Javascript must be enabled to continue!
Phakic Intraocular Lens Explantation: Causes in 240 Cases
View through CrossRef
PURPOSE:
To describe the main causes of explantation of phakic intraocular lenses (PIOLs) according to the anatomical site of implantation (angle supported, iris fixated, or posterior chamber).
METHODS:
This multicentric, retrospective, and consecutive study sponsored by the Spanish Ministry of Health comprised a total of 240 eyes (226 patients) explanted due to PIOL complications. Clinical data of 144 angle-supported lenses, 24 iris-fixated lenses, and 72 posterior chamber lenses explanted were recorded preoperatively and postoperatively.
RESULTS:
Mean age of the patients at explantation was 46.30 ± 11.84 years (range: 25 to 80 years). The mean time between implantation and explantation was 381.14 ± 293.55 weeks (range: 0.00 to 1,551.17 weeks). It was 422.33 ± 287.81 weeks for the angle-supported group, 488.03 ± 351.95 weeks for the iris-fixated group, and 234.11 ± 4,221.60 weeks for the posterior chamber group. It was 8.10 ± 5.52 years for the angle-supported group, 9.36 ± 6.75 years for the iris-fixated group, and 4.49 ± 4.25 years for the posterior chamber group. This period of time was significantly shorter in the posterior chamber group (
P
< .001). Overall, the main causes of explantation were cataract formation (132 eyes, 55%), endothelial cell loss (26 eyes, 10.83%), corneal decompensation (22 eyes, 9.17%), PIOL dislocation/decentration (16 eyes, 6.67%), inadequate PIOL size or power (12 eyes, 5%), and pupil ovalization (10 cases, 4.17%). Cataract development was the cause of explantation in 51.39% of angle-supported cases, 45.83% of iris-fixated cases, and 65.28% of posterior chamber cases. Endothelial cell loss was the cause of explantation in 15.97% of angle-supported PIOLs, 8.33% of iris-fixated PIOLs, and 1.39% of posterior chamber PIOLs.
CONCLUSIONS:
Cataract is the main cause of PIOL explantation, especially in posterior chamber PIOLs. In the angle-supported group, endothelial cell loss was the second cause of explantation.
[
J Refract Surg
. 2015;31(1):30–35.]
Title: Phakic Intraocular Lens Explantation: Causes in 240 Cases
Description:
PURPOSE:
To describe the main causes of explantation of phakic intraocular lenses (PIOLs) according to the anatomical site of implantation (angle supported, iris fixated, or posterior chamber).
METHODS:
This multicentric, retrospective, and consecutive study sponsored by the Spanish Ministry of Health comprised a total of 240 eyes (226 patients) explanted due to PIOL complications.
Clinical data of 144 angle-supported lenses, 24 iris-fixated lenses, and 72 posterior chamber lenses explanted were recorded preoperatively and postoperatively.
RESULTS:
Mean age of the patients at explantation was 46.
30 ± 11.
84 years (range: 25 to 80 years).
The mean time between implantation and explantation was 381.
14 ± 293.
55 weeks (range: 0.
00 to 1,551.
17 weeks).
It was 422.
33 ± 287.
81 weeks for the angle-supported group, 488.
03 ± 351.
95 weeks for the iris-fixated group, and 234.
11 ± 4,221.
60 weeks for the posterior chamber group.
It was 8.
10 ± 5.
52 years for the angle-supported group, 9.
36 ± 6.
75 years for the iris-fixated group, and 4.
49 ± 4.
25 years for the posterior chamber group.
This period of time was significantly shorter in the posterior chamber group (
P
< .
001).
Overall, the main causes of explantation were cataract formation (132 eyes, 55%), endothelial cell loss (26 eyes, 10.
83%), corneal decompensation (22 eyes, 9.
17%), PIOL dislocation/decentration (16 eyes, 6.
67%), inadequate PIOL size or power (12 eyes, 5%), and pupil ovalization (10 cases, 4.
17%).
Cataract development was the cause of explantation in 51.
39% of angle-supported cases, 45.
83% of iris-fixated cases, and 65.
28% of posterior chamber cases.
Endothelial cell loss was the cause of explantation in 15.
97% of angle-supported PIOLs, 8.
33% of iris-fixated PIOLs, and 1.
39% of posterior chamber PIOLs.
CONCLUSIONS:
Cataract is the main cause of PIOL explantation, especially in posterior chamber PIOLs.
In the angle-supported group, endothelial cell loss was the second cause of explantation.
[
J Refract Surg
.
2015;31(1):30–35.
].
Related Results
Indications for exchange or explantation of phakic implantable collamer lens with central port in patients with and without keratoconus
Indications for exchange or explantation of phakic implantable collamer lens with central port in patients with and without keratoconus
AIM: To evaluate the causes of phakic implantable collamer lens (ICL) exchange/explantation in patients with and without keratoconus (KC) at two tertiary hospitals in Riyadh, Saudi...
Calculation of ocular magnification in phakic and pseudophakic eyes based on anterior segment OCT data
Calculation of ocular magnification in phakic and pseudophakic eyes based on anterior segment OCT data
Abstract
Purpose
The purpose of this study is to develop a straightforward mathematical concept for determination of obje...
The ethics of explantation
The ethics of explantation
Abstract
Background
With the increased use of implanted medical devices follows a large number of explantations. Implants are removed for a wide ran...
Clinical Applications of Anterior Segment Optical Coherence Tomography in Managing Phakic and Secondary Intraocular Lens Implants: A Comprehensive Review
Clinical Applications of Anterior Segment Optical Coherence Tomography in Managing Phakic and Secondary Intraocular Lens Implants: A Comprehensive Review
Anterior segment optical coherence tomography (AS-OCT) has emerged as a crucial imaging technique in ophthalmology, particularly for evaluating intraocular structures and the behav...
Outcomes of Pseudophakic, Phakic, and Triple DMEK
Outcomes of Pseudophakic, Phakic, and Triple DMEK
Purpose:
To evaluate the outcome of phakic and pseudophakic eyes treated by Descemet membrane endothelial keratoplasty (DMEK) versus DMEK combined with cataract surgery...
Comparative Clinical Performance of Rigid versus Soft Hyper Dk Contact Lenses Used for Continuous Wear
Comparative Clinical Performance of Rigid versus Soft Hyper Dk Contact Lenses Used for Continuous Wear
ABSTRACT
Purpose.
The purpose of this study is to investigate the comparative clinical performance of two hyper Dk contac...
Retinitis Pigmentosa
Retinitis Pigmentosa
In studying the cases with typical and atypical pigmentary degeneration of the retina we strived to analyse in the clinical material all ophthalmoscopic and ocular changes together...
The Duet-Kelman Lens: A New Exchangeable Angle-supported Phakic Intraocular Lens
The Duet-Kelman Lens: A New Exchangeable Angle-supported Phakic Intraocular Lens
ABSTRACT
PURPOSE: We describe a new model of anglesupported phakic intraocular lens, the DuetKelman lens, the design of which incorporates two separate supporting parts. ...