Can the Oxygenator Screen Filter Reduce Gaseous Microemboli?

View through CrossRef
Gaseous microemboli (GME) define small bubbles as <200 μm in size. GME are reported to increase morbidity after cardiopulmonary bypass (CPB) and cardiac surgery. To prevent intrusion of GME into the systemic circulation during CPB, arterial line filtration is generally recommended. New trends in oxygenator design promote location of arterial filtration as an integral part of the oxygenator housing. The present experimental study aimed to evaluate the GME removal properties of an integrated arterial screen filter in a standard microporous oxygenator. The GME properties of Terumo Capiox® FX25 with an integrated arterial screen filter was assessed in an experimental setup and compared with Capiox® RX25, in which no arterial screen filter is present. A blood analog prime solution was recirculated using a roller pump at 4 and 6 L per minute flow rate, respectively, through a customized CPB circuit comprising oxygenator, reservoir, and connecting tubing. A controlled volume of air was introduced into the circuit. The GME activity was measured and computed using a Gampt BCC200® ultrasonic device placing one probe at the venous inlet and one other at the arterial outlet of the oxygenator. Transmembrane delta values of GME activity were used to calculate the removal efficacy based on counts and volume of GME. Use of screen filtration reduced the GME volume by 99.1% ± .1% compared with 98.0% ± .1% for controls at 4 L/min flow rate (p< .001). At 6 L/min, the reduction was 97.9% ± .1% compared with 97.0% ± .1% (p< .001). In contrast, the reduction of GME counts was less effective after screen filtration compared with controls: 89.6 ± .6% versus 91.4 ± .4% at 4 L/min and 55.6% ± 1.6% versus 76.0% ± 1.4% at 6 L/min, respectively (p< .001). The tested oxygenator with incorporated arterial screen filter reduced GME activity based on the calculated volume at the same time as counts of GME increased.
Title: Can the Oxygenator Screen Filter Reduce Gaseous Microemboli?
Description:
Gaseous microemboli (GME) define small bubbles as <200 μm in size.
GME are reported to increase morbidity after cardiopulmonary bypass (CPB) and cardiac surgery.
To prevent intrusion of GME into the systemic circulation during CPB, arterial line filtration is generally recommended.
New trends in oxygenator design promote location of arterial filtration as an integral part of the oxygenator housing.
The present experimental study aimed to evaluate the GME removal properties of an integrated arterial screen filter in a standard microporous oxygenator.
The GME properties of Terumo Capiox® FX25 with an integrated arterial screen filter was assessed in an experimental setup and compared with Capiox® RX25, in which no arterial screen filter is present.
A blood analog prime solution was recirculated using a roller pump at 4 and 6 L per minute flow rate, respectively, through a customized CPB circuit comprising oxygenator, reservoir, and connecting tubing.
A controlled volume of air was introduced into the circuit.
The GME activity was measured and computed using a Gampt BCC200® ultrasonic device placing one probe at the venous inlet and one other at the arterial outlet of the oxygenator.
Transmembrane delta values of GME activity were used to calculate the removal efficacy based on counts and volume of GME.
Use of screen filtration reduced the GME volume by 99.
1% ± .
1% compared with 98.
0% ± .
1% for controls at 4 L/min flow rate (p< .
001).
At 6 L/min, the reduction was 97.
9% ± .
1% compared with 97.
0% ± .
1% (p< .
001).
In contrast, the reduction of GME counts was less effective after screen filtration compared with controls: 89.
6 ± .
6% versus 91.
4 ± .
4% at 4 L/min and 55.
6% ± 1.
6% versus 76.
0% ± 1.
4% at 6 L/min, respectively (p< .
001).
The tested oxygenator with incorporated arterial screen filter reduced GME activity based on the calculated volume at the same time as counts of GME increased.

Related Results

CFD Simulation and Optimization of a Cake Filtration System
CFD Simulation and Optimization of a Cake Filtration System
Abstract This study presents a simulation of filter cake formation during the filtration of rice hull ash and liquid mixture using ANSYS Fluent software. Filter cake...
Air handling characteristics of five membrane oxygenators
Air handling characteristics of five membrane oxygenators
This project looked at the potential of five different membrane oxygenators to allow passage of catastrophic quantities of air in a clinically simulated environment. All the oxygen...
Cerebral Microemboli and Brain Injury During Carotid Artery Endarterectomy and Stenting
Cerebral Microemboli and Brain Injury During Carotid Artery Endarterectomy and Stenting
Background and Purpose— Cerebral microembolic signals detected by transcranial Doppler are frequent during carotid angioplasty with stentin...
Implementasi Bandpass Filter M-derifed pada pemancar radio FM
Implementasi Bandpass Filter M-derifed pada pemancar radio FM
Pemancar siaran radio FM adalah salah satu sistem komunikasi nirkabel, bekerja rentang frekuensi 88-108 MHz membutuhkan sebuah filter bandpass untuk menghindari adanya interferensi...
ECMO circuit monitoring
ECMO circuit monitoring
Extracorporeal membrane oxygenation or ECMO in short, is an external circuit that acts as a heart-lung machine or life support system. This treatment is administered to critically ...
The relative safety of an oxygenator
The relative safety of an oxygenator
Analysis of prospectively registered incidents related to cardiopulmonary bypass (CPB) was initiated to establish the incident rate for 10 different oxygenator brands employed over...
IoT‐based mock oxygenator for extracorporeal membrane oxygenation simulator
IoT‐based mock oxygenator for extracorporeal membrane oxygenation simulator
Abstract Background Training is an essential aspect of providing high‐quality treatment and ensuring patient safety in an...

Back to Top