COMPARATIVE STUDY OF TWO OCCLUDER-RELEASE TECHNIQUES IN TRANSCATHETER CLOSURE OF PERIMEMBRANOUS VENTRICULAR SEPTAL DEFECT
View through CrossRef
Objectives
Reported experience with the retained guidewire technique in transcatheter closure of perimembranous ventricular septal defect (VSD) is limited. To compare two occluder-release techniques in transcatheter closure of perimembranous VSD we reviewed our experience in 48 procedures performed on 48 patients.
Methods
From January 2009 to January 2011, 48 patients (26 males and 22 females) with perimembranous VSD underwent an attempt of transcatheter closure using the amplatzer occluder (made in China). The mean age of patients was (9.1±4.9) years (ranged from 3 to 18 years). The mean body weight of patients was (31.1±17.9) kg (ranged from 12 to 52 kg). The mean diameter of VSD measured by transthoracic echocardiography (TTE) was (6.4±3.6) mm (ranged from 3 to 11 mm). The patients were grouped by either the retained guidewire technique (n=24) or non-retained guidewire technique (n=24). Occluder was released through the right heart system. During the operation, an arteriovenous guide wire loop with a 0.035 inch exchange length guide wire (Cordis) was set up through a femoral vein approach. A long sheath was advanced to the left ventricle through the arteriovenous guide wire loop and positioned beneath the aortic valve. Then the guide wire was removed traditionally in non-retained guide wire group, or was not removed creatively in retained guide wire group. The VSD occluder was deployed through the long sheath under fluoroscopic control and echo cardiographic guidance. Procedure time, X-ray exposing time, technical successful rate and complication rate were compared between two groups. All patients were followed up in 1, 6 and 12 months after procedures of TTE, X-ray and electrocardiography.
Results
No significant differences were found in gender, age, weight, the average diameter of VSD (6.3±3.5 vs 6.5±3.7 mm, p>0.05), the distance from VSD to aortic valve (2.3±1.9 vs 2.5±2.1 mm, p>0.05) or the average diameter of amplatzer occluder (9.5±4.1 vs 9.7±5.2 mm, p>0.05). No significant difference was found in technical successful rate (91.7% vs 95.8%, p>0.05), number of applications, or major complication rate (2.1% vs 2.1%, p>0.05). The procedure time in retained guidewire group was less than that in non-retained guide wire group (66.2±31.4 vs 97.5±54.5 min, p<0.05). The X-ray exposing time in retained guide wire group was less than that in non-retained guide wire group (12.4±6.3 vs 21.6±10.7 min, p<0.05). Except 7 cases with mild residual shunt that disappeared in 12 months, no serious complication were found in the other patients, such as device transposition, haemolysis, atrioventricular block, valvular regurgitation and cardiac perforation.
Conclusions
These data suggest that although outcomes and major complication rates are similar for the two groups, transcatheter closure of perimembranous ventricular septal defect with retained guide wire technique would shorter operative time, and reduce the amount of x-ray irradiation of the doctors and patients.
Title: COMPARATIVE STUDY OF TWO OCCLUDER-RELEASE TECHNIQUES IN TRANSCATHETER CLOSURE OF PERIMEMBRANOUS VENTRICULAR SEPTAL DEFECT
Description:
Objectives
Reported experience with the retained guidewire technique in transcatheter closure of perimembranous ventricular septal defect (VSD) is limited.
To compare two occluder-release techniques in transcatheter closure of perimembranous VSD we reviewed our experience in 48 procedures performed on 48 patients.
Methods
From January 2009 to January 2011, 48 patients (26 males and 22 females) with perimembranous VSD underwent an attempt of transcatheter closure using the amplatzer occluder (made in China).
The mean age of patients was (9.
1±4.
9) years (ranged from 3 to 18 years).
The mean body weight of patients was (31.
1±17.
9) kg (ranged from 12 to 52 kg).
The mean diameter of VSD measured by transthoracic echocardiography (TTE) was (6.
4±3.
6) mm (ranged from 3 to 11 mm).
The patients were grouped by either the retained guidewire technique (n=24) or non-retained guidewire technique (n=24).
Occluder was released through the right heart system.
During the operation, an arteriovenous guide wire loop with a 0.
035 inch exchange length guide wire (Cordis) was set up through a femoral vein approach.
A long sheath was advanced to the left ventricle through the arteriovenous guide wire loop and positioned beneath the aortic valve.
Then the guide wire was removed traditionally in non-retained guide wire group, or was not removed creatively in retained guide wire group.
