P1090DETERMINATION OF ABSOLUTE BLOOD VOLUME USING ONLINE DIALYSATE DILUTION: WHEN SHOULD BE MEASURED?

View through CrossRef
Abstract Background and Aims Current on-line haemodiafiltration (HDF) machines equipped with a blood volume monitor (BVM) and an on-line bolus function have the potential for measuring absolute blood volume (aBV). Recently, we developed a simple method to determine absolute BV in everyday dialysis sessions. The aim of the present study was to evaluate the reproducibility of measurements. Method Intra-individual reproducibility was studied in 10 patients during a single dialysis session by 4 measurements of absolute BV: immediately after beginning before ultrafiltration (UF) was started, and after one, two and three hours. ABV was determined by indicator dilution. A defined volume bolus of 240 mL dialysate was infused into the venous blood line by pressing the emergency button of the HDF machine 5008 (FMC). For this reason, total UF volume was increased by 1L. UF was automatically stopped during and after the infusion. The resulting increase in relative blood volume (RBVpost-RBVpre) was measured by the ultrasonic relative BVM incorporated in the dialysis machine. ABV was measured in hourly intervals and for assessment of reproducibility the volume at treatment start (t=0) where RBV is 100% was calculated for all measurements as: aBV in mL = bolus volume 240 mL x 100% / increase RBV in % ABV data were normalized for body mass at dry weight (in mL/kg). Additionally, in 5 patients the RBV graph was monitored immediately at the beginning of dialysis without UF in a separate dialysis session. Results ABV at t=0 were consistently larger when calculated from measurements done immediately after the beginning compared to measurements obtained after 1 h (6.52 ± 1.40 L or 80.6 ± 14.5 mL/kg vs. 5.16 ± 1.40 L or 63.9 ± 14.3 mL/kg). Specific BV derived from 2 and 3 h measurements did not significantly differ from the measured volumes after 1 hour (61.4 ± 13.8 mL/kg, and 60.9 ± 13.9 mL/kg). The standard deviations of the 3 examinations in the same study patient during a further course of dialysis were between 0.6 and 5.3 ml/kg (ø 2.6 ml/kg). In a separate session, RBV decreases without UF at the beginning of dialysis in the first 3 minutes by 0.5 % and in 5 minutes by 0.6 %. Conclusion If BV is diluted by additional priming volume and bolus volume, a part of this volume will leave the circulation. This represented the time frame where the bolus was initially infused and the measurements were carried out. This loss is caused by the reduction in plasma colloid osmotic pressure induced by the dilution thereby changing the microvascular filtration equilibrium. The increase in RBV display is not solely caused by the bolus volume in this time and, and therefore, calculated BV would be overestimated by about 17 mL/kg. If measurement is performed at a later time, UF will take place and, consequently, refilling. This inward drive matches the outward bolus escape as a counterforce. BV measurement during a further course of dialysis is well reproducible with a deviation of only ± 2.6 ml/kg. The method would therefore be sufficiently precise in clinical practice. Therefore, we propose the determination of aBV only after 1 hour dialysis when a sufficient refilling takes place. With a software modification, the BV measurement could be routinely automated during each dialysis treatment. Manufacturers are asked to implement this technology in their devices.
Title: P1090DETERMINATION OF ABSOLUTE BLOOD VOLUME USING ONLINE DIALYSATE DILUTION: WHEN SHOULD BE MEASURED?
Description:
Abstract Background and Aims Current on-line haemodiafiltration (HDF) machines equipped with a blood volume monitor (BVM) and an on-line bolus function have the potential for measuring absolute blood volume (aBV).
Recently, we developed a simple method to determine absolute BV in everyday dialysis sessions.
The aim of the present study was to evaluate the reproducibility of measurements.
Method Intra-individual reproducibility was studied in 10 patients during a single dialysis session by 4 measurements of absolute BV: immediately after beginning before ultrafiltration (UF) was started, and after one, two and three hours.
ABV was determined by indicator dilution.
A defined volume bolus of 240 mL dialysate was infused into the venous blood line by pressing the emergency button of the HDF machine 5008 (FMC).
For this reason, total UF volume was increased by 1L.
UF was automatically stopped during and after the infusion.
The resulting increase in relative blood volume (RBVpost-RBVpre) was measured by the ultrasonic relative BVM incorporated in the dialysis machine.
