SU‐GG‐T‐145: Dose Response of Spectral Change of AS1000 EPID

View through CrossRef
Purpose: Amorphous silicon electronic portal imaging devices have been used as 2D dosimeters due to its excellent dose linearity response. However, dose response (cGy/signal) may be spectra‐dependent when patients are in the beam. The aim of this study was to evaluate the dosimetric sensitivity of a Varian aS1000 EPID to spectral change caused by the presence of a patient and give conclusive recommendation for clinical applications. Method and Materials: The signal‐to‐absolute‐dose response curves of the EPID were established for 0 to 40 cm thick of solid water phantoms on couch. The thickness can well cover the water equivalent thicknesses of most patients from head (11 to 19 cm) to pelvis (AP 15 to 25 cm, lateral 26 to 37 cm). Monte Carlo simulation (BEAMnrc 2006) was used to evaluate spectral variation at the EPID plane. Results: The results show that spectra at the EPID plane change dramatically for phantom thickness from 0 to 5 cm, and become more stable with further increased phantom thickness. Direct measurements have shown that large discrepancy up to 9.6% can be found for dose response within phantom thickness between 0 and 30 cm. It was also shown that the max variation between 10 to 40 cm which cover thicknesses of head and pelvis, 10 to 20 cm which cover thicknesses of head, 15 to 40 cm which cover thicknesses of pelvis, were 2.7%, 1.3%, 1.1% respectively. Conclusion: We thus concluded that for clinical absolute dose measurements, opened field dosimetry (without patient in the beam) and transit dosimetry should have identical dose response curves. From the different variations of dose response among 10 to 40 cm, 10 to 20 cm, and 15 to 40 cm, we suggest that for transit dosimetry, head and pelvis should have its own one dose response curve.
Title: SU‐GG‐T‐145: Dose Response of Spectral Change of AS1000 EPID
Description:
Purpose: Amorphous silicon electronic portal imaging devices have been used as 2D dosimeters due to its excellent dose linearity response.
However, dose response (cGy/signal) may be spectra‐dependent when patients are in the beam.
The aim of this study was to evaluate the dosimetric sensitivity of a Varian aS1000 EPID to spectral change caused by the presence of a patient and give conclusive recommendation for clinical applications.
Method and Materials: The signal‐to‐absolute‐dose response curves of the EPID were established for 0 to 40 cm thick of solid water phantoms on couch.
The thickness can well cover the water equivalent thicknesses of most patients from head (11 to 19 cm) to pelvis (AP 15 to 25 cm, lateral 26 to 37 cm).
Monte Carlo simulation (BEAMnrc 2006) was used to evaluate spectral variation at the EPID plane.
Results: The results show that spectra at the EPID plane change dramatically for phantom thickness from 0 to 5 cm, and become more stable with further increased phantom thickness.
Direct measurements have shown that large discrepancy up to 9.
6% can be found for dose response within phantom thickness between 0 and 30 cm.
It was also shown that the max variation between 10 to 40 cm which cover thicknesses of head and pelvis, 10 to 20 cm which cover thicknesses of head, 15 to 40 cm which cover thicknesses of pelvis, were 2.
7%, 1.
3%, 1.
1% respectively.
Conclusion: We thus concluded that for clinical absolute dose measurements, opened field dosimetry (without patient in the beam) and transit dosimetry should have identical dose response curves.
From the different variations of dose response among 10 to 40 cm, 10 to 20 cm, and 15 to 40 cm, we suggest that for transit dosimetry, head and pelvis should have its own one dose response curve.

Related Results

Uusi suomalais-virolainen sanakirja
Uusi suomalais-virolainen sanakirja
Kirja-arvioPihel, Kalju & Pikamäe, Arno: Soome-eesti sõnaraamatKielenaineksetaallokas (kieli: suomi, sivulla: 144)aaltoinen (kieli: suomi, sivulla: 144)aamen (kieli: suomi, siv...
SU‐E‐T‐05: A 2D EPID Transit Dosimetry Model Based On An Empirical Quadratic Formalism
SU‐E‐T‐05: A 2D EPID Transit Dosimetry Model Based On An Empirical Quadratic Formalism
Purpose:To describe a 2D electronic portal imaging device (EPID) transit dosimetry model, based on an empirical quadratic formalism, that can predict either EPID or in‐phantom dose...
Abstract 5025: Regulation of miR-145 expression by Foxo3a, C/EBP-β and p53 in cancer cells
Abstract 5025: Regulation of miR-145 expression by Foxo3a, C/EBP-β and p53 in cancer cells
Abstract Background: MicroRNAs are master gene regulators that control gene expression through the posttranscriptional repression mechanism. Accumulating evidence in...
High through-put identification of miR-145 targets in human articular chondrocytes
High through-put identification of miR-145 targets in human articular chondrocytes
ABSTRACTObjectiveMicroRNAs play a key role in biological processes, including cartilage development and homeostasis and are dysregulated in many diseases, including osteoarthritis....
Abstract 2350: MicroRNA-145 suppresses cell invasion and metastasis by directly targeting mucin 1
Abstract 2350: MicroRNA-145 suppresses cell invasion and metastasis by directly targeting mucin 1
Abstract MicroRNAs (miRNAs) are important gene regulators that could play a profound role in tumorigenesis. Our previous studies indicate that miR-145 is a tumor sup...

Back to Top