Refining Measurements of Power Absorbance in Newborns: Probe Fit and Intrasubject Variability

View through CrossRef
Objectives: Because unresolved debris in the ear canal or middle ear of newborns may produce high false positive rates on hearing screening tests, it has been suggested that an outer/middle ear measure can be included at the time of hearing screening. A potential measure is power absorbance (absorbance), which indicates the proportion of power in a broadband acoustic stimulus that is absorbed through the outer/middle ear. Although absorbance is sensitive to outer/middle dysfunction at birth, there is large variability that limits its accuracy. Acoustic leaks caused by poor probe fitting further exacerbate this issue. The objectives of this work were to: (1) develop criteria to indicate whether a change in absorbance occurs in association with probe fit; (2) describe the variability in absorbance due to poor fitting; and (3) evaluate test–retest variability with probe reinsertions, excluding poor fits. Design: An observational cross-sectional design was used to evaluate changes in absorbance due to probe fit and probe reinsertion. Repeated measurements were recorded in 50 newborns (98 ears) who passed TEOAE screenings and were <48 hours of age. One absorbance measurement was chosen as the baseline that served as a best-fit reference in each ear. Changes in absorbance, called absorbance probe-fit Δ, were calculated relative to the baseline in each ear. Correlations were assessed between the absorbance probe-fit Δ and low-frequency absorbance, impedance magnitude, impedance phase, and equivalent volume, to determine which measures predicted poor fits. Criteria were derived from the strongest of these correlations and their performance was analyzed. Next, measurements with poor/leaky fits were identified, and the changes in absorbance that they introduced were analyzed. Excluding the poor fits, test–retest differences in absorbance, called reinsertion Δ, were determined. Variability was assessed using the SDs associated with absorbance, absorbance probe-fit Δ, and reinsertion Δ. Results: Based on the analysis of 12 moderate–strong correlations, the following criteria were adopted to identify measurements with poor fits: (1) impedance phase-based criterion (500 to 1000 Hz) > −0.11 cycles and (2) absorbance-based criterion (250 to 1000 Hz) > 0.58. Poor-fit measurements introduced statistically significant increases in absorbance up to 0.1 for 1000 to 6000 Hz, and up to 0.4 for frequencies <1000 Hz. Reinsertion Δ were ≤0.02, and were significant for 500 to 5000 Hz. The SDs of absorbance probe-fit Δ were greatest and similar to overall absorbance SD in the low frequencies. Separately, the SDs of reinsertion Δ were also greatest and similar to low-frequency absorbance SD. Conclusions: Poor probe fits introduced the greatest inflation in absorbance for frequencies < 500 Hz, and a smaller but significant inflation for higher frequencies, consistent with controlled experiments on acoustic leaks in adults. Importantly, inflation of absorbance in diagnostically sensitive 1000 to 2000 Hz may impact its clinical performance. Test–retest with probe reinsertion contributed significantly to absorbance variability, especially in the low frequencies, consistent with reports in adults, even though changes were smaller than those associated with poor probe fit. The results indicate that variability in absorbance was reduced by minimizing acoustic leaks. Pending further validation, the probe-fit criteria developed in this work can be recommended to ensure proper probe fit.
Ovid Technologies (Wolters Kluwer Health)
Title: Refining Measurements of Power Absorbance in Newborns: Probe Fit and Intrasubject Variability
Description:
Objectives: Because unresolved debris in the ear canal or middle ear of newborns may produce high false positive rates on hearing screening tests, it has been suggested that an outer/middle ear measure can be included at the time of hearing screening.
A potential measure is power absorbance (absorbance), which indicates the proportion of power in a broadband acoustic stimulus that is absorbed through the outer/middle ear.
Although absorbance is sensitive to outer/middle dysfunction at birth, there is large variability that limits its accuracy.
Acoustic leaks caused by poor probe fitting further exacerbate this issue.
The objectives of this work were to: (1) develop criteria to indicate whether a change in absorbance occurs in association with probe fit; (2) describe the variability in absorbance due to poor fitting; and (3) evaluate test–retest variability with probe reinsertions, excluding poor fits.
Design: An observational cross-sectional design was used to evaluate changes in absorbance due to probe fit and probe reinsertion.
Repeated measurements were recorded in 50 newborns (98 ears) who passed TEOAE screenings and were <48 hours of age.
One absorbance measurement was chosen as the baseline that served as a best-fit reference in each ear.
Changes in absorbance, called absorbance probe-fit Δ, were calculated relative to the baseline in each ear.
Correlations were assessed between the absorbance probe-fit Δ and low-frequency absorbance, impedance magnitude, impedance phase, and equivalent volume, to determine which measures predicted poor fits.
Criteria were derived from the strongest of these correlations and their performance was analyzed.
Next, measurements with poor/leaky fits were identified, and the changes in absorbance that they introduced were analyzed.
Excluding the poor fits, test–retest differences in absorbance, called reinsertion Δ, were determined.
Variability was assessed using the SDs associated with absorbance, absorbance probe-fit Δ, and reinsertion Δ.
Results: Based on the analysis of 12 moderate–strong correlations, the following criteria were adopted to identify measurements with poor fits: (1) impedance phase-based criterion (500 to 1000 Hz) > −0.
11 cycles and (2) absorbance-based criterion (250 to 1000 Hz) > 0.
58.
Poor-fit measurements introduced statistically significant increases in absorbance up to 0.
1 for 1000 to 6000 Hz, and up to 0.
4 for frequencies <1000 Hz.
Reinsertion Δ were ≤0.
02, and were significant for 500 to 5000 Hz.
The SDs of absorbance probe-fit Δ were greatest and similar to overall absorbance SD in the low frequencies.
Separately, the SDs of reinsertion Δ were also greatest and similar to low-frequency absorbance SD.
Conclusions: Poor probe fits introduced the greatest inflation in absorbance for frequencies < 500 Hz, and a smaller but significant inflation for higher frequencies, consistent with controlled experiments on acoustic leaks in adults.
Importantly, inflation of absorbance in diagnostically sensitive 1000 to 2000 Hz may impact its clinical performance.
Test–retest with probe reinsertion contributed significantly to absorbance variability, especially in the low frequencies, consistent with reports in adults, even though changes were smaller than those associated with poor probe fit.
The results indicate that variability in absorbance was reduced by minimizing acoustic leaks.
Pending further validation, the probe-fit criteria developed in this work can be recommended to ensure proper probe fit.

