure time, and contrast volume.
Related Results
Safety and Efficacy of Atezolizumab in Ovarian Cancer
Safety and Efficacy of Atezolizumab in Ovarian Cancer
Abstract
Introduction
Although the efficacy of PD-L1 blockade has been evaluated in analyses that combine pharmacologically distinct antibodies, the specific efficacy and safety of...
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Abstract
The Physical Activity Guidelines for Americans (Guidelines) advises older adults to be as active as possible. Yet, despite the well documented benefits of physical activi...
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Abstract
Introduction
Tarlatamab is a Delta-like ligand 3 (DLL3) -directed bispecific T-cell engager recently approved for use in patients with advanced small cell lung cancer (SCL...
Do evidence summaries increase health policy‐makers' use of evidence from systematic reviews? A systematic review
Do evidence summaries increase health policy‐makers' use of evidence from systematic reviews? A systematic review
This review summarizes the evidence from six randomized controlled trials that judged the effectiveness of systematic review summaries on policymakers' decision making, or the most...
Pembrolizumab and Sarcoma: A meta-analysis
Pembrolizumab and Sarcoma: A meta-analysis
Abstract
Introduction: Pembrolizumab is a monoclonal antibody that promotes antitumor immunity. This study presents a systematic review and meta-analysis of the efficacy and safety...
Effectiveness of radiation shields to minimize operator dose in the bronchoscopy suite: a phantom study and clinical cases
Effectiveness of radiation shields to minimize operator dose in the bronchoscopy suite: a phantom study and clinical cases
Abstract
Background
The purpose of this study was to 1) evaluate the effectiveness of radiation shields to minimize radiation exposure to the bronchoscopist location in a ...
Digital Mental Health Landscaping in Low- and Middle-Income Countries
Digital Mental Health Landscaping in Low- and Middle-Income Countries
Introduction
The aim of this project was to map the landscape of who is doing what and where in digital mental health, and to pr...
Perancangan Beban Kerja Proses Produksi Pabrik Tahu Ciburial dengan Metode Work Load Analysis
Perancangan Beban Kerja Proses Produksi Pabrik Tahu Ciburial dengan Metode Work Load Analysis
Abstract. Excessive workload can create an uncomfortable working atmosphere for workers because it can trigger the emergence of work stress more quickly. On the other hand, a lack ...