The VSD occluder was deployed through the long sheath under fluoroscopic control and echo cardiographic guidance.
Procedure time, X-ray exposing time, technical successful rate and complication rate were compared between two groups.
All patients were followed up in 1, 6 and 12 months after procedures of TTE, X-ray and electrocardiography.
Results
No significant differences were found in gender, age, weight, the average diameter of VSD (6.
3±3.
5 vs 6.
5±3.
7 mm, p>0.
05), the distance from VSD to aortic valve (2.
3±1.
9 vs 2.
5±2.
1 mm, p>0.
05) or the average diameter of amplatzer occluder (9.
5±4.
1 vs 9.
7±5.
2 mm, p>0.
05).
No significant difference was found in technical successful rate (91.
7% vs 95.
8%, p>0.
05), number of applications, or major complication rate (2.
1% vs 2.
1%, p>0.
05).
The procedure time in retained guidewire group was less than that in non-retained guide wire group (66.
2±31.
4 vs 97.
5±54.
5 min, p<0.
05).
The X-ray exposing time in retained guide wire group was less than that in non-retained guide wire group (12.
4±6.
3 vs 21.
6±10.
7 min, p<0.
05).
Except 7 cases with mild residual shunt that disappeared in 12 months, no serious complication were found in the other patients, such as device transposition, haemolysis, atrioventricular block, valvular regurgitation and cardiac perforation.
Conclusions
These data suggest that although outcomes and major complication rates are similar for the two groups, transcatheter closure of perimembranous ventricular septal defect with retained guide wire technique would shorter operative time, and reduce the amount of x-ray irradiation of the doctors and patients.
Related Results
Cardiovascular Malformations Among Preterm Infants
Cardiovascular Malformations Among Preterm Infants
Objective. Preterm birth and cardiovascular malformations are the 2 most common causes of neonatal and infant death, but there are no published population-based reports on the rela...
The clinical evaluation of transcatheter closure of perimembranous ventricular septal defects with the Amplatzer duct occluder II
The clinical evaluation of transcatheter closure of perimembranous ventricular septal defects with the Amplatzer duct occluder II
Objectives
Investigate the feasibility of perimembranous ventricular septal defects closure with the Amplatzer duct occluder II.
...
Advanced three-dimensional surgical planning of complex ventricular anatomy in transposition of the great arteries
Advanced three-dimensional surgical planning of complex ventricular anatomy in transposition of the great arteries
Abstract
This is a case of a 6-week-old male with D-TGA and multiple large ventricular septal defects who presented with oxygen desaturations into the 60s. He had initially unde...
Transcatheter and hybrid closure of muscular ventricular septal defects using the KONAR-MF™ occluder: a multicentre paediatric experience in India
Transcatheter and hybrid closure of muscular ventricular septal defects using the KONAR-MF™ occluder: a multicentre paediatric experience in India
Abstract
Background:
The KONAR-MF™ occluder, with its flexible medium-profile design, has broadened the feasibility of tr...
ASSA13-15-6 Transcatheter Closure of Patent Ductus Arteriosus with Severe Pulmonary Hypertension Using Domestic Occluder in Adults
ASSA13-15-6 Transcatheter Closure of Patent Ductus Arteriosus with Severe Pulmonary Hypertension Using Domestic Occluder in Adults
Background
To evaluate the clinical efficacy of transcatheter closure of patent ductus arteriosus (PDA) with severe pulmonary hypertention using domestic occluder...
The Incidence of Isolated Ventricular Septal Defect in Libyan Newborns
The Incidence of Isolated Ventricular Septal Defect in Libyan Newborns
This study aimed to evaluate the incidence and spontaneous closure of ventricular septal defects in a randomly selected newborn population, using color Doppler echocardiographic sc...
Off‐label use of duct occluder devices to close hemodynamically significant perimembranous ventricular septal defects: A multicenter experience
Off‐label use of duct occluder devices to close hemodynamically significant perimembranous ventricular septal defects: A multicenter experience
AbstractObjectiveWe sought to evaluate the feasibility, technical aspects, and outcome of transcatheter perimembranous ventricular septal defect (pmVSD) closure using duct occluder...
Outcomes of closure of doubly committed subarterial ventricular septal defects in adults
Outcomes of closure of doubly committed subarterial ventricular septal defects in adults
AbstractBackground:Outcome data of doubly committed subarterial ventricular septal defect closure in adults are limited.Methods:A review was made of the inpatients >18 years of ...