ABV was measured in hourly intervals and for assessment of reproducibility the volume at treatment start (t=0) where RBV is 100% was calculated for all measurements as: aBV in mL = bolus volume 240 mL x 100% / increase RBV in % ABV data were normalized for body mass at dry weight (in mL/kg).
Additionally, in 5 patients the RBV graph was monitored immediately at the beginning of dialysis without UF in a separate dialysis session.
Results ABV at t=0 were consistently larger when calculated from measurements done immediately after the beginning compared to measurements obtained after 1 h (6.
52 ± 1.
40 L or 80.
6 ± 14.
5 mL/kg vs.
5.
16 ± 1.
40 L or 63.
9 ± 14.
3 mL/kg).
Specific BV derived from 2 and 3 h measurements did not significantly differ from the measured volumes after 1 hour (61.
4 ± 13.
8 mL/kg, and 60.
9 ± 13.
9 mL/kg).
The standard deviations of the 3 examinations in the same study patient during a further course of dialysis were between 0.
6 and 5.
3 ml/kg (ø 2.
6 ml/kg).
In a separate session, RBV decreases without UF at the beginning of dialysis in the first 3 minutes by 0.
5 % and in 5 minutes by 0.
6 %.
Conclusion If BV is diluted by additional priming volume and bolus volume, a part of this volume will leave the circulation.
This represented the time frame where the bolus was initially infused and the measurements were carried out.
This loss is caused by the reduction in plasma colloid osmotic pressure induced by the dilution thereby changing the microvascular filtration equilibrium.
The increase in RBV display is not solely caused by the bolus volume in this time and, and therefore, calculated BV would be overestimated by about 17 mL/kg.
If measurement is performed at a later time, UF will take place and, consequently, refilling.
This inward drive matches the outward bolus escape as a counterforce.
BV measurement during a further course of dialysis is well reproducible with a deviation of only ± 2.
6 ml/kg.
The method would therefore be sufficiently precise in clinical practice.
Therefore, we propose the determination of aBV only after 1 hour dialysis when a sufficient refilling takes place.
With a software modification, the BV measurement could be routinely automated during each dialysis treatment.
Manufacturers are asked to implement this technology in their devices.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
What is the Proper Composition of the Dialysate Magnesium and How Much Magnesium is Removed During Pre-Dilution Online Hemodiafiltration?
What is the Proper Composition of the Dialysate Magnesium and How Much Magnesium is Removed During Pre-Dilution Online Hemodiafiltration?
Aim: The aim of the study was to determine the ideal dialysate magnesium during hemodialysis, knowing that the solution used is not satisfactory and is often associated with low se...
Haemodialysis with low dialysate flow rates: A comparison of high performance membranes and conventional membranes
Haemodialysis with low dialysate flow rates: A comparison of high performance membranes and conventional membranes
Summary: We performed experimental studies to examine the effects of low dialysate flow rates on the clearance of small molecular weight substances, a middle molecular weight subs...
Comparison of Two Different Kt/V Methods in Continuous Ambulatory Peritoneal Dialysis Patients
Comparison of Two Different Kt/V Methods in Continuous Ambulatory Peritoneal Dialysis Patients
Dialysis adequacy has gained particular interest for the assessment of the quality of dialysis in patients undergoing hemodialysis (HD) and continuous ambulatory peritoneal dialysi...
Could cooling dialysate improve inflammatory and nutritional status of hemodialysis patients?
Could cooling dialysate improve inflammatory and nutritional status of hemodialysis patients?
AbstractBackgroundIt has been shown that dialysate cooling (lowering the dialysate temperature to 0.5 °C below central body temperature) reduces the incidence of intradialytic hypo...
Dialysate Sodium: Rationale for Evolution over Time
Dialysate Sodium: Rationale for Evolution over Time
AbstractOligo‐anuric individuals receiving hemodialysis (HD) are dependent on the dialysis machine to regulate sodium and water balance. Interest in adjusting the dialysate sodium ...
pO2 and pCO2 Increment in Post‐dialyzer Blood: The Role of Dialysate
pO2 and pCO2 Increment in Post‐dialyzer Blood: The Role of Dialysate
Abstract:  Blood returning from a dialyzer during hemodialysis has a higher pO2 and pCO2 content than blood entering the dialyzer, and this has been attributed to the dialysate. Th...

Back to Top