Correlation and variation of cuff inflating volumes and pressures in the adult laryngeal masks: a prospective randomized trial
View through CrossRef
Abstract
Background: Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications. The manufacturer recommends inflating cuff until the intracuff pressure reaches 60 cmH2O, or inflate with the volume of air to not exceed the maximum recommended volume. We prospectively assessed the correlation of cuff inflating volumes and pressures, and the appropriated the cuff inflating volumes to generate an intracuff pressure of 60 cmH2O in the adult laryngeal masks from different manufacturers.
Methods: Two groups of eighty patients requiring laryngeal mask size 3 and 4 during general anesthesia were randomized into 4 subgroups for each size of the laryngeal mask: Soft Seal® (Portex®), AuraOnceTM (Ambu®), LMA-ClassicTM (Teleflex®) and LMA-ProSealTM (Teleflex®). After insertion, the cuff was inflated with 5-ml increments of air up to the maximum recommended volume. After each 5-ml intracuff pressure was measured, the volume of air that generated the intracuff pressure of 60 cmH2O was recorded.
Results: Mean(SD) volume of air required to achieve the intracuff pressure of 60 cmH2O in Soft Seal®, AuraOnceTM, LMA-ClassicTM, LMA-ProSealTM laryngeal mask size 3 were 11.80(1.88), 9.20(1.88), 8.95(1.50) and 13.50(2.48) ml, respectively, and these volumes in laryngeal mask size 4 were 14.45(4.12), 12.55(1.85), 11.30(1.95) and 18.20(3.47) ml, respectively. The maximum recommended volume resulted in high intracuff pressures (>60 cmH2O) in all laryngeal mask types and sizes studied.
Conclusion: Approximately half of the maximum recommended volume is required to achieve the intracuff pressure of 60 cmH2O except LMA-ProSealTM which required two-thirds of the maximum recommended volume.
Trial Registration: Thai Clinical Trials Registry, TCTR20150602001, May 28, 2015.
Keywords: Laryngeal mask, cuff inflating volume, intracuff pressure.
Springer Science and Business Media LLC
Title: Correlation and variation of cuff inflating volumes and pressures in the adult laryngeal masks: a prospective randomized trial
Description:
Abstract
Background: Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications.
The manufacturer recommends inflating cuff until the intracuff pressure reaches 60 cmH2O, or inflate with the volume of air to not exceed the maximum recommended volume.
We prospectively assessed the correlation of cuff inflating volumes and pressures, and the appropriated the cuff inflating volumes to generate an intracuff pressure of 60 cmH2O in the adult laryngeal masks from different manufacturers.
Methods: Two groups of eighty patients requiring laryngeal mask size 3 and 4 during general anesthesia were randomized into 4 subgroups for each size of the laryngeal mask: Soft Seal® (Portex®), AuraOnceTM (Ambu®), LMA-ClassicTM (Teleflex®) and LMA-ProSealTM (Teleflex®).
After insertion, the cuff was inflated with 5-ml increments of air up to the maximum recommended volume.
After each 5-ml intracuff pressure was measured, the volume of air that generated the intracuff pressure of 60 cmH2O was recorded.
Results: Mean(SD) volume of air required to achieve the intracuff pressure of 60 cmH2O in Soft Seal®, AuraOnceTM, LMA-ClassicTM, LMA-ProSealTM laryngeal mask size 3 were 11.
80(1.
88), 9.
20(1.
88), 8.
95(1.
50) and 13.
50(2.
48) ml, respectively, and these volumes in laryngeal mask size 4 were 14.
45(4.
12), 12.
55(1.
85), 11.
30(1.
95) and 18.
20(3.
47) ml, respectively.
The maximum recommended volume resulted in high intracuff pressures (>60 cmH2O) in all laryngeal mask types and sizes studied.
Conclusion: Approximately half of the maximum recommended volume is required to achieve the intracuff pressure of 60 cmH2O except LMA-ProSealTM which required two-thirds of the maximum recommended volume.
Trial Registration: Thai Clinical Trials Registry, TCTR20150602001, May 28, 2015.
Keywords: Laryngeal mask, cuff inflating volume, intracuff pressure.
Related Results
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial
Abstract
Background
Hyperinflation of laryngeal mask cuffs may carry the risk of airway complicati...
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial
Abstract
Background: Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications. The manufacturer recommends inflating cuff until the intracuff press...
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask airway: a prospective randomized trial
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask airway: a prospective randomized trial
Abstract
Background : Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications. The manufacturer recommends inflating cuff until the intracuff pres...
Safety and Efficacy of Atezolizumab in Ovarian Cancer
Safety and Efficacy of Atezolizumab in Ovarian Cancer
Abstract
Introduction
Although the efficacy of PD-L1 blockade has been evaluated in analyses that combine pharmacologically distinct antibodies, the specific efficacy and safety of...
“ENDOTRACHEAL TUBE CUFF PRESSURE MONITORING USING PREFIXED VOLUME AIR VERSUS MANOMETER IN GENERAL ANAESTHESIA”
“ENDOTRACHEAL TUBE CUFF PRESSURE MONITORING USING PREFIXED VOLUME AIR VERSUS MANOMETER IN GENERAL ANAESTHESIA”
Background And Objective: The endotracheal tube (ETT) cuff seals the airway, preventing leakage and
aspiration of pharyngeal contents into the trachea during ventilation. The Endot...
Comparison of endotracheal tube cuff inflation techniques by stethoscope and “just seal” method in patients undergoing surgeries under general anesthesia
Comparison of endotracheal tube cuff inflation techniques by stethoscope and “just seal” method in patients undergoing surgeries under general anesthesia
Background: Overinflation of the endotracheal tube cuff affects tracheal mucosa blood supply that causes postoperative complications like cough, sore throat and hoarseness. There i...
Arthroscopic cuff repair: footprint remnant preserving versus debriding rotator cuff repair of transtendinous rotator cuff tears with remnant cuff
Arthroscopic cuff repair: footprint remnant preserving versus debriding rotator cuff repair of transtendinous rotator cuff tears with remnant cuff
Abstract
Background
In transtendinous full thickness rotator cuff tears (FTRCT) with remnant cuff, conventionally, cuff remnant of the greater tuber...
A Literature Review on Strategies Preventing Ventilator-Associated Pneumonia: Role of Endotracheal Cuff Pressure, Cuff Material, and Cuff Shape
A Literature Review on Strategies Preventing Ventilator-Associated Pneumonia: Role of Endotracheal Cuff Pressure, Cuff Material, and Cuff Shape
Objectives: To provide an overview of existing literature regarding the effects of ETT cuff pressure, cuff material, and cuff shape on VAP prevention in adult ventilated patients. ...