Related Results

Wideband tympanometry and absorbance measurements in otosclerotic ears
Wideband tympanometry and absorbance measurements in otosclerotic ears
ObjectivesThis study aimed to assess absorbance using a thorough analysis of individual points of its plot in ears with intraoperatively confirmed otosclerosis. To the best of our ...
Maternal cigarette smoking and its effect on neonatal lymphocyte subpopulations and replication
Maternal cigarette smoking and its effect on neonatal lymphocyte subpopulations and replication
AbstractBackgroundSignificant immunomodulatory effects have been described as result of cigarette smoking in adults and pregnant women. However, the effect of cigarette smoking dur...
Perceived Community Acceptance of Maternal - Newborns Care Continuum and its Correlates in Ethiopia
Perceived Community Acceptance of Maternal - Newborns Care Continuum and its Correlates in Ethiopia
Abstract Background: Maternal and newborns care continuum service use decision making process is influenced by seeking validation and the sole approval from significant oth...
Incidence, Course, and Prediction of Hyperbilirubinemia in Near-Term and Term Newborns
Incidence, Course, and Prediction of Hyperbilirubinemia in Near-Term and Term Newborns
Objective. In this study, we investigated prospectively the incidence of significant hyperbilirubinemia and demographic and laboratory characteristics and pattern of serum bilirubi...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...

Back